1 / 6

20 cm and 33mm MCP Test Progress

20 cm and 33mm MCP Test Progress. Lifetest with #164/163 continues – now close to 2 C cm -2 New batch of 33mm test MCPs arrived Chem-3 first tests begun Need to choose MCPs for next lifetests : - Bake , and scrub with a pair of the new MCPs

mervyn
Download Presentation

20 cm and 33mm MCP Test Progress

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 20 cm and 33mm MCP Test Progress • Lifetest with #164/163 continues – now close to 2 C cm-2 • New batch of 33mm test MCPs arrived • Chem-3 first tests begun • Need to choose MCPs for next lifetests:- • Bake, and scrub with a pair of the new MCPs • Find out if air exposure affects post bake/scrub SEE • Investigate temp stability of gain • Evaluate background sources • No new 20cm MCP data – awaiting new MCPs • Have made changes to 20cm detector to improve • imaging for next set of tests. • Continue to electrode for 20cm MCPs, 2 more coming

  2. ALD Chem 3, 20µm Microchannel Plates 12317-120/117, 50MΩ/180MΩ, Pair with 0.7mm gap/400v 900v on each MCP, gain high, >107. Numerous black spot defects, lots of bright spots due to removed/missing ALD layer exposing electrode material. Cleanliness issues? Gain map, multifiber pattern modulation, typical of 12317 boule material with pore size variations. Not a good substrate material.

  3. Chem-3 ALD 20µm Microchannel Plate Previous data on 12317 boule 20 µm pore #160 ALD MCP with NiCrelectrode . 12317-120/117 Chem-3 background 100 sec background, ~4 events cm-2 sec-1. Some hot patches, uniform background rate is 3.5 events cm-2 sec-1 ----not very good. #160, 12317-113, 700v MCP, earlier data shows problems with this material

  4. ANL ALD-MCP Samples Started with Chem-3 12137 boule has bad pore stacking uniformity 11863 boulehas good pore stacking uniformity 12258 boule is unknown 11863-88

  5. Preconditioning Test of 33mm 20µm Pore MCPs ANL MCPs 164 and 163, 20µm, 60:1, 8° bias. INCOM serial #s, 11863-99, 11863-95 Resistances 159MΩ, 138MΩ (NiCr over) XDL detector with MCPs gapped by 0.001” spacer. Completed 350°C bake, with RGA scans. Scrub in progress, with RGA scans. Have gone from 1µA to 2µA, now 3µA >1.9 C cm-2 extracted.

  6. MCP 164-163 Scrub Record Almost all of the gain variation is due to thermal changes in the lab Scrub @ 3 x 105 gain, 700v per MCP

More Related