48 hour tool. Step by Step. Purpose of 48-hour tool. Available for NC and RC Removal of placed implant from patient Substitute for transfer part after transfer part failure Implant needs to be discharged after removal
E N D
48 hour tool Step by Step
Purpose of 48-hour tool • Available for NC and RC • Removal of placed implant from patient • Substitute for transfer part after transfer part failure • Implant needs to be discharged after removal • Can be done up to 48 hours after placement, afterwards osseointegration is too far advanced! • Do not use for height correction Education
Situation • Implant stuck & transfer piece breaks Education
Situation • Implantstuck& transfer piece breaks • Transfer part has predetermined breakage point: • RC: 100 Ncm • NC: 80 Ncm • Up until this point, the internal connection will not be damaged • The breakage point is located at the head of the transfer screw Education
Removing the Implant • Remove adapter with screw head Education
Removing the Implant • Remove transfer nut from implant (no unscrewing needed) • The remaining screw remains secure within the implant CAUTION • Make sure the transfer nut is not aspirated Education
Removing the Implant • Remove remaining screw with tweezers or pliers by counter clockwise rotations CAUTION • Make sure the remaining screw is not aspirated Education
Removing the Implant • Mount 48-hour tool like an abutment • Secure tool by hand-tightening the mounting screw Education
Removing the Implant • Mount ratchet and slowly unscrew implant • Remove and sacrifice the implant Education