An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Without reference, identify principles about volume/pressure and high frequency ventilators with at least 70 percent accuracy.

  2. Volume/Pressure and High Frequency Ventilators • Purpose of Volume/Pressure Ventilators • Device used to move gas into the lungs • Required when • Acute respiratory failure • No absolute rule

  3. Volume/Pressure and High Frequency Ventilators • Categories • Based on the type of pressure used to move gas into the lungs • Negative pressure ventilators • Apply a sub-atmospheric pressure around the chest • Air at atmospheric pressure is drawn in • An example is an iron lung • Major disadvantages • Cannot provide adequate ventilation in all patients • Problems with fit and comfort during long-term use

  4. Volume/Pressure and High Frequency Ventilators • Pressure ventilators • Apply a positive pressure directly to the airway • Similar to mouth-to-mouth artificial ventilation • Provides better control over ventilation • Most common • Disadvantage is that the increased pressure in the thoracic cavity decreases venous blood return to the heart

  5. Volume/Pressure and High Frequency Ventilators • Patient Connection Methods • Face mask • Disadvantages • Difficult to get a good seal • Airway blockage may occur • Acceptable for short term emergency use only • Intubation • A rigid tube inserted into the trachea • Types • Endotracheal tubes may be inserted through the mouth or nose • Tracheostomy tube is surgically inserted into the trachea • Use a balloon-like cuff which seals against the walls of the trachea • All gas flow is through the intubation tube

  6. Volume/Pressure and High Frequency Ventilators • Limiting Devices - Required to prevent barotraumas (lung damage) • Pressure limiter • Pressure developed by the drive mechanism may be high enough to injure the lungs • Methods • Vent excess pressure • End the inspiration completely

  7. Volume/Pressure and High Frequency Ventilators • Flowrate control • When the generated pressure is high there must be some means of controlling flow rate • Methods • Increasing resistance • With injectors, decreasing the driving pressure

  8. Volume/Pressure and High Frequency Ventilators • Phases of the Ventilation Cycle • Inspiratory phase - moves gas into the lungs • Changeover from inspiration to expiration • Cycling mechanisms determine when to end an inspiration and begin an expiration • There are four ways of cycling the ventilator into an expiratory phase • Pressure cycled ventilators • Inspiration is ended when a predetermined pressure is reached • Volume, flow rate, and inspiratory time may all vary from one inspiration to another

  9. Volume/Pressure and High Frequency Ventilators • Flow cycled ventilators • Inspiration is ended when the flow of gas into the patient falls below a specific point • Volume, pressure, and inspiratory time may vary from one inspiration to another • Volume cycled ventilators • Inspiration is ended when a predetermined volume of gas has been delivered to the patient • Pressure, flow rate, and inspiratory time may vary from one inspiration to another

  10. Volume/Pressure and High Frequency Ventilators • Time cycled ventilators • Inspiration is ended after a predetermined time • Volume, pressure, and flow rate may vary from one inspiration to another • Inspiratory supplemental functions • Sigh- a periodic deep breath, usually delivered 6 to 10 times an hour • Inflation hold • It is a period of time at the end of an inspiration that the lungs are kept inflated • Also known as inspiratory hold, inspiratory pause, or inspiratory plateau

  11. Volume/Pressure and High Frequency Ventilators • Expiratory phase • Allows the lungs to empty • Normally not restricted by the ventilator • Adjuncts of the expiratory phase • Expiratory resistance • Application of mechanical resistance to the exhalation port of the breathing circuit • Increases the mean airway pressure

  12. Volume/Pressure and High Frequency Ventilators • Positive end-expiratory pressure (PEEP) • Maintains a positive pressure in the lungs throughout the respiration cycle • Used during mechanical breathing • Purposes • Increases the mean airway pressure and functional residual capacity • May improve lung compliance

  13. Volume/Pressure and High Frequency Ventilators • Continuous positive airway pressure (CPAP) • Maintains a positive pressure in the lungs throughout the respiration cycle • Used during spontaneous breathing • Purposes: • Keeps airways open • Increases the mean airway pressure and functional residual capacity

  14. Volume/Pressure and High Frequency Ventilators • Changeover from expiration to inspiration • Cycling mechanisms determine when to end an expiration and begin an inspiration • Three ways of cycling the ventilator into an inspiratory phase • Manual cycling - expiration is ended and inspiration is begun by operator intervention • Patient cycled • Started when the patient attempts to take a breath • Once initiated the ventilator delivers a controlled breath • Breath is known as an "assisted" breath • The sensitivity or trigger sensitivity parameter must be set to the patient's efforts

  15. Volume/Pressure and High Frequency Ventilators • Timed cycled • Inspiration begins after a predetermined time has passed • Timer is independent of the patient • Controlled by the "breaths/min" setting • Classified as a "strict controller" if no assist mechanism is used • Often used as a backup to patient cycling

  16. Volume/Pressure and High Frequency Ventilators • Special Ventilator Modes • Developed to make it less tedious to wean patient from the ventilator • Intermittent mandatory ventilation (IMV) • Allows the patient to breathe spontaneously between controlled breaths • May be divided into two phases • IMV phase • Spontaneous phase • Controlled IMV breath is delivered at the beginning of the IMV phase • Disadvantage is that the ventilator can inspire against the patient • Uncommon

  17. Volume/Pressure and High Frequency Ventilators • Synchronized intermittent mandatory ventilation (SIMV) • Allows the patient to breathe spontaneously between assisted breaths, with a mandatory breath as backup • May be divided into two phases • SIMV phase • Spontaneous phase • Assisted breath in SIMV phase is synchronized to patient efforts • If no assisted breath is taken during the SIMV phase • A mandatory SIMV breath will be given • At the beginning of the next spontaneous phase

  18. Volume/Pressure and High Frequency Ventilators • Mandatory minute ventilation (MMV) • Additional option to SIMV • Number of mandatory breaths is increased or decreased dynamically • Used to ensure a specific minute ventilation • High Frequency Ventilation (HFV) • Uses breath rates of 1 HZ to 20 HZ • Purpose • Low diffusion • Patient requires higher pO2 (partial pressure of oxygen) gradient to force O2 to diffuse • Allows the Paw to be reduced but keeps pO2 high • Infants • Require a higher breathing rate • Have a lower compliance

  19. Volume/Pressure and High Frequency Ventilators • GENERAL CALIBRATION PROCEEDURES FOR VOLUME/PRESSURE VENTILATORS • Ensure annual manufacturer’s calibration • Ensure equipment is designed to accurately measure the delivery from the unit under test • High flow adult • Low flow pediatric • High frequency • Obtain ventilator information • What settings must the ventilator be placed in to measure volumes and flows • Test device must be set accordingly to ensure proper measurements • Failure to do so may result in the maladjustment of the ventilator

  20. Volume/Pressure and High Frequency Ventilators • Ventilator Measurement Standards • Flow and volume measurements are affected by • Type of gas (air, oxygen or mixed) • Temperature • Pressure • Water vapor • BTPS – Body Temperature Pressure Saturated • Body temperature – 37° C • Pressure – ambient barometric pressure • Saturated – water vapor at 100% relative humidity • Usually used when the flow of gases has been changed to patient conditions (exhalation)

  21. Volume/Pressure and High Frequency Ventilators • STPD – Standard Temperature Pressure Dry • Standard temperature – usually 0 or 21° C • Pressure – ambient barometric pressure • Dry – water vapor at 0% relative humidity • Usually used when the flow of gases has been changed to standard conditions (wall supply/inhalation)