540 likes | 2.58k Views
SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS . SERVO-i System Version 4.0. Heliox option Low MV alarm audio off option Possibility to switch off back-up ventilation Other usability improvements. HELIOX – GAS PROPERTIES. What is Heliox?.
E N D
GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 • Heliox option • Low MV alarm audio off option • Possibility to switch off back-up ventilation • Other usability improvements
HELIOX – GAS PROPERTIES What is Heliox? • Heliox – HeO2 – is a mixture of helium (He) and oxygen (O2) (e.g. 80/20; 70/30). • Helium is an inert, low density gas (~1/7 the density of air and oxygen). • Lower density leads to a more laminar flow (less turbulence) and thus reduces resistance in the airways.
HELIOX – GAS PROPERTIES • Laminar gas flow (A) is more efficient than turbulent flow (B).
HELIOX – EFFECTS ON VENTILATION Effects of Heliox therapy according to the literature • Established positive effects on ventilation • Reduces work of breathing • Increases CO2 diffusion • Potential positive effects on ventilation • Reduces the peak pressure needed to maintain flow • Increases CO2 elimination and improves gas exchange • Increases expiratory flow • Reduces respiratory rate and auto PEEP • Increases drug deposition in distal parts of the airways.
HELIOX – USES AND APPLICATIONS For which patients is Heliox most commonly used? • Upper airway obstruction • Lower airway obstruction
HELIOX – POTENTIAL BENEFITS Potential benefits of Heliox therapy according to the literature • Avoiding endotracheal intubation • Protecting the lungs • Improving patient outcome • Enhancing patient comfort
SERVO-i WITH HELIOX OPTION General • Available as an option for the SERVO-i • Field upgrade • Factory installed • The option consists of: • Heliox SW • Heliox adapter kit
SERVO-i WITH HELIOX OPTION Functionality • Can be used with all pediatric and adult patients, as well as neonates • Compatible with all invasive/non-invasive ventilation modes • Supporting mobility from emergency room to MR environment • Connection of 80/20; 79/21; 78/22 mixtures possible • Adjustable O2 concentration (21 to 100%).
SERVO-i WITH HELIOX OPTION Key benefits of using SERVO-i for Heliox therapy • Unique solution for automatic gas identification • Automatic compensation of flow delivery • Automatic compensation of CO2 and volume monitoring
SERVO-i WITH HELIOX OPTION Key benefits of using SERVO-i for Heliox therapy • Quick and easy start of therapy • Quick and easy switch from air to Heliox (and vice versa) without going to standby • User friendly interface • Cost efficient Heliox therapy thanks to the SERVO-i’s low gas consumption.
SERVO-i WITH HELIOX OPTION Heliox adapters • Four different adapters are available: • DISS (both air (A) and Heliox (B)) • NIST/NIST • AGA/NIST • French/NIST
SERVO-i WITH HELIOX OPTION Heliox adapters • Color marked for Heliox • Brown/green for Heliox DISS • Brown/white for Heliox NIST • Attached to ventilator with safety wire • Attach to air module when using Heliox
SERVO-i WITH HELIOX OPTION Connecting Heliox to the ventilator – Heliox adapters NIST / AGA / French • Remove adapter from holder. • Connect adapter to Heliox high pressure hose. • Connect adapter to air module inlet.
SERVO-i WITH HELIOX OPTION Preparations before using Heliox with SERVO-i
SERVO-i WITH HELIOX OPTION Automatic gas identification • Performed by the ultrasonic expiratory flow sensors • Available during Pre Use Check and during ventilation • Active as long as the set O2 concentration is 75% or below • Time for identification is determined by tidal/minute volume and O2 concentration used
SERVO-i WITH HELIOX OPTION Pre Use Check
SERVO-i WITH HELIOX OPTION Automatic identification when changing from air to Heliox during ventilation
SERVO-i WITH HELIOX OPTION Gas type change – manually initiated compensation for Heliox gas
SERVO-i WITH HELIOX OPTION Gas type change – manually initiated compensation for Heliox gas
SERVO-i WITH HELIOX OPTION Gas type change – manually initiated compensation for Heliox gas
SERVO-i WITH HELIOX OPTION Non-invasive application
SERVO-i WITH HELIOX OPTION Infant application
SERVO-i WITH HELIOX OPTION Order information • The Heliox option is not compatible with the O2 Sensor and Y Sensor measuring.
SERVO-i 4.0 – OTHER IMPROVEMENTS/CHANGES Low Minute Volume alarm audio off
SERVO-i 4.0 – OTHER IMPROVEMENTS/CHANGES Low Minute Volume alarm audio off
SERVO-i 4.0 – OTHER IMPROVEMENTS/CHANGES Low Minute Volume alarm audio off
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES Possibility to disable back-up ventilation
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES Possibility to disable back-up ventilation
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES Possibility to disable back-up ventilation
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES Showing previous Pre Use Check/Suction Support FiO2 retention
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES Maintenance notifications/Changed calculation of time for next PM
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES Previous mode information/Initiating a Pre Use Check
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES New alarm symbol/Clearer difference between medium and high priority alarms
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES New patient category symbols
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES Time scale on the waveforms Art. no. 6679957 rev. 01