Biometry: Long & Short eyes

Biometry: Long & Short eyes. Alireza Peyman, MD Isfahan University of Medical Sciences. Normal Range:. Axial length Biggest source of error in IOL power calculations. 96% of axial lengths fall within the range 21.0 to 25.5 mm 60% of AL is between 22.5 and 24.5 mm.

moanna
Download Presentation

Biometry: Long & Short eyes

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

  2. Normal Range: • Axial lengthBiggest source of error in IOL power calculations. • 96% of axial lengths fall within the range 21.0 to 25.5 mm • 60% of AL is between 22.5 and 24.5 mm

  3. Ultrasounds are calibrated with average velocities for normal length eyes. These velocities are incorrect for short eyes, causing significant measurement errors. • Accurate measurement of axial length in hyperopic eyes is especially important since any error is greatly magnified in proportion to the length of the eye. • Immersion or LASER biometry can provide superior results in these cases.

  4. In a hyperopic eye with axial length of 20 millimeters each mm error in measurement cause 3.75 D post operative refractive error.

  5. Difference in axial length of more than 0.3mm requires confirmation • Most individuals have similar axial lengths in each eye except: • unilateral refractive error • coloboma • Staphyloma

  6. Myopia

  7. In the axial myopia, the presence of a staphyloma should be considered, especially if there is difficulty obtaining a distinct retinal spike during A-scan. • an immersion A/Bscan approach to axial length measurement in the setting of a posterior staphyloma

  8. nearly all eyes with pathologic myopia have some form of posterior staphyloma. • For highly myopic eyes (axial > 28mm) a B-scan should be carried out to determine the presence or otherwise of staphylomata

  9. Long eye, foveal deep

  10. Very long eye, posterior pole staphyloma

  11. Buphthalmic globe: very long eye (37·5 mm): deep anterior chamber.

  12. Megalocornea: average length eye (23·1 mm): deep anterior chamber (5·2 mm).

  13. The Ascan vector is adjusted to pass through the middle of the cornea as well as the middle of the anterior and posterior lens echoes to assure that the vector will intersect the retina in the region of the fovea. • IOL Master permits accurate measurements when posterior staphylomata are present. Since the patient fixates along the direction of the measuring beam.

  14. IOL power calculation formula

  15. Calculation formula: • The SRK/T and Holladay formulae worked best overall. • SRK I and II were found to be least accurate • No statistical difference was found between SRK/T, Hoffer Q and the Holladay formulae

  16. When using IOL Master do not forget • For AL < 22 do immersion A scan • For AL > 30 do A and B scan

  17. Silicone oil filled vitreous cavity: eye measures 49 mm on B-scan, but actual axial length is 34·3 mm.