Doppler echocardiography & tissue doppler

Doppler echocardiography & tissue doppler. Dolly mathew. Properties of blood. Hemodynamics - physical principles of blood flow & circulation Density – mass per unit volume(g/ml) Resistance to acceleration

nerina
Download Presentation

Doppler echocardiography & tissue doppler

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Doppler echocardiography & tissue doppler Dolly mathew

  2. Properties of blood • Hemodynamics- physical principles of blood flow & circulation • Density – mass per unit volume(g/ml) Resistance to acceleration • Viscosity – ability of molecules to move past one another by overcoming frictional forces 0.035 poise at 37◦c • Flow occurs from high pressure to low pressure end

  3. R= 8Lv/ ∏r2 • V viscosity of blood • R radius of lumen • L length of the vessel • Q= ∆P/R poiseuille’s law • Q= ∆P∏r2/ 8Lv • Q= ∆P∏r2

  4. Types of flow

  5. Laminar flow • Acceleration of flow- flat flow profile / plug flow • Converging flow- flat profile parabolic profile • Diverging flow - multiple flow patterns(uniform high velocity flow, stagnant flow, eddy flow) • Vessel curvature – high velocity in the inner part of curve in the ascending limb, outer part of the curve in descending limb

  6. Turbulant flow • Valve obstruction, regurgitation,septal defects • Increased velocities , flow vortices • Predicted by reynold’s number Re= ᵨcd/v

  7. Laminar flow • narrow spectral envelope-most cells travel over a narrow range of velocities • Large spectral window-echo free area under spectral doppler trace • Turbulant flow • Spectral widening-direction and range of velocities increase-greater range of doppler shift frequencies • Spectral window diminished

  8. Continuity principle fig

  9. Doppler echocardiography utilizes ultrasound to record -direction,velocity and pattern of blood flow • based upon the changes in frequency of the backscatter signal from small moving structures, ie, red blood cells, intercepted by the ultrasound beam

  10. Comparison of 2-D echo and doppler • 2-D echocardiography • target is tissue • Type of information is structural • Optimal alignment is perpendicular • Preferred transducer frequency is high • Doppler imaging • Blood is target • Obtain information on physiology • Parallel alignment between beam and target • Low transducer frequency is preferred

  11. BASIC PRINCIPLES • A moving target will backscatter an ultrasound beam to the transducer • the frequency observed when the target is moving toward the transducer is higher • the frequency observed when the target is moving away from the transducer is lower than the original transmitter frequency

  12.  Doppler shift (F[d]) = F[r] - F[t] • F[d] = 2 x F[t] x [(V x cos ø)] ÷ C Blood flow velocity (V) speed of sound in blood (C) ø, the intercept angle between the ultrasound beam A factor of 2 is used to correct for the "round-trip" transit time to and from the transducer.

  13. This equation can be solved for V, by substituting (F[r] - F[t]) for F[d]: •   V = [(F[r] -F[t]) x C] ÷ (2 x F[t] x cos ø) • the angle of the ultrasound beam and the direction of blood flow are critically important in the calculation • ø of 0º and 180º (parallel with blood flow), cosine ø = 1 • ø of 90º (perpendicular to blood flow), cosine ø = 0 , the Doppler shift is 0 • ø up to 20º, cos ø results in a <10 percent change in the Doppler shift • ø of 60º, cosine ø = 0.50

  14. Angle of doppler beam in relationship to direction of blood flow

  15. Doppler effect

  16. Spectral analysis — •  the difference between the transmitted and backscattered signal is determined by comparing the two waveforms with the frequency content analyzed by fast Fourier transform (FFT). The display generated by this frequency analysis is termed spectral analysis. • By convention, time is displayed on the x axis and frequency shift on the y axis.

  17. Spectral display • Intensity/amplitude -proportional to the number of blood cells with that velocity-represented by brightness of the signal.

  18. Pressure determination • ∆P pressure difference between 2 points • V1 proximal,v2 distal velocity • ᵨ density of liquid • dv change in velocity over the time period • ds distance over which pressure decrease • R viscous resistance in the vessel • v velocity of blood flow • Simplified ∆P = 4(V2²-V1²) =4V²

  19. Doppler formats • Continous wave doppler • Pulse wave doppler • Colour flow imaging • Tissue doppler

  20. Continous wave doppler • Continuous transmission of doppler signal towards the moving RBCs & continuous reception of reflected signals • Blood flow along the entire beam is observed • Range resolution not possible • High velocities can be recorded • Aliasing does not occur

  21. Pulse wave doppler • Short intermittent bursts of ultrasound emitted • Listens only at a fixed and brief time interval to receive signals from a specific distance • Select doppler information from a particular location using sample volume • PRF-number of pulses transmitted from transducer each second,determines the sampling rate • PRF limits the max.velocity detectable-Nyquist limit :PRF/2 • PRF/2= ±∆f = 2 f○ v cosø/C • V = C PRF/ 4 f○ v cosø

  22. Sample volume • three-dimensional, teardrop shaped portion of the ultrasound beam  • width is determined by the width of the ultrasound beam at the selected depth. • length determines the length of time that the transducer is activated to receive information from sv location

  23. main disadvantage of PW - inability to accurately measure high blood flow velocities , limitation technically known as "aliasing" • inability of pulsed Doppler to faithfully record velocities above 1.5 to 2 m/sec when the sample volume is located at standard ranges in the heart • Aliasing is represented on the spectral trace as a cut-off of a given velocity with placement of the cut section in the opposite channel or reverse flow direction.

  24. Aliasing

  25. Control of aliasing ◦ V = C PRF/4f◦cosø • Increasing the PRF D = c t /2 ; c propagation speed thro’tissue t time taken for us signal to return to the transducer 2 pulse must travel to the structure & then back again T= 2D/C PRF = 1/T = C/2D • Decreasing the transmitted frequency

  26. Baseline shift ("zero shift" or "zero off-set" ) Repositioning of zero baseline effectively increases the maximum velocity in one direction, at the expense of other direction

  27. Changing from PW to CW • Utilising high PRF mode: transmission of any given pulse occurs before the reception of all the echoes from the previous pulse • Use multiple sample gates at various locations • Signals received at different depths simultaneously • Disadvg- exact location of doppler shift is not known

  28. The spectral outputs from PW and CW appear differently • When there is no turbulence, a laminar (narrow band) spectral output- PW • CW - all the various velocities encountered by the ultrasound beams are detected by CW

  29. Comparison between CW & PW

  30. Doppler audio signal • audio outputs • High pitched sounds -large Doppler shifts -high velocities • low pitched sounds -lesser Doppler shifts • Flow direction information • stereophonic audio output • Flow • Laminar flow -smooth, pleasant tone –uniform V. • Turbulent flow -high-pitched and whistling or harsh and raspy sound- different velocities

  31. It can usually be said that when an operator wants to know where a specific area of abnormal flow is located that PW Doppler is indicated. • When accurate measurement of elevated flow velocity is required, CW Doppler should be used

  32. The Use of the Doppler Controls • Gray Scale : the gray scale control provides a means of altering the various ranges of gray (from white to black) on the spectral display and has no effect on the audio output of the Doppler system. • Different Doppler instruments have from two to more than eight different ranges of gray scale display.   • Lighter shades of gray indicate that there are fewer red cells moving at that velocity in comparison to darker shades of gray or black where many red cells are moving at that velocity.

  33. The use of this control at eight different gray scale settings is demonstrated • A spectral recording of normal aortic flow with properly set gain and gray scale

  34. The problem with leaving the gray scale control at the maximum setting is that a light level of gray is assigned to low amplitude background noise in the spectral trace. • balanced adjustment between the gain control and the gray scale control ,so that the cleanest spectral trace with the most shades of gray is displayed.

  35. Uses • Doppler echo may offer a valid alternative to invasive cardiac catheterization for hemodynamic assessment of patients with advanced heart failure • it may assist in monitoring and optimization of therapy in heart transplant recipients.

  36. Optimisation of doppler signals • Angle dependency-parallel alignment of blood flow & ultrasound beam • Sample volume position-placed where the blood flow is most parallel to us beam - increasing the depth of the sample volume;reduces PRF; lowers the maximum velocity that can be displayed • Velocity scale adjusts the maximum velocity that can be displayed • Baseline is a horizontal line with zero doppler shift; velocities towards the transducer above the baseline; away from transducer below the baseline

  37. Wall filters- eliminates the low frequency doppler shifts that typically occurs due to motion of heart valves or heart walls • Gain function adjusts the degree of amplification of received doppler signals • Should be adjusted to optimally display the entire doppler spectrum without any excessive background noise • Sample volume length- 2-5mm, avoid spectral broadening

  38. Doppler examination of mitral valve inflow • PW sample volume at the tip of mv leaflets. • IVRT (Ao valve closure –m v opening)-can be measured by displaying mv inflow signal & LVOT signal in the same spectral trace • E- early rapid filling • Deceleration slope • Diastasis- equalisation of pressures-low uniform velocities close to the baseline • Atrial filling phase- Awave • Stroke volume thro’ mv-sv at mitral annulus

  39. Doppler examination of LVOT &Ao • Apical 5 chamber view – LVOT, Ao • Pw sample vol (3-5mm) just proximal to the aortic valve • Ao flow – cw ideal • Desc Ao- suprasternal long axis, sv 1cm distal to the origin of left subclavian artery • Below the baseline,v shaped, steep acceleration & deceleration

  40. Doppler examination of PV flow • Apical 4 chamber view • Sv (3-5mm) placed 1-2cm into RUPV • Systolic forward flow • Diastolic forward flow • Atrial flow reversal

  41. Doppler examination of TV inflow • Plax RV inflow tract/apical 4 chamber view • Positioned centrally between open leaflet tips on the RV side

  42. Doppler examination of RVOT & PA • Sv within the outflow tract , 1cm prox to pulmonary valve leaflets • PA – centre of PA 1cm below the pulmonary valve

  43. Doppler examination of hepatic veins • subcostal long axis view of ivc • Sv placed 1-2cm into the hepatic vein proximal to its jn with IVC • Systolic forward flow • Diastolic forward flow • Atrial flow reversal

  44. Doppler examination of svc • Suprasternal notch or rtsupraclavicularfossa • Sv 5-7 cm into ivc

  45. Doppler artifacts • Mirror imaging or crosstalk • Symmetric spectral image on opposite side • Less intense • Decrease power output,optimise alignment • Ghosting • Brief displays of colour painted over regions of tissues • Usually solid colour,occur into tissue area of image • Produced by motion of strong reflectors

  46. Basic principles of colour doppler imaging • Produced by using multiple sample gaits along multiple scan lines • Where doppler signals are detected, pixels representing that areas are designated a colour • Colour coding relative to the transducer is direction sensitive

  47. Blood flow direction – BART system • Blood flow velocity- low velocity flow indicated by colours closest to colour baseline - Appear in deeper colour hues - High velocity flow – towards the end of colour bar, appears brighter - No angle correction -Peak velocity estimations are not possibe -Only mean doppler velocities are depicted

  48. Frequency aliasing -appears as colour reversal • Timing of the colour doppler signals- achieved by observing CFI in relation to ECG • Laminar vsturbulant flow – smooth homogenous pattern; RBCs move at about the same velocity & in the same general direction • Turbulant flow- disorganised mosaic pattern containing all colours on the colour bar

  49. Optimisation of colour flow doppler images • Frame rate- no of frames produced per second • Depends upon depth,colour sector width,& line density • Velocity scale- djusts the maximum velocity that can be displayed • Wall filters • Gain • Angle dependency-best doppler signal when direction of blood parallel to the us beam