1 / 47

Staying Focussed..

Staying Focussed. An introduction to stable isotope mass spectroscopy. Stable Isotope analysis. Sample preparation Chemically convert sample material (ie rocks, water, biological materials) into gas Quantitative Measurement of isotope ratios Mass spectroscopy

niesha
Download Presentation

Staying Focussed..

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Staying Focussed.. An introduction to stable isotope mass spectroscopy

  2. Stable Isotope analysis • Sample preparation Chemically convert sample material (ie rocks, water, biological materials) into gas Quantitative • Measurement of isotope ratios Mass spectroscopy Laser cavity molecular spectroscopy • Normalization of results Laboratory references International standards

  3. Stable Isotope mass specs are gas-source D/H H2 18O/16O CO2, CO, O2 13C/12C CO2, CO 15N/14N N2 34S/32S SO2, SO, SF6 37Cl/35Cl CH3Cl

  4. Combust: (C6H10O5)n + O2 CO2 + H20 (C6H10O5)n + C CO + H2 Reduce: H20 + Zn H2 + ZnO React: SiO2 + BrF5 O2 + SiF2 O2 + C CO2 Equilibrate: C16O2 + 2H218O C16O18O + 2H216O18O

  5. Purification Vacuum lines Cryogenic (LN2) traps for separation of gasses Reaction vessels for chemical reactions in vacuum Usually used in conjunction with “off-line” isotope analysis Necessary for some analyses ie silicate analyses

  6. Gas Chromatography Uses a GC column to separate gasses He TCD detectors usually in a singe instrument as a preparatory inlet to a mass spectrometer Combustion/Reduction Automation “On-Line” Mass Spec

  7. Mass spectrometers • JJ Thomspon 1910 – parabola spectrograph • Discovered first stable isotopes (Ne mass 20-22) • Discovered the electron • Awarded nobel prize 1906 • F.W.Aston – mass spectrograph • Discovered 21Ne • 212 out of 287 naturally occurring isotopes • Mass defect – binding energies • Awarded nobel prize 1922

  8. A Nier • First stable isotope abundance instrument • Electron-impact source • Dual detectors • Magnetic sector • Electronic rather than photographic ion counting

  9. A lot has changed, a lot has stayed the same……

  10. Inlet System: DI or CF Source of ions M+ Pumping system Diffusion or turbo pumps Analyser = Magnetic sector Detectors -Faraday cups -electronic ion counting

  11. Ion Source Electron Impact source: M + e- M+ + 2e- Electron energies ca. 100 eV Electron emission 1mA or 6x1015 e-/s Efficiency = 1 in 2400 molecules ionized Problems: -Linearity current ≠ const measured ratios -memory -stability -chemical inertness of hot filiament

  12. Ionization efficiency - sensitivity Ca 70V Electrons about 70 eV – de Broglie l is about equal to molecule bond lengths About 1 in 1000 impacts give ionization Emission is about 1 mA Cross section is low – 10-7 mm2 About 1nA ion current from 1mA emission Increased source pressure- closed ion box Potential across ion box – too high – variable ion energies – too low ion-molecule interactions. fragmentation

  13. Ion optics • About 50% efficient • Burn marks • Extraction • Half plate focussing • Fine-tuning – • Generally empirically tuned

  14. Problem areas of source design • Deviations from linear behavior – “discrimination” • ion-molecule interactions forming isobaric interferences – ie H3+ • collimating magnetic field can lead to non-linear response • changes in number of ions – affects space charge of ion-source – therefore extraction conditions • careful source design • more important in CF instruments.

  15. Problem areas of source design • 2. Gas exchange • Minimize gas exchange in ion source • Pumping efficiency of source region • Avoidance of dead volumes • Chemical inertness of source materials • Filament – chemical inertness and conditioning

  16. Problem areas in ion sources 3. High stability over time • electrostatic potentials need to be stable to 200 ppm • insulating surface layers lead to charging – source cleaning

  17. Analyser • Magnetic Sector • U=HT, B=Mag Field, z = # charge, e= charge, m=mass • m/z 44 (CO2) at 5KV = 13.5 cm, B=0.5T • Permanent/electromagnets • Magnetic field more uniform with electromagnets • Need two magnets for low mass • Image broadening by inhomogeneous ion energy • Large-radius – high energy (10kV) less affected • DE/U less in large radius instruments • Early instruments x-only focussing • All modern instruments X-Y (cross)focussing

  18. Cross focussing Stigmatic focussing Ions enter and exit magnet at an offset angle rather than 90° Fringing fields at the magnet pole gap result in y-direction focussing Mat 250 – 1977

  19. Permanent vs Electromagnets From a theoretical veiwpoint both are identical Limited mass selection with permanent magnets – 5KV for N2 (typically designed at high end of HT range) 3.2 kV for CO2 2.2 kV for SO2 (v. low HT- lower resolution) cannot scan lower than mass 28 without magnet change HD separate magnet

  20. Detectors Faraday cups Mechanically simple Named after Micheal Farrady who first theorized ions about 1830 • Error sources • Secondary electrons • backscattering Circuit where charged ions are the charge carriers in vacuum Cup gains charge that can be measured as current when discharged N/t = I/e N/t= #of ions/sec, I= current, e= elementary charge(1.6x10-19 C) 1nA = 6x109 ions/sec

  21. Electrometers Measure charge or currrent (charge/sec) Solid state – transistors Ohms law E=IR 1V = 10-9A x 109 ohms Different resistors – different currents - similar voltages Measured by ADC High amplifications – shielded

  22. Masses not evenly spaced – so cant get a collector array for more than element -compromise – triple array or moving collectors

  23. Flat-topped peaks – defining slit width

  24. Inlets

  25. McKinney (1950) – introduced change-over valve, thereby eliminating most instrumental effects allowed measurement of O2 and CO2 to 0.1 per mil • Measured d13C to precisions of about 0.1 per mil • Has essentially remained unchanged in 50 years • smallest sample limited by requirement to maintain viscous-flow conditions • Practical limit about 15-20 mbar • Cold fingers for small volumes. • Smallest sample size about 0.2 mmol • - With a few exceptions, most sample preparations are “off-line”

  26. Continuous Flow or IR monitoring inlet • GC techniques coupled to MS • No change over or dual inlet • Viscous flow in GC stream • Smaller sample sizes • Completely taken over most modern analyses • Well suited for automated analyses • Things to be aware of: • Linearity effects • small measurement times • absolute sensitivity • isotope chromatography • statistical limits on precision • large He background (HD) • Background corrections

  27. Isotope Chromatography Transport of gas through GC not only separates chemical species but also isotopic species Cannot measure instantaneous isotope ratios, but must integrate entire peaks to “count ions” Makes correct background subtraction, and peak integration algorithms essential

  28. HD measurement in He Large mass 4 (He+) tails into m/z 3 Generally reduced by modern instrument design Differential pumping Increased abundance sensitivity by increased dispersion Energy filters to homogenize minimize DU

  29. Statistical Limits to precision • IRMS is basically measuring ion currents • Ion currents have a standard deviation as a result of “shot noise” • For low ion currents - poisson distribuion • s- = 1/√N

  30. Implications for CF-IRMS • Typically small total ion numbers are measured • For example typical 10nA CO2 peak • – about 3x1011 ions for mass 44 • -about 3x109 ions for mass 45 • s= 1/√3x109 = 2x10-5 or about 0.02 per mil • Reference peak has similar precisions so that minimum statistical limit of s is about 0.05 permil or so. Minimum.

  31. Mass Resolution

  32. Abundance Sensitivity About 10-5 on modern instruments for m/z=45 =0.001 per mil Basically, how much does one mass peak overlap the other

  33. Instrument Corrections Corrections to measured d values based on instrument properties Not as important on newer instruments as manufacturing and materials have improved Should be monitored to evaluate instrument performance • Leak correction (zero enrichment) • - corrects for differences in the two viscous leaks • - crimp is adjusted so that enrichment is zero • Abundance sensitivity • - effect of one mass on the adjacent mass • - principally controlled by instrument design • - is different for each mass • - dependent on inlet pressure • Valve mixing • -mixing of reference and sample gasses in the changeover valve by cross seat leakage • - new changeovers minimize this

  34. Computation of d values Mass spectrometers measure abundance ratios or mass enrichments Need to correct for isobaric interferences to get isotope ratios

  35. Optical Methods • Two Competing technologies: • Wavelength-Scanned Cavity Ring Down Spectroscopy (WS-CRDS) Picarro Inc • Off-axis integrated cavity output spectroscopy (OA-ICOS) • Los Gatos Research HDO H2O

  36. •Absorption spectrometry is a direct measure of concentration •Very selective - C2H2 absorbs light between 1510 - 1545 nm •Fast – laser can be reproducibly swept at > 100 Hz •For a 1 meter sample containing 100 torr of 1 ppm acetylene, ΔI/I0 ~ 10-5 •Increase pathlength by (1-R)-1 ~ 10,000 times, giving several kilometers of effective path •Single-pass ΔI/I0 ~ 10-5 􀄺 Multipass ΔI/I0 ~ 10-1 (a considerable absorption)

  37. Pros - cons Pros Cons • Do one thing • Difficult to calibrate • Like IRMS – instrumental effects Do one thing really well – eg H2O No compressed gasses No moving parts Cheap Simple mechanically

  38. Wikipedia – A modern stable isotope ratio mass spectrometer

More Related