An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. EOG for REM Sleep Detection Robert Slavicek & Andrew Wassef

  2. Description of Problem • Over 12 million people suffer from sleep apnea • Americans average 6.22 hours of sleep a night, well below the recommended 7 to 8 hours • Currently sleep studies of patients must be done outside of the home and are expensive

  3. Objectives • Wanted to create a cheap, portable sleep observation system • Taking input signals from temperature and eye movement • Be able to identify when patient is in REM sleep

  4. Original Design Test Subject (Electrodes) Signal Filtering and Amplification Signal Filtering and Amplification Test Subject (Thermistor) Summer Circuit Data Acquisition & Storage C Program & Display

  5. Original Design • Obtain clear, useable signal • Have system be functional for multiple patients • Need to insure patient safety • Need to sample at a frequency high enough so that no relevant eye data is lost

  6. Overview of Information Flow

  7. Eye Dipole

  8. Biosignal: Corneal Retinal Potential • Natural mV dipole between the cornea and the retina of the eye • Front of eye is positively charged, while back is negatively charged • Measured by placing electrodes directly lateral to each eye on the canthi and a reference electrode to the forehead

  9. Eye Motion If you look right: - right electrode reads positive voltage - left electrode reads negative voltage - circuit takes difference of electrodes - net voltage is positive (pos-neg) If you look left: - exact opposite of looking right - net voltage is negative (neg-pos)

  10. Circuit

  11. Circuit Diagram

  12. Buffer Circuit: isolates the output from the signal source Differential Amplifier: Rejects signals common to both electrodes 4th Order Low Pass Filter: attenuates signals from frequencies greater than 35Hz

  13. Variable Amplifier: allows operator to control the gain for each subject Summer Circuit: allows operator to block DC offset Inverting Amplifier: increases signal to better differentiate eye movement

  14. Wein Bridge Oscillator Electronic oscillator used to generate a sine wave so that the temperature sensor is at a different frequency spectrum

  15. Temperature Sensor Implementing the temperature signal with the eye signal

  16. Data Acquisition

  17. Data Acquisition • Used a 16-bit sound card to acquire data • Data was collected through a program Audacity • Audacity collected the data into a .wav file at 8000KHz sampling • Files of a night’s sleep were about 230 MBs of storage.

  18. Data Acquisition • From Audacity, we import the .wav file into Matlab • Matlab’s “wavread()” command sorts the .wav data into a M by 2 Matrix

  19. Data Acquisition • When we sort this data into a matrix, we can perform the FFT (Fast Fourier Transform) • We isolate the frequency range of the eye signal from the temperature signal

  20. Data Acquisition • Using Matlab’s “sptool” command, we could isolate each signal for analysis • Sptool contains FIR bandpass and LS low pass filters that can filter out either signal

  21. Data Acquisition • Used LS low pass filter, order of 80 cutoff at 170 Hz

  22. Data Acquisition • After filtering the original signal we can send the time domain signal back to matlab in a data array • The data array can be exported as a .txt file to be further analyzed. Filtered

  23. C Program

  24. C Program Analysis of Data Output file from Matlab

  25. C Program Analysis of Data

  26. C Program Analysis of Data

  27. C Program Analysis of Data remsun1> eog Enter input file name: eogin.txt Enter sampling frequency (Hz): 20 Would you like to average the data (y/n)?: n The output file 'out.txt' was written successfully! Would you like to view it (y/n)?: y Left at 0.4500 s due to -0.2017 V Right at 1.9500 s due to 0.2179 V Left at 3.7500 s due to -0.2174 V Right at 5.5000 s due to 0.2285 V Left at 7.2500 s due to -0.2547 V Right at 9.0000 s due to 0.2144 V Left at 10.7000 s due to -0.2121 V Right at 12.5000 s due to 0.2120 V Left at 14.3000 s due to -0.2259 V Right at 16.0000 s due to 0.2297 V Left at 18.0500 s due to -0.2486 V Right at 18.4500 s due to 0.2181 V Left at 18.8500 s due to -0.2195 V Right at 19.3500 s due to 0.2295 V Left at 19.9000 s due to -0.2308 V

  28. Final Output

  29. Final Output File Left at 0.4500 s due to -0.2017 V Right at 1.9500 s due to 0.2179 V Left at 3.7500 s due to -0.2174 V Right at 5.5000 s due to 0.2285 V Left at 7.2500 s due to -0.2547 V Right at 9.0000 s due to 0.2144 V Left at 10.7000 s due to -0.2121 V Right at 12.5000 s due to 0.2120 V Left at 14.3000 s due to -0.2259 V Right at 16.0000 s due to 0.2297 V Left at 18.0500 s due to -0.2486 V Right at 18.4500 s due to 0.2181 V Left at 18.8500 s due to -0.2195 V Right at 19.3500 s due to 0.2295 V Left at 19.9000 s due to -0.2308 V

  30. Functional Tests Left at 0.0500 s due to -0.2066 V Right at 0.5500 s due to 0.4051 V Left at 1.2000 s due to -0.2246 V Right at 1.9500 s due to 0.4172 V

  31. Functional Tests Left at 0.0500 s due to -0.4279 V Right at 0.1500 s due to 0.1525 V Left at 0.4000 s due to -0.3228 V Right at 0.6500 s due to 0.1831 V Left at 0.8500 s due to -0.3472 V Right at 1.0500 s due to 0.1239 V Left at 1.3500 s due to -0.3410 V Right at 1.5500 s due to 0.1944 V Left at 1.8500 s due to -0.3426 V Right at 2.1000 s due to 0.2452 V Left at 2.3000 s due to -0.3210 V Right at 2.5500 s due to 0.1284 V Left at 2.8000 s due to -0.3957 V Left at 3.2500 s due to -0.3876 V Right at 3.5500 s due to 0.2024 V Left at 3.7500 s due to -0.3952 V Right at 4.0500 s due to 0.2044 V Left at 4.2500 s due to -0.4400 V Left at 4.7000 s due to -0.3433 V Right at 4.9500 s due to 0.1645 V Left at 5.1500 s due to -0.3439 V

  32. Functional Testing

  33. Functional Testing remsun1> eog Enter input file name: eogin.txt Enter sampling frequency (Hz): 20 Would you like to average the data (y/n)?: n The output file 'out.txt' was written successfully! Would you like to view it (y/n)?: y Left at 1.5500 s due to -0.2017 V Right at 1.9500 s due to 0.2179 V Left at 2.4500 s due to -0.2174 V Right at 2.8000 s due to 0.2285 V Left at 3.2500 s due to -0.2547 V Right at 3.8000 s due to 0.2144 V Left at 4.4500 s due to -0.2121 V Right at 5.1000 s due to 0.2120 V Right at 6.0500 s due to 0.2297 V Left at 7.9500 s due to -0.2486 V Right at 8.4500 s due to 0.2181 V Left at 8.8500 s due to -0.2195 V Right at 9.3500 s due to 0.2295 V Left at 9.9000 s due to -0.2308 V Right at 11.3500 s due to 0.2295 V Right at 13.0000 s due to 0.2235 V Right at 18.4500 s due to 0.2102 V

  34. Successes • Effectively captured lateral eye motion and temperature changes in human subjects in a portable device • Measured and stored changes in these signals over an entire night of sleep • Ensured patient safety • Display in a user friendly fashion when a subject displays mannerisms of REM sleep

  35. Challenges • Actually recovering a signal from the lateral eye motion • Offset of electrodes pushed signal out of viewable range and signals were lost due to saturation • Electrodes slipped off during sleep / uneasy sleep • Digitally sample the two signals in one sound card • Temperature not responsive enough to accurately gauge when REM sleep occurs

  36. Recommendations • Could add more functionality by adding EEG, EMG, HR, or BP monitors to help better determine the exact time of REM sleep

  37. Recommendations • A feedback loop in the circuit to normalize the retinal-corneal dipole signal from all users rather than having to manually adjust the potentiometer in the circuit and the threshold values in the program • A more accurate and sensitive temperature sensor

  38. Questions?