Study of Photon Sensors using the Laser System

Study of Photon Sensors using the Laser System. ACFA2005@Daegu,Korea 05/7/12 Niigata University, Japan Sayaka Iba, Editha P. Jacosalem, Hiroaki Ono, Noriko Nakajima, Hitoshi Miyata GLD Calorimeter Group. Contents. Introduction Outline of new photon sensor – MPC

Download Presentation

Study of Photon Sensors using the Laser System

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Study of Photon Sensors using the Laser System ACFA2005@Daegu,Korea 05/7/12 Niigata University, Japan Sayaka Iba, Editha P. Jacosalem, Hiroaki Ono, Noriko Nakajima, Hitoshi Miyata GLD Calorimeter Group

  2. Contents • Introduction • Outline of new photon sensor – MPC • Setup of MPC measurement using laser system • Voltage dependence of signal and noise • Laser intensity dependence of signal • Pixel signal uniformity in one pixel • Sensor signal uniformity dependent on the pixel location • Summary • Future Plan

  3. Fine segmentation scintillator Read out by photon sensor -> Study of photon sensor 1. Introduction • Present design of GLD Calorimeter • We might need smaller segmentation calorimeter • 10x40x2mm strip type scintillator X, Z-layer strip scintillator : 10x200x2mm Tile-layer : 40x40x1mm This granularity will be checked by simulation soon

  4. MPC 400pixels ~100um ~85um MPC 100pixels (10x10pixels) 2. New photon sensor --MPC-- • MPC :Multi pixels Photon Counter • Made by Hamamatsu Photonics and under development • Each pixel is in Geiger mode and as a whole pixels they worked in avalanche mode • Compact device • Works with much lower voltage than PMT • Suitable for wavelength shifter fiber • We have two types of MPC • 100pixels : 10x10pixels • 400pixels : 20x20pixels

  5. 3. Setup of MPC measurement using laser system Logic readout YAG Laser MPC YAG Laser Wave length & power: 532nm (10mJ/cm2), 1064nm (20mJ/cm2) Filter : Laser intensity is down to10-8 Trigger : from Laser system Pulse width : <10nsec Laser beam minimum spot size : <2um Precision of laser position : ±2um MPC Output circuit

  6. 0pixel (0photon) 1pixel (1photon) 2pixel (2photon) Signal • From the ADC graph, we calculated the Charge output of 1pixel which is ~ 2.5pc and Gain is ~107 for 100pixels MPC 10mV signal 500nsec trigger View from oscilloscope MPC : 100pixels From ADC signal MPC : 100pixels

  7. 4. Sensor bias dep. of 1pixel PH at 532nm 100pixels 400pixels PH PH 49.6V 48.4V 48.5V 51.0V The laser light covered all pixels (100pixels) • Below 48.6V of 100pixels, signal was low rate • Above 49.4V of 100pixels and above 50.6V of 400pixels, there were long tail of signal and exceeds the gate width (Gate full width: 1us)

  8. Noise level • Noise level = Sensor bias dependent of Pedestal sigma • We wanted to know the best operation voltage, so we used Noise level and S/N for above-mentioned condition • 100pixels : Noise level rapidly increases as bias voltage increases • 400pixels : Noise increase is more gently than that of 100pixels 30 3 400pixels 100pixels σ σ 49.6V 48.4V 48.5V 51.0V

  9. 100pixels 400pixels much tail highest region much tail flat region 49.0V 50.3V 48.5V 49.6V 48.4V 51.0V S/N • S/N = Pulse Height / Pedestal sigma • 100pixels: S/N peak was ~49.0V as the best operation voltage • 400pixels: S/N peak was above 50.5V. But it had long signal tail exceeding the gate width. So we decided to use the lower voltage (50.3V) as the best operation voltage • 400pixels operation voltage is higher than that of 100pixels and also operation voltage range is wider than that of 100pixels S/N S/N

  10. 5. Laser intensity dependence • For getting the best laser intensity corresponding to 1photon injection • Measurement conditions • MPC : 100pixels • Laser hitting area : within the 1pixel • Wavelength : 532nm • Used filter : for laser intensity down to10-8 • We think laser intensity 160 corresponds to 1photon injection, because this value is beginning of max of 1photon and min of 0photon Laser intensity 160 for 1photon injection # of event (0pe, 1pe) vs. Laser intensity ● : # of 0photon event ▲ : # of more than 1photon event

  11. 6. Pixel signal uniformity in one pixel (Position dependence between two pixels) • Checked efficiency between two pixels as uniformity measurement • Scanned 7points between two pixels • Wavelength : 532nm • Sensor bias : 49.0V • Efficiency of more than 1photon event becomes minimum at the boundary line between 2pixels Efficiency vs Position ▲: Efficiency of 0photon event ●:Efficiency of ≧1photon event Efficiency = # of 0 or 1photon event / # of All events pixel pixel

  12. ~85um ~30um ~35um Laser hitting area (smaller than 1pixel) 7. Sensor signal uniformity dependent on the pixel locations • Injected laser single photon to each pixel and got response • Measurement conditions • MPC : 100pixels • Sensor bias : 49.0V • Laser wavelength : 532nm • Laser hitting area is smaller than 1pixel area • Measured points are 50points that are shown as gray area Measured points : 50points (Gray pixels)

  13. Pulse Height 1photon mean 0photon mean • Pulse height = 1photon mean value – 0photon mean value • Deviation of PH (RMS) : 10% • Laser long term fluctuation : ≦5% Pulse height vs Pixel position # of pixels Distribution of the PH

  14. Cross sectional view (Central part) • Cross section of X-axis (line-5 + line-6) and Y-axis (line-5 + line-6) which are shown in previous slide • Central part of sensor Y-axis line-5+6 Cutting X-axis line-5▲ + line-6▲ X-axis line-5 +6 Cutting Y-axis line-5▲ + line-6▲

  15. Y-axis line-9+10 X-axis line-9 +10 Cross sectional view (Edge part) Cutting X-axis line-9▲ + line-10▲ • Cross section of X-axis (line-9 + line-10) and Y-axis (line-9 + line-10) • Edge of sensor • We can see that pulse height level and deviation look same as previous slide Cutting Y-axis line-9▲ + line-10▲

  16. Position dependence of efficiency • Compare efficiencies between 0photon events and more than 1photon events • Efficiency = #of 0photon (or ≧1photon) event / #of all photon events • Edge of sensor shows low signal efficiency • Central area of sensor has good efficiency Efficiency : 0photon Efficiency : ≧1photon

  17. 8. Summary • We measured two types of Hamamatsu MPC which are under development; 100pixels and 400pixels • Bias voltage dependence of PH, noise level and S/N were measured • Good operation voltage was 49.0V for 100pixels and 50.3V for 400pixels • Operation voltage range of 100pixels is smaller than that of 400pixels • Position dependence of PH, number of photons were measured with 532nm laser for 100pixels MPC • PH uniformity was 10% (RMS), while laser long term fluctuation was less than 5% • Central part of a sensor had good efficiency while edge part showed inefficiency in the light collection

  18. 9. Future plan • Try to connect scintillator strip and MPC through Wavelength-shifter fiber and to do beta-ray test • Measure more detail for 400pixels • Position dependence • Intensity dependence etc. • Supply our test results to Hamamatsu Photonics for further improvement MPC Setup of connection scintillator and MPC Thank you very much for listening !