An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Introduction:Optical coherence tomography Peter E. Andersen Optics and Plasma Research Department Risø National Laboratory, Denmark E-mail: peter.andersen@risoe.dk

  2. Outline • Part I: • Introduction to optical coherence tomography (OCT). • Part II: • Modeling the light propagation in the OCT geometry. • Part III: • Biomedical applications, • Extracting optical properties. Peter E. Andersen, Optics and Plasma Research Department

  3. Introduction to optical coherence tomography (OCT) Part I

  4. Outline – Part I • Physical principle behind OCT • based on interference, • tomography comes about by scanning. • Components and systems Peter E. Andersen, Optics and Plasma Research Department

  5. Optical biopsy – definition • The in situ imaging of tissue microstructure with a resolution approaching that of histology, but without the need for tissue excision and processing Peter E. Andersen, Optics and Plasma Research Department

  6. Optical coherence tomography • Three-dimensional imaging technique with ultrahigh spatial resolution even in highly scattering media • Based on measurements of the reflected light from tissue discontinuities • e.g. the epidermis-dermis junction. • Based on interferometry • involves interference between the reflected light and the reference beam. Peter E. Andersen, Optics and Plasma Research Department

  7. Resolution (log) 1 mm Ultrasound 100 mm 10 mm Confocalmicroscopy 1 mm Penetration depth (log) 1 mm 1 cm 10 cm OCT vs. standard imaging Standardclinical Highfrequency OCT Peter E. Andersen, Optics and Plasma Research Department

  8. Ophthalmology diagnosing retinal diseases. Dermatology skin diseases, early detection of skin cancers. Cardio-vascular diseases vulnerable plaque detection. Endoscopy (fiber-optic devices) gastrology, … Functional imaging Doppler OCT, spectroscopic OCT, optical properties, PS-OCT. OCT in non-invasive diagnostics • Guided surgery • delicate procedures • brain surgery, • knee surgery, • … Peter E. Andersen, Optics and Plasma Research Department

  9. The OCTsetup Fiber-optic beamsplitter Broadband source Tissue Scanning reference mirror Detector Computer Amplifier Bandpass filter Peter E. Andersen, Optics and Plasma Research Department

  10. Interference Coherent source Michelson interferometer light source Detector Partially coherent source Peter E. Andersen, Optics and Plasma Research Department

  11. Construction of image Peter E. Andersen, Optics and Plasma Research Department

  12. Normal Eye 250 microns Humphrey Nominal width of scan: 2.8 mm Peter E. Andersen, Optics and Plasma Research Department

  13. UHR-OCT versus commercial OCT W. Drexler et al., “Ultrahigh-resolution ophthalmic optical coherence tomography”, Nature Medicine 7, 502-507 (2001) Peter E. Andersen, Optics and Plasma Research Department

  14. The OCTsetup Fiber-optic beamsplitter Broadband source Tissue Scanning reference mirror Detector Computer Amplifier Bandpass filter Peter E. Andersen, Optics and Plasma Research Department

  15. System perspective (links clickable) • Light sources • Superluminescent diodes • Semiconductor amplifiers • Femtosecond lasers • … • Beam delivery and probes • Hand-held probe • Catheter • Ophthalmoscope • Microscope • OCT imaging engine • Resolution • Reference delay scanning • Doppler/polarization/spectroscopy • Detection • Frequency domain continue • Computer control • Drive system • Real-time display • Data management • Image & signal processing • Motion reduction • Speckle reduction • Image enhancement • Rendering algorithms • … Peter E. Andersen, Optics and Plasma Research Department

  16. Choosing the light source • Four primary considerations • wavelength, • bandwidth, • power (in a single-transverse-mode), • stability; • portability, ease-of-use, etc. Peter E. Andersen, Optics and Plasma Research Department

  17. Choose light source – wavelength Peter E. Andersen, Optics and Plasma Research Department

  18. Choose light source – wavelength • Light propagation (Monte Carlo simulation) Absorption “Snake” component Incident light Ballistic component Diffuse reflectance Diffuse transmittance Peter E. Andersen, Optics and Plasma Research Department

  19. Choose light source – wavelength Peter E. Andersen, Optics and Plasma Research Department

  20. Raw signals • Scattering mainly attenuates the signal Peter E. Andersen, Optics and Plasma Research Department

  21. Light source spectrum • Basic property • the temporal coherence envelope function G(t) is related to the power spectral function S(n) through G(t) = FT{S(n)} • Wiener-Kinchine theorem • broadband source  high axial resolution Peter E. Andersen, Optics and Plasma Research Department

  22. Source spectrum and envelope Peter E. Andersen, Optics and Plasma Research Department

  23. Axial resolution • The axial resolution is • notice that Dl is the 3dB-bandwidth! Peter E. Andersen, Optics and Plasma Research Department

  24. Plotting the relation Peter E. Andersen, Optics and Plasma Research Department

  25. Light sources for OCT (1/2) • Continuous sources • SLD/LED/superfluorescent fibers, • center wavelength; • 800 nm (SLD), • 1300 nm (SLD, LED), • 1550 nm, (LED, fiber), • power: 1 to 10 mW (c.w.) is sufficient, • coherence length; • 10 to 15 m (typically), • Example • 25 nm bandwidth @ 800 nm 12 m coherence length (in air). Peter E. Andersen, Optics and Plasma Research Department

  26. Light sources for OCT (2/2) • Pulsed lasers • mode-locked Ti:Al2O3 (800 nm), • 3 micron axial resolution (or less). • Scanning sources • tune narrow-width wavelength over entire spectrum, • resolution similar to other sources, • advantage that reference arm is not scanned, • advantage that fast scanning is feasible. Peter E. Andersen, Optics and Plasma Research Department

  27. OCT spatial resolution • Axial and lateral resolutions are decoupled • The axial resolution is • notice that Dl is the 3dB-bandwidth! • The lateral resolution is determined by the focusing conditions • optics • dynamic vs. static focusing Peter E. Andersen, Optics and Plasma Research Department

  28. Plotting the relation Peter E. Andersen, Optics and Plasma Research Department

  29. Low NA Lateral resolution High NA Dz Dz Dz Dz b b Depth of focus Lateral resolution (1/2) Peter E. Andersen, Optics and Plasma Research Department

  30. Lateral resolution (2/2) Peter E. Andersen, Optics and Plasma Research Department

  31. Ultra-high resolution OCT • Broad bandwidth sources • solid-state lasers, • sub-5 fs pulse; • Ti:Al2O3 (Spectral bandwidth: 350 nm demonstrated), • other lasers/wavelengths available or needed. • Special interferometers and fiber optics • support for broad spectral range, • dispersion balanced, • current system used for OCT: 260 nm bandwidth, ~1.5µm resolution. • Chromatically corrected optics • aberrations can decrease resolution and SNR. • Broad bandwidth detectors and electronics • dual balance detection, • low noise circuitry necessary. Peter E. Andersen, Optics and Plasma Research Department

  32. Scanning devices • Piezo or motorized scanning devices • ideal for both longitudinal and lateral scanning. • Galvanic mirrors • Resonance scanners • Helical mirrors • longitudinal scanning. • Fiber stretcher • longitudinal scanning. Peter E. Andersen, Optics and Plasma Research Department

  33. Fourier domain rapid scanning optical delay line (RSOD) • The technique was originally developed for femtosecond pulse measurements • based on Fourier-transform pulse shaping techniques. • Relies on the basic property of the Fourier transform • phase ramp in the Fourier domain corresponds to a group delay in the time domain. Peter E. Andersen, Optics and Plasma Research Department

  34. RSOD setup Peter E. Andersen, Optics and Plasma Research Department

  35. RSOD in the lab Peter E. Andersen, Optics and Plasma Research Department

  36. Free-space group pathlength p: the grating pitch, f: focal length, s: mirror angle. Interferogram central frequency Bandwidth RSOD characteristics Peter E. Andersen, Optics and Plasma Research Department

  37. RSOD – figures-of-merit • Scanning capabilities (galvo) •  200 Hz, •  5 mm. • Scanning capabilities (resonant) •  4-8 kHz, •  5 mm. • Advantages • dispersion compensation feasible since phase and group delays may be separated, • center frequency of interferogram is adjusted through axial position. Peter E. Andersen, Optics and Plasma Research Department

  38. Light propagation in sample • The sample • need to describe light-tissue interaction taking temporal and spatial coherence properties into account; • can transport theory be used? • Light-tissue interaction to be modelled using the extended Huygens-Fresnel principle • Part II of this lecture, • correlation between tissue, state of tissue (lesions) and optical properties? • Part III of this lecture. Peter E. Andersen, Optics and Plasma Research Department

  39. Our setup (mobile OCT unit) Peter E. Andersen, Optics and Plasma Research Department

  40. Clinically adapted systems Peter E. Andersen, Optics and Plasma Research Department

  41. BCC II • Layers • Thinning oflayers L. K. Jensen, MSc thesis (in Danish), 2003 [data obtained atLund Medical Laser Centre, courtesy K. Svanberg]. Peter E. Andersen, Optics and Plasma Research Department

  42. Dynamic range 100 dB (or better). Resolution (typical) 1-10 micrometers. Penetration depth depending on wavelength/tissue, 1-2 mm (typically) for 1300 nm in skin tissue. Axial and lateral resolutions are decoupled important for applications. Pixel density is related to spatial resolution and image acquisition time Nz=2*Lz/dz, Nx=2*Lx/dx, image acq. time: T=Nx*fs, scan velocity: vs=Lz*fs. Image acquisition seconds or less, real-time OCT. Clinical adaptation interfaced to standard equipment, fiber-optic devices, endoscopes. OCT: Figures-of-merit – summary Peter E. Andersen, Optics and Plasma Research Department

  43. Ophthalmology diagnosing retinal diseases. Dermatology skin diseases, early detection of skin cancers. Cardio-vascular diseases vulnerable plaque detection. Endoscopy (fiber-optic devices) gastrology, … Functional imaging Doppler OCT, spectroscopic OCT, optical properties, PS-OCT. OCT in non-invasive diagnostics • Guided surgery • delicate procedures • brain surgery, • knee surgery, • … Peter E. Andersen, Optics and Plasma Research Department

  44. OCT: suggested reading • OCT • D. Huang et al., Science 254, 1178 (1991) • A. Fercher, J. Biomed. Optics 1, 157 (1996) • J. M. Schmitt, “Optical coherence tomography (OCT): A review”, IEEE J. Select. Topics Quantum Electron. 5, 1205-1215 (1999) • J. G. Fujimoto et al., “Optical coherence tomography: An emerging technology for biomedical imaging and optical biopsy”, Neoplasia 2, 9-25 (2000) • W. Drexler et al., “Ultrahigh-resolution ophthalmic optical coherence tomography”, Nature Medicine 7, 502-507 (2001) • W. Drexler, “Ultrahigh-resolution optical coherence tomography”, J. Biomed. Opt. 9, 47-74 (2004) Peter E. Andersen, Optics and Plasma Research Department