480 likes | 772 Views
RECTIFICATION. X-RAY CIRCUIT. Whenever the art of medicine is loved, there is also love for humanity. Hippocrates. RECTIFICATION. PROCESS BY WHICH ALTERNATING CURRENT IS CHANGED TO PULSATING DIRECT CURRENT. ALTERNATING CURRENT 1 PHASE. A/C ( UNRECTIFIED)- FREQUENCY 60 Hz.
E N D
Whenever the art of medicine is loved, there is also love for humanity. Hippocrates
RECTIFICATION PROCESS BY WHICH ALTERNATING CURRENT IS CHANGED TO PULSATING DIRECT CURRENT
ALTERNATING CURRENT1 PHASE A/C ( UNRECTIFIED)- FREQUENCY 60 Hz
60 POSITIVE PULSES/SEC 60 NEGATIVE PULSES/SEC
60 POSITIVE PULSES 60 NEGATIVE PULSES 120 PULSES 60 FULL CYCLES
BATTERY – CURRENT OUTPUT STEADY D/C
D/C HALF WAVE RECTIFIED D/C HALF-WAVE – 60 POSITIVE PULSES/SEC NEGATIVE PULSES ARE SUPPRESSED 1 PULSE/CYCLE
D/C FULL WAVE RECTIFIED D/C HALF-WAVE – 120 POSITIVE PULSES/SEC 2 PULSES/CYCLE
D/C -3 PHASE 6 PULSE CURRENT D/C 3 PHASE 6 PULSE – 6 PULSES/CYCLE
D/C - 3 PHASE 12 PULSE D/C 3 PHASE 12 PULSE – 12 PULSES/CYCLE
DIODES OR VALVE TUBES HELP RECTIFY A/C CURRENT ELECTRONS DIODE VALVE TUBE
REVERSE BIAS + - NO X-RAY
RECTIFICATION TYPES • HALF-WAVE • FULL WAVE • THREE PHASE SIX PULSE • THREE PHASE TWELVE PULSE • HIGH FREQUENCY SINGLE PHASE
RECTIFICATION NO RECTIFICATION HALF-WAVE FULL-WAVE
VOLTAGE RIPPLE • HALF WAVE------------------- 100% • FULL WAVE ------------------ 100% • 3 PHASE 6 PULSE ----------- 13% • 3 PHASE 12 PULSE ---------- 4% • HIGH FREQUENCY ---------- <1%
RIPPLE 100% 100% 100%
RIPPLE-3 PHASE 13%
CHECKING FAILURE OF RECTIFICATION SYSTEM • SPINNING TOP TEST – 1 PHASE EQUIPMENT • SYNCHRONOUS SPINNING TOP TEST- 3 PHASE WITH LONGER EXPOSURE TIMES • OSCILLOSCOPE- 3 PHASE EQUIPMENT WITH SHORT EXPOSURE TIMES
HALF- WAVE RECTIFICATION # OF DASHES = TIME (SEC) x 60 PULSES
FULL- WAVE RECTIFICATION # OF DASHES = TIME (SEC) x 120 PULSES
3 PHASE RECTIFICATION-LONGER EXPOSURE TIME DEG. OF ARC = 360°/ sec x TIME (sec)