UNIT 2 X-ray Tube Review & Dedicated X-ray machines BUSHONG- Ch. 6 & 7 Merrills’s V. 3 Ch. 32 RT 244 Week 10
Major Topics • Review the components of the x-ray tube • Saturation current • Line focus principal/Anode heel affect • Anode cooling/HU • Off focus radiation • Dedicated x-ray machines
A B D F C G E X-ray Tube Construction WHAT TYPE OF MOTOR IS ROTOR? The energy of electrons comprising the tube current is measured in ______________ ?
What makes the Anode spin? • Electromagnetic Induction Motor • How fast does it turn? • ______________ rpm
A motor is a device that converts • electrical energy into mechanical energy
Cathode - Filaments • Negative Electrode • Tungsten with thorium • Filament circuit heats (>2200°C)with 3-5 amps • Thermionic Emission • Vaporization • Space charge • Saturation current • Focusing cup
Molybdenum or Graphite base Anodes - Target • Rotating Anodes • 2” to 5” disk (focal track) • Induction motor • Speed 3000 to 10000 rpm Cu W • Other target materials • Tungsten-rhenium • Molybdenum • Rhodium
Angle (5°-15°) e- effective actual Line-Focus Principle • Effective FS < actual FS • Beveled anode • Improves heat capacity • Small angle = FS • Small FS = resolution • AHE • Field coverage • Heat loading on smaller anode surface area
Anode Angles and Heat • Small the angle = more heat load • Larger angle – less heat load • Different anode angles are used for different types of equipment: • Diagnostic vs special procedures
ANODE HEEL EFFECT • MORE PRONOUNCED WITH SMALLER ANODE ANGLE • (sm anode angle = larger heel effect) • 120% ~75% difference of intensity across beam • How does this affect positioning? FAT – CAT HIGHER AT CATHODE See pg 139 Bushong
Leakage Radiation <100 mR/hr @ 1 m Extra-focal Radiation rebounding e- Off Focus Radiation Protective Housing Miscellaneous Terms
SHADOW OF SOMEONE’S HEAD = OFF FOCUS FROM TUBE
LEAKAGE RADIATIONmay not EXCEED • TUBE HOUSING 100mR / HR @ 1 meter
Tube Failures & Prevention • Failure Causes • Anode pitting/cracking after single excessive exposure • Bearing damage from numerous long exposures • Vaporization of the filament • Filament break • Coating of glass envelope with tungsten • Prevention • Warm up tube before use • Avoid “boost & hold” exposures if possible • Use acceptable levels of exposure (e.g. tube rating charts)
Heat Units • HU – measure of thermal energy applied to the x-ray tube from an exposure • Formula based on generator HU1Ø = kVp x mAs x 1 HU3Ø6p = kVp x mAs x 1.35 HU3Ø12p = kVp x mAs x 1.41 • Multiply if consecutive exposures are made, HUT = HU x # exposures
TYPES OF EQUIPMENT • FIXED • MOBILE • DEDICATED • Bone densitometry, Chest, Head, Mammography, Panorex & Tomography
1 second exposure time Need to adjust mA How do you adjust the fulcrum?
Which terms do not apply ? • Linear • Curvilinear • Hypocycloidal • Trisprial • Oscillating • Circular • Figure 8 • Reciprocating • Elipictal
Conventional tomo vs. CT images Questions?