What difference does a difference make? Elizabeth Little, Ph.D. 26-Oct- 2010
Talk overview • Introduction • Tissue thickness variation • Using best histological practices • Stain intensity variation due to tissue thickness • The difference matters • Could impact algorithm functionality
Systems integration source: www.vagabondish.com
The Hematoxylin & Eosin (H&E) slide • Numbers • In 2009, 330 million histology slides were produced in the United States • 83% (274 million) were stained with H&E • Pathologist • Potential first look at the disease state • Cost • Dollars vs. thousands of dollars for more advanced testing
Impacts of H&E stain variability • Pathologist workflow is impacted by staining variability • Repeat slides • Imaging workflow is also impacted by staining variability • Algorithms can by impacted by stain variability
Antecedents that are helpful for H&E slide image analysis • Control of the stain variation • Under best practices we can control stain variability to a certain degree • Algorithms that are robust against stain variation
Staining variables we cannot control- tissue type affects stain intensity
Pixel count (N) Intensity Level
Staining variables that we have some control over - tissue thickness impacts stain intensity 2 micron 4 micron
Pixel count (N)
Talk overview • Introduction • Tissue thickness variation • Using best histological practices • Stain intensity variation due to tissue thickness • The difference matters • Could impact algorithm functionality
Possible sources of variations in section thickness in the histology laboratory • Fixative • Duration of fixation • Tissue processing • Paraffin • Tissue block • Microtome • Histologist
Objective – measure the sectioning process impact on tissue thickness • 1 tissue block used • 1 microtome • 2 settings • Automated (32 slides per histologist) • Manual (32 slides per histologist) • 2 histologists • 22 years of experience vs. 4 years of experience
Tissue thickness variability testing outline • Section • Tissue was sectioned using a microtome setting of 4 microns • Measure Section Thickness • Interferometry • Stain • H&E • Measure intensity • Whole slide imaging
Measuring tissue thickness using vertical scanning interferometry source: cnx.org
Tissue thickness using interferometric measurements • Glass vs. paraffin • Tissue was not measured • Interferometer limitation • Glass level variability • Measurements taken at 6 locations repeatedly
How good is our tissue thickness measuring system? - gage R & R Equipment variation – 0.79% Operator variation – 0.01% Sample variation – 99.20%
Slice thickness variation – by histologist • Nominal setting was 4 microns • Both Histologists cut significantly thicker than 4 microns • Both Histologists cut at significantly different thicknesses from each other
Manual vs. automated microtomy impact on tissue thickness • Histologist 1 mean thickness was not impacted by microtome setting • Both histologists had statistically significant more variability using the • automated setting as compared to the manual setting
Block influences tissue thickness • Histologist 1 was the cutter • Automated setting used • Tissue 3 was cut significantly thinner than tissues 1 & 2
Summary of tissue thickness measurement results • Histology (location within block, slice selection, soaking, etc.) • Difference in mean tissue thickness • Microtome setting – automated vs. manual • Both histologists were impacted by setting • Block • Blocks 1 and 2 were cut more thickly than block 3
Talk overview • Introduction • Tissue thickness variation • Using best histological practices • Stain intensity variation due to tissue thickness • The difference matters • Could impact algorithm functionality
Stain intensity variation due to tissue thickness - normal breast lymph node study 3 micron 4 micron
Objective – measure tissue thickness impacton stain intensity • Tissue was sectioned and measured for thickness • All slides were stained using the same method • All slides were scanned using whole slide imaging and their average intensities were measured
Talk overview • Introduction • Tissue thickness variation • Using best histological practices • Stain intensity variation due to tissue thickness • The difference matters • Could impact algorithm functionality
Grey scale intensity differences Pixel count (N)
Summary • Expected vs. measured is different • The difference is quantifiable • Tissue thickness • Stain intensity • The difference matters • Could impact algorithm functionality • Tissue thickness and stain intensity correlate as expected
Further studies • Intensity vs. tissue type • Microtome bounce • Histology vs. • Drift • Knife • Location in block • Degrees of fixation
Acknowledgments Cindy Connolly Wendy Lange Allison Cicchini Heather Free Aaron Ewoniuk Jonathan Hall Mike Cohen, Ph.D. David Clark, Ph.D.