1 / 14

Goal

Physics Testing for Performance Based Protocol version 3.0 (Now based on ACRIN 6678) QIBA Group 1C. Goal. From a specification of image quality performance, determine scanner settings which achieve that performance. Statement of Image Quality Performance.

petej
Download Presentation

Goal

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Physics Testing for Performance Based Protocolversion 3.0(Now based on ACRIN 6678) QIBA Group 1C

  2. Goal From a specification of image quality performance, determine scanner settings which achieve that performance

  3. Statement of Image Quality Performance • QIBA 1C Image Quality Performance Specification: • 120 kVp • Reconstructed slice thickness between 0.75 to 1.25 mm • Beam Collimation (Detector configuration), Pitch and rotation time set to that scan through an entire thorax could be completed in under 15 seconds* • Spatial Resolution of 6 lp/cm on ACR phantom spatial resolution section in ACR Phantom Module 4 • Standard Deviation (proxy for noise) of 17±1 HU in ACR Phantom water equivalent Uniformity Module- Module 3 * Example of a breathold constraint; technically not an image quality performance specification, but related.

  4. General Outline of Testing Protocol • Record required data on accompanying spreadsheet, • Note that spreadsheet has two tabs to allow two Scan Field of View (or Patient Size) settings to be used if necessary. • If Scanner has multiple Patient size (Toshiba) or SFOV (GE) settings, then perform measurements twice, once each for: • SFOV or Patient Size that is appropriate for size of phantom (20 cm diameter) AND • SFOV or Patient Size that would be appropriate for adult thorax (approx. 35 cm diameter) • Set Scan Field of View or Patient Size (or equivalent). • Set 120 kVp • Set Slice thickness between 0.75 to 1.25 mm (depending on scanner’s available reconstructed slice thickness) • Set nominal beam collimation (NxT such as 16 x 0.5mm, or 128x0.6mm, 320 x 0.5 mm) rotation time and pitch such that scan can cover a 35 cm thorax in 15 seconds or less • Values near pitch 1 are preferred. • ITERATE (hopefully only a few times) on reconstruction kernels to meet spatial resolution spec. • ITERATE (again, hopefully just a few times) on mAs or effective mAs setting, given beam collimation, pitch and rotation time.

  5. Specific Instructions for Testing Protocol • On accompanying spreadsheet, record: • Scanner Manufacturer (e.g GE) • Model (e.g. VCT) • Site id (e.g Scanner 95) • Software version • Testing date (12-26-09) • Set kVp to 120 or just lower if there is no 120 kVp setting • Record selected setting on accompanying spreadsheet • If Scanner has multiple Patient size (Toshiba) or SFOV (GE) settings, then perform measurements twice, once each for: • SFOV or Patient Size appropriate for phantom size (20 cm diameter) AND • SFOV or Patient Size appropriate for adult thorax (approx. 35 cm diam) • Record selected setting on accompanying spreadsheet • Use different tabs for each set of measurements • Set reconstructed slice thickness between 0.75 to 1.25 mm (depending on scanner’s available reconstructed slice thickness) • Record selected setting on accompanying spreadsheet

  6. Specific Instructions for Testing Protocol • Set the following parameters such that the scan could cover a 35 cm thorax in ≤ 15s AND that reconstructed slice thickness selected in step 4 is possible. The accompanying spreadsheet calculates scan time for 35 cm long thorax given selected parameters Select: • Nominal beam collimation (NxT such as 16 x 0.5mm, or 64x0.625mm, 320 x 0.5 mm, etc.) • Rotation time (in seconds) • Pitch (Values near pitch 1 are preferred)

  7. Specific Instructions for Testing Protocol • ITERATE on recon kernels to get 6 lp/cm in ACR phantom 6.1 Scan phantom once through section 4 of ACR CT accreditation phantom, using: • kVp selected in step 2 • Patient size or SFOV selected in step 3 • Reconstructed slice thickness selected in step 4 (use contiguous reconstructions) • Other acquisition parameters (nominal beam collimation, rotation time, pitch) as selected in step 5. • Select mA, mAs/rotation or effective mAs – for this step, use high enough mA to get 200 effective mAs; spreadsheet has calculator to confirm 200 effective mAs (even if you specify mA). • Record mA or mAs or eff. mAs setting on spreadsheet. • Instructions continued on next page

  8. Specific Instructions for Testing Protocol 6.2Reconstruct images using several recon kernels, display with Level =1100, Window=100 • For each recon filter used: • Record the name of recon filter • Record the size of the line pair object (e.g. 6 lp/cm etc.) observed for that filter • Select smoothest kernel that allows user to see 6 lp/cm: • Record the name of recon filter selected • Provideimage (DICOM AND jpeg) reconstructed with this filter Zoomed in Zoomed in B30 B45 6 lp/cm 6 lp/cm 7 lp/cm 7 lp/cm ACR Spatial Resolution Section (Module 4)

  9. Specific Instructions for Testing Protocol 7 ITERATE on mA or eff. mAs settings to meet Standard Deviation Specification (17±1 HU) 7.1 Scan and reconstruct images of section 3 (homogeneity) of ACR CT accreditation phantom, using: • kVp selected in step 2 • Patient size or SFOV selected in step 3 • Reconstructed slice thickness selected in step 4 (use contiguous reconstructions) • Other acquisition parameters (nominal beam collimation, rotation time, pitch) as selected in step 5. • Reconstruction filter as selected in Step 6 • Select an initial mA (or mAs or effective mAs) setting • Record initial mA or eff. mAs setting on spreadsheet • From Reconstructed image, measure standard deviation by placing an ROI of approximately 400 mm2 at the center of the image • Record standard deviation for this setting on spreadsheet

  10. Specific Instructions for Testing Protocol 7.2 If standard deviation is within specification (17±1 HU), then: • Record mA (or eff. mAs setting) • Record scanner reported CTDIvol • Provideimage (DICOM AND jpeg) using this setting. 7.3 If standard deviation is NOT within specification (17±1 HU), then: • Adjust mA (or eff. mAs) setting and repeat steps 7.1.f and 7.1.g until standard deviation is within specification. • When specification is met, then perform step 7.2

  11. Example of Testing Protocol to Determine Parameters That Meet Specification • Step 7 – Example on Siemens Sensation 64. • 100 eff. mAs • Tech. Params: • 120 kVp • 1mm • Pitch 1 • 0.5 sec • B30 Spec: std dev= 17±1 HU MEETS SPEC! 100 eff. mAs; std dev = 17

  12. Testing Protocol to Determine Parameters That Meet Specification • Conclusion: For Siemens Sensation 64, these settings meet QIBA 1C specifications: • 120 kVp • Reconstructed Slice thickness 1.0 mm • Nominal beam collimation 64x0.6 (actually 32x0.6 with flying focal spot) • Rotation time 0.5 seconds • Pitch 1.0 • Recon kernel B30 • Effective mAs setting 100 • CTDIvol = 7.7 mGy

  13. For Scanners with Multiple Patient Size (Toshiba) or SFOV (GE) Settings • Repeat Steps 1-7 for 2nd patient size setting • Record results on second tab of spreadsheet

  14. Next Scanner?

More Related