Radiation damage of 1600Pixel MPPC by neutrons

Radiation damage of 1600Pixel MPPC by neutrons. ILC-CAL meeting University of tsukuba H.Yamazaki. Dose rate. Flux = 1.5×10 9 neutron/cm 2 /Wh (W = watt, h = hour) …… calculated MCNP ( Monte Carlo Neutron and Photon Transport Code System ) Code. 4 dose rate patterns

psyche
Download Presentation

Radiation damage of 1600Pixel MPPC by neutrons

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Radiation damage of 1600Pixel MPPC by neutrons ILC-CAL meeting University of tsukuba H.Yamazaki

  2. Dose rate Flux = 1.5×109 neutron/cm2/Wh (W = watt, h = hour) …… calculated MCNP(Monte Carlo Neutron and Photon Transport Code System) Code 4 dose rate patterns                      ↓ 1.5×109 neutron/cm2 /Wh×12.5/60 h×1W =3.125×108 neutron/cm2 1.5×109 neutron/cm2 /Wh×12.5/60 h×10W =3.125×109 neutron/cm2 1.5×109 neutron/cm2 /Wh×12.5/60 h×100W =3.125×1010 neutron/cm2 1.5×109 neutron/cm2 /Wh×12.5/60 h×1kW=3.125×1011 neutron/cm2

  3. Uniformity of Sample (Noise Rate,Cross-talk Probability ) Noise Rate (Vbias) Cross-talk Probability Noise Rate (Vover ) • Sample irradiated with 3.125×108n/cm2 • Sample irradiated with 3.125×109n/cm2 • Sample irradiated with 3.125×1010 n/cm2 • Sample irradiated with 3.125×1011 n/cm2 The individual difference of several percent was confirmed to four samples.

  4. Measurement after the irradiation • Before the irradiation • 3.125×109 n/cm2 • 3.125×109 n/cm2 • Before the irradiation • 3.125×108 n/cm2 • 3.125×109 n/cm2 Noise Rate Cross-talk Probability (at Vbias = -77.5V) before 108 n/cm2 109 n/cm2 NoiseRate(Hz) 346,236 569,263 3,157,690 Increased of Noise Rate 160% 910% Crosstalk probability 0.158 0.144 0.171 Noise Rate rises as dose rate goes up, Cross-talk dose not change so much by irradiation

  5. Measurement after the irradiation (3.125×108 n/cm2) several times measurement after irradiation Noise Rate Cross-talk Probability • 2007 11/14 Before irradiation • 11/16 Just after irradiation • 12/05 Before annealing at 60゜C • 12/06 After annealing at 60゜C • 12/15 The additional measurement 1 • 12/15 The additional measurement 2

  6. Measurement after the irradiation( 3.125×109 n/cm2) several times measurement after irradiation Noise Rate Cross-talk Probabirity • 2007 11/14 Before irradiation • 11/16 Just after irradiation • 12/05 Before annealing at 60゜C • 12/06 After annealing at 60゜C • 12/15 The additional measurement 1 • 12/15 The additional measurement 2

  7. Summary • Noise Rate rises as dose rate gose up • Cross-talk dose not change so much by irradiation • There are some problems for the noise rate measurement • Plan • Another analysis …….Gain,NoiseRate(Vbias correction),After Pulse

  8. Back Up

  9. Noise Rate (Reference)