H2 upstream XTAX grid connection broken on 05/6/2012

H2 upstream XTAX grid connection broken on 05/6/2012. Pictures here : \cern.chdfsWorkspacessSBANORTHTCC2PhotosH2upstream XTAX grid connection broken on 20120605. How it should be. Grid down. Broken rotule. Broken part . Environment detail. Mechanical analysis : adjustment.

regis
Download Presentation

H2 upstream XTAX grid connection broken on 05/6/2012

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. H2 upstream XTAX gridconnectionbroken on 05/6/2012 Pictureshere: \\cern.ch\dfs\Workspaces\s\SBA\NORTH\TCC2\Photos\H2upstream XTAX grid connection broken on 20120605

  2. How itshouldbe

  3. Grid down

  4. Broken rotule

  5. Broken part

  6. Environmentdetail

  7. Mechanicalanalysis: adjustment • X towards EHN1, Y towards Jura, Z Up Z X Y

  8. Mechanicalanalysis: degree of freedom • X towards EHN1, Y towards Jura, Z Up Z X Y • 2 degrees of freedom are missing in the system

  9. Rotule data sheet

  10. Mechanicalanalysis: rupture load • M8 rupture load (shear mode): σr= 240 Mpa, As= 38 mm2 10kN • Stress in function of beam offset: f=159.PL3/I • (L = 3 m, P = 10 kN) • Hard to believethisfailure mode • Failure of a swith on the gridshouldbeinvestigated (rupture by traction then: tests on a sparegrid)

  11. First idea for new design

  12. Check in TCC2 • Laser scan: check verticality of the beams • If not possible, survey of beam motion on full range (sticker) check verticality • Remove broken part + analysis (shear or traction rupture) • Check offset, install new connection and adjust via 4 holes • Check grid