1 / 41

Method Adjustment Analyses for VA DEQ Nutrient Determinations

Method Adjustment Analyses for VA DEQ Nutrient Determinations. Introduction. 1984 until 1994 TN calculated as TKNW + NO 23 F(NO 2 F + NO 3 F) TP measured directly (analyzed by DCLS). January 1994 TN calculated as TDN + PN (analyzed by VIMS).

reidar
Download Presentation

Method Adjustment Analyses for VA DEQ Nutrient Determinations

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Method Adjustment Analyses forVA DEQ Nutrient Determinations

  2. Introduction • 1984 until 1994 TN calculated as TKNW + NO23F(NO2F + NO3F) TP measured directly (analyzed by DCLS). • January 1994 TN calculated as TDN + PN (analyzed by VIMS). TP calculated as TDP + PP (analyzed by VIMS). • February 1995 to the present DCLS adopts VIMS methods for PC, PN, PP, and TDN DCLS uses EPA method 365.2 for TDP. • Changes resulted in step trends in both parameters. • Adversely affects statistical analyses.

  3. Pre-Method Change Post-Method Change Example – TN at Station LE5.4

  4. Methods • 33 Tidal Monitoring Stations (1998-2003). • 46 Pfiesteria Monitoring stations (1998-2002). • Focus on Tidal Monitoring data only but Pfiesteria Monitoring data also analyzed. • For CBP TP, there was an average of 33 with a min. of 11 and max. of 42 samples per station. • For CBP TN, there was an average of 28 samples per station with a min. of 10 and max. of 75 samples per station. • Sample collection and processing reflect historical methods used except: change in instrumentation.

  5. Definition of Bias • New Method – Old Method = Difference Between Methods. • A negative value indicates Old Method is biased high relative to the New Method i.e. consistent with historical bias. • A positive value indicates Old Method is biased low relative to the New Method in contrast with the historical bias.

  6. Total Nitrogen

  7. TN - Paired Comparisons • Difference Between Methods significantly different from zero: Student’s t= -26.66; Prob. >| t | <0.01, Wilcoxon Signed Rank S=-500812; Pr >|S| <0.01.

  8. TN - Screening Analyses + 2 Std Mean Difference • Mean difference between methods: -0.320.52. • 75% of all observations at or below 0.0 mg/L. • Old Method biased high. • There were two distinct groups of values for the differences. - 2 Std

  9. TN – Screening Analyses • First group of differences Mean: 0.00 mg/L Range: –0.50 to 0.80 mg/L Range constant regardless of mean conc. • Second group of differences Mean: -1.00 mg/L Range: -0.50 to –2.00 Difference decreases with increasing mean conc.

  10. TN - Screening Analyses • Frequency histogram confirms two groups in bias.

  11. TN – Source of Groups? • Procedural Effects: • Two Monitoring Programs • Three DEQ Offices (TRO, PRO, NRO) • Spatial Effects • Temporal Effects • Environmental Effects • Combinations

  12. Mean diff. = -0.22 Positive Bias CBP Pfiesteria TN – Monitoring Program Effect Mean diff. = -0.41 Positive Bias

  13. Mean diff. =-0.480.45 Old Method Biased High Mean diff. =0.080.18 Old Method Biased Low TN – Collection Office Effect (CBP) Mean diff. =0.030.33 Old Method Biased Low • Was collection agency (PRO, TRO, NRO) responsible? • Grouping persists in TRO data with Old Method biased high. • Grouping disappears in PRO data but now Old Method biased low. • Grouping disappears in NRO data but now Old Method biased low.

  14. TN – Spatial Effects on Bias (CBP) • Negative bias at most Tidal Fresh and Oligohaline stations. • PRO and NRO responsible for collection at most of these stations. • Positive bias at higher salinity (mostly TRO) stations.

  15. TN – Spatial and Temporal Effects (CBP) • Spatial pattern persists between years. • Difference between methods higher in 2002 ( 0.00 mg/l).

  16. TN – Temporal Effects (CBP - TRO) Mean diff. =-0.130.30 Old Method Biased High • Two groups observed appear to be two different time periods. • Old Method is biased high for both time periods. Mean diff. =-0.650.43 Old Method Biased High

  17. TN – Environmental Effects (CBP - TRO) Shown are Pearson’s |R| and associated p values. All correlations based on > 400 observations except CHL a (221). • Several significant correlations but none explain patterns observed. • Other potential predictors?

  18. TN – Environmental Effects (CBP - TRO) • Prior to 2003, difference was much higher even in high salinities. • For 2003, values of difference at 0 salinities were higher. • Significant but slight correlation with salinity (|R|=-0.12;p<0.01).

  19. TN – Component Variables (CBP – TRO)

  20. TN – Component Variables (CBP - TRO) • Plots of component variables indicate TKNW as problem. • Plot of bias confirms this observation. • Correlation analysis revealed no apparent environmental causes.

  21. TN – Component Variables (All Data)

  22. TN – Component Variables (CBP - TRO) • When TKNW < 0.8 then Old Method bias is negative. • When TKNW >=0.8 then Old Method bias is positive.

  23. |R| = NS |R| = 0.25 |R| = 0.29 |R| = 0.08 TKNW – Environmental Effects (All Data)

  24. TN - Conclusions • Overall significant difference between methods with Old Method biased high relative to the New Method. • Two groups of differences in the data set. • Two groups were not due to differences between monitoring programs or DEQ Offices. • Old Method was biased low at most Tidal Fresh and Oligohaline stations and at low concentrations of TKNW. • Old Method was biased high when concentrations of TKNW were high. • No clear effect of environmental factors on results.

  25. TN - Conclusions • Method correction development is not recommended using these data • Bias in Old Method low at Tidal Fresh/Oligohaline areas • Bias is inconsistent in all other areas • Bias changes in relation to TKNW • No apparent environmental predictor of this change • Changes in instrumentation are still at issue.

  26. TN - Recommendations • Use Blocked Seasonal Kendall until questions are answered. • No other recommendations… • Are there additional analytical approaches that might be useful for exploring these data? • Are there any explanations for the patterns observed?

  27. Total Phosphorus

  28. TP - Paired Comparisons (CBP Only) • Mean difference between methods significantly different from zero: Student’s t=8.00; Prob. >| t | <0.01 Wilcoxon Signed Rank S value =219432; Prob. >|S| <0.01 • Mean difference between methods = 0.020.035 mg/L. • 75% of all differences at or above 0.00.

  29. + 2 Std Mean Difference - 2 Std TP - Screening Analyses (CBP Only) • Mean difference between methods: 0.020.035 mg/L. • Old TP Method biased low relative to New TP Method up to mg/L. • This conflicts with historical pattern in the data.

  30. TP - Conclusions • Significant difference between methods but Old Method biased low relative to the New Method. • Opposite of pattern in historical data. • Spatial, temporal or environmental effects do not explain the difference between current and historical bias. • Difference may be due to a change in instrumentation and/or other procedures. • Current bias may be due to difference in accuracy between methods at high levels of TSS.

  31. TP – Recommendations • Data are not conducive to method adjustment. • Blocked Seasonal Kendall for TP until additional studies are available. • Another paired study? • Use old instrumentation or more trouble than it is worth? • Do we control for season or not? • Do we control locations, TSS, salinity or not? • Are there other effects?

  32. CSSP Data • Collected at Three Stations • Two Tidal Fresh stations PMS10 and Wilcox Wharf • One Mesohaline station CB4.4. • May ’95 through June ’04. • 328 total observations. • Surface measurements only.

  33. CSSP TP Data – Screening • Mean difference between methods • significantly different from zero: • Student’s t: • t value =3.79; Prob. >| t |=0.0002 • Wilcoxon Signed Rank: • S value =-4076; Prob. >|S|=0.0002 • Mean difference between methods = • - 0.010.036 mg/L. • 70% of all differences at or below 0.00 with 25% = 0.00. • No significant effect on results by removing outliers. • Data appear to be comparable with slight positive bias for Old Method. + 2 STD Mean - 2 STD

  34. CSSP TP Data – Station PMS10 Only • Mean difference between methods • NOT significantly different from zero: • Student’s t: • t value =-0.968; Prob. >| t |=0.3367 • Wilcoxon Signed Rank: • S value =119; Prob. >|S|=0.1741 • Mean difference between methods = • - 0.010.064 mg/L. • Bias of Old Method positive. • 70% of all differences at or above 0.00 with 29% = 0.00. • Three probable outliers. + 2 STD Mean - 2 STD

  35. CSSP TP Data – Station PMS10 Only • Remove outliers and there is a significant difference: • Student’s t: • t value =2.562; Prob. >| t |=0.013 • Wilcoxon Signed Rank: • S value =264; Prob. >|S|=0.0180 • Mean Difference between methods= 0.010.015 mg/L. • With outliers removed, bias for Old Method is negative. + 2 STD Mean - 2 STD

  36. CSSP TP Data – Station CB4.4 Only • Mean difference between methods significantly different from zero: • Student’s t: • t value =2.562; Prob. >| t |=0.013 • Wilcoxon Signed Rank: • S value =264; Prob. >|S|=0.0180 • Mean Difference between methods= • -0.010.025 mg/L. • No significant effect on results by removing outliers. + 2 STD Mean - 2 STD

  37. CSSP TP Data – Wilcox Wharf • Mean difference was significantly difference from zero: • Student’s t: • t value =-3.44; Prob. >| t |=0.0013 • Wilcoxon Signed Rank: • S value =261; Prob. >|S|=0.0015 • Mean Difference between methods= • -0.010.021 mg/L.

  38. CSSP - Conclusions • Data indicate a slight positive bias in Old Method. • Bias was consistently positive at all stations except at PMS10 when outliers were removed. • Bias was very small (–0.01 mg/L) and does not change in relation to concentration. • Data are not compatible with a method correction.

  39. DCLS Instrument Comparison Studies • Data were part of a larger study comparing 8 variables between the Technicon AA and Skalar AA: • 94 total phosphorus samples • 176 PO4F samples. • Documentation indicates: • Study Conducted in the Summer of 1997 • No information on collection locations.

  40. DCLS Instrument Comparison for TP • Mean difference was NOT significantly different from zero: • Student’s t: t value =-1.58; Prob. >| t |=0.1173 • Wilcoxon Signed Rank:S value =196.5; Prob. >|S|=0.0155. • Mean difference =0.000.01 with most values around 0.00 mg/L. • No bias for the Old Method.

  41. DCLS Instrument Comparison for PO4F • Mean difference between methods • significantly different from zero: • Student’s t: • t value =-11.58; Prob. >| t | <0.0001 • Wilcoxon Signed Rank: • S value =-6341; Prob. >|S| <0.0001. • Mean difference was -0.010.01 mg/L • 83% percent of all observations were at or below zero.

More Related