An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Continuously Monitoring the RPC Gas Mixture with a Gas Chromatograph C. Lu Princeton University

  2. Run GC in sequence mode We created a sequence, which repeatedly execute a GC method file (program) and collect a gas chromatogram every 21’. The corrected gas mixing ratio is directly wrote into a data base. We have collected data continuously for more than 2 weeks. Isobutane Ar See the interpretation of the plot on next slide. SF6 R134A

  3. Interpretation of the results The GC tested mixing ratio for each gas component is not so stable. Is this due to gas mixing panel or GC? Most likely it is due to the gas mixing panel. The repeatability of the MKS flow controller is 0.2% of full range. The full range for the mixing panel we are using is for the Daya Bay near hall RPC, Ar full range is ~2800 sccm, thus the precision would be 2800*0.2% = 5.6 sccm. Now we only use 65 sccm for Ar controller, so we should expect the fluctuation at the level of 5.6/65 = 8.6%. The other three controllers are having similar limitation. To confirm if GC can distinguish small increase of Isobutane percentage, we adjusted the gas mixing ratio at two points. At the first point we decreased the flow rate for Ar and R134A, at the second point we change their flow rates back to original. The small percentage increase of Isobutane can be clearly identified. It gives us a confidence that the GC will be able to carry out the task as a safety device as well.

  4. Dark current and ambient parameters We recorded an RPC chamber’s current with a Co-60 source: HV = 7600V, Daya Bay gas mixture. We also recorded the ambient parameters. Looks to me the most sensitive ambient parameter to the current is relative humidity. We can see that the current follows R.H. nicely. pressure temperature humidity Current

  5. The Varian 3900GC system is a reliable, high-quality tool to check Daya Bay RPC gas system, including the concentration of isobutane. 0.5% of Isobutane mixing ratio change can be reliably detected by GC. Ambient parameters are sensitive to RPC operating current. Since the humidity can change wildly, its effect on the dark current is very important. Conclusion