730 likes | 2.68k Views
MAMMOGRAPHY QA. PURPOSE OF QC. Quality Assurance, Quality Control & Equipment Bushong Ch. 19 & 20. ASSURES OPTIMUM PATIENT CARE by avoiding unnecessary repeat exposures due to artifacts, processing problems and equipment failure. Rev. Spring 2009/2010. MQSA National Statistics.
E N D
MAMMOGRAPHY QA PURPOSE OF QC Quality Assurance, Quality Control & Equipment Bushong Ch. 19 & 20 ASSURES OPTIMUM PATIENT CARE by avoiding unnecessary repeat exposures due to artifacts, processing problems and equipment failure Rev. Spring 2009/2010
MQSA National Statistics • FEDERAL LEGISLATION • OCT. 1, 1994 • EQUIPMENT DESIGN & OPERATION STANDARDS (DEDICATED UNITS) • INTENSIVE QUALITY CONTROL PROGRAM WHAT DOES MQSA REPRESENT? __________________
MAMMO & Quality Control • Quality control is an equipment and processing evaluation program in diagnostic imaging centers. • QC involves testing, record keeping and evaluation of imaging equipment and processing • QC Team includes Radiographer/Mammographer, Physicists and Radiologist • QA involves people – scheduling, reports, waiting time and follow up exams • “double reads” See Bushong – CH. ____ PG ____
TASKS Darkroom Clean Processor QC – sensitometry/densitometry and DEVELOPER temp Screen Cleaning VISUAL CHECK Phantom Images Repeat Analysis (should be less than 2% repeat rate - diagnostic rad = 5%) FREQUENCY Daily Weekly Monthly/Quarterly Elements of QC pg 333
SEMIANNUALLY DARKROOM FOG SCREEN/FILM CONTACT COMPRESSION NOT MORE THAN 40 PSI (25 PSI is average) MAMMO QC MONTHLY Checks
QUALITY CONTROL TOOLS • PHANTOM • DIGITAL THERMOMETER • SCREEN CLEANER • LINT FREE CLOTHS • FIXER RETENTION TESTING KIT • CHARTS
Hypo RetentionChecks for what? Fixer left on film – “Archival Properties”
WIRE MESH Checks for what How long should you wait after loading a cassette to use it? 15 min & placed in an upright position
Types of Safelights? • Wratten 6b • GBX • Check for leaks/cracks? How often? Bulb Wattage?
For those still using film…clean crossovers daily Use “dedicated” processor for only mammo
These specks were produced by flakes trapped between the film and the screen – Not calcifications in the breast
PHANTOM IMAGES • WEEKLY - to assure film density, contrast, uniformity and image quality are optimum – Using the SAME cassette and technical factors each time • 4/5 masses • 3/4specks • 4/6 fibers • should be seen • (see pg 500 review book)
REPEAT ANALYSIS • Mammography should be less than 2% Number of repeated film • Repeat Rate = --------------------------- X 100 Total # of films • Diagnostic Radiology = less than 5% • Students = less than 10 %