2.71k likes | 15.12k Views
“Quality Control” in Hematology. June 14, 2013. M R Tiwari M.Sc -TQM. Laboratory test results (Good laboratory practice, GLP) Clinical diagnosis Patient management. Good Laboratory Practice. Quality can be assured at Pre-analytical stage Analytical stage Post-analytical stage.
E N D
“Quality Control” in Hematology June 14, 2013 M R Tiwari M.Sc-TQM
Laboratory test results (Good laboratory practice, GLP) Clinical diagnosis Patient management
Good Laboratory Practice Quality can be assured at • Pre-analytical stage • Analytical stage • Post-analytical stage
Terminologies = QC at Analytical stage
PRECISION (REPRODUCIBILITY) • Definition Precision refers to the reproducibility of a result. • Comparing QC terms to a targetFigure illustrates that the results are precise (close together) but not accurate (they are not in the bull’s-eye). • Checking precision is required while -calibration -troubleshooting QC at Analytical stage
ACCURACY • Definition Closeness of a result to the true (accepted) value. NOTE: Before determining accuracy, first determine precision. • Comparing QC terms to a targetFigure illustrates that the results are accurate (in the bull’s-eye) and precise (close together). NOTE • You cannot have accuracy without precision. • However, you can have precision without accuracy. QC at Analytical stage
NEITHER ACCURACY NOT PRECISION • This figure illustrates that the results are neither accurate nor precise. • None of the results are close together, and none of them are in the bull’s-eye. QC at Analytical stage
Statistics involved… • Mean • Standard deviation (SD) • ± 1SD • ± 2SD • ± 3SD • Coefficient of variation (%CV) QC at Analytical stage
Hematology analyzers QC at Analytical stage
Technical Specifications QC at Analytical stage
Performance specification (Linearity) QC at Analytical stage
Precision QC at Analytical stage
Test selectivity / Throughputs QC at Analytical stage
Sample Modes / Volumes QC at Analytical stage
Calibration (setting accuracy) Hematology analyzer
Setting accuracy Calibration Is done to standardize the instrument for accuracy. Calibrator Certified Reference Material (CRM) used to calibrate a measurement on an analyzer. Cal-Factors If any deviation from calibration references is observed necessary calibration correction factors are applied to set the accuracy of the instrument. QC at Analytical stage
Calibration • Pre-calibration check • Total maintenance of the instrument • Reagents (replenish or replace) • Calibrator (Check for expiry) • Calibration procedure • Precision check • Carryover check • Calibration in both (open and closed modes) • Post-calibration validation • Run calibrator as samples • Run 3 level controls QC at Analytical stage
CARRYOVER • Carryover is defined as a number of cells remaining behind following the cycling of a blood sample. • This test is performed to determine if one sample interferes with the accurate analysis of the next sample. • Ideally, carryover shall be very low. • Measure a specimen with a high concentration in triplicate, immediately followed by a specimen with a low concentration in triplicate. • Where l1 and l3 are the results of the first and third measurements of the samples with a low concentration and h3 is the third measurement of the sample with a high concentration. QC at Analytical stage
Daily internal quality control (IQC) Hematology analyzers
Daily IQC in hematology analyzers • Commercial controls • Retained sample • Moving averages (Bull’s algorithm) QC at Analytical stage
DAILY IQC of Hematology analyzer QC at Analytical stage
In 1931, Dr. Walter Shewhart, a scientist at the Bell Telephone Laboratories, proposed applying statistical based control charts to interpret industrial manufacturing processes. In 1950, S. Levey& E.R. Jennings suggested the use in the clinical laboratory. Control Chart's Inventor QC at Analytical stage
Establishing Laboratory’s acceptable ranges for controls Reference : CELL-DYN Sapphire Operator’s Manual, 11-16 Establishing laboratory control mean Commercial controls • New lot of control material should be analyzed in parallel with current lot. • This may be accomplished by running the new controls twice a day for five days (we run three times for three days and one time on fourth day). • The mean of the ten runs is then used. • A control file is set up for the new mean. • The same file is used to run control for the reminder dating of period. QC at Analytical stage
Establishing Laboratory’s acceptable ranges for controls Reference : CELL-DYN Sapphire Operator’s Manual, 11-16 Establishing laboratory acceptable ranges (SD) Commercial controls • The expected ranges published by the manufacturers do not represent standard deviations and are generally too broad for effective quality control. • These ranges are determined by evaluating three to six months of data (data from the previous IQC) for a particular level of control. • The individual SD values may be averaged as follows: N = number of values in a group SD = Standard Deviation of the values in that group i = the last group of values The resultant long-term instrument SD and the laboratory established mean is used to monitor instrument performance QC at Analytical stage
L-J chart Interpretation Dr. James O Westgard QC at Analytical stage
Modification of 10x Dr. James O Westgard QC at Analytical stage
× I prefer the ‘xyz’ analyzer because its 2SD QC charts are amazingly flat 2SD QC at Analytical stage
Retained sample testing • A previous days (retained) sample, stored at 2-80C, with normal counts is run as 1st sample after the controls are analyzed. This is done on a daily basis. • This sample is considered as the precision sample. • This sample is then analyzed every one hour or / after 30 patient samples and also as the last sample before the analyzer is shut down. • The mean, SD and CV is monitored. QC at Analytical stage
When a control point is outlier… Warning rule = use other rules to inspect the control points Rejection rule = “out of control” • Stop testing • Identify and correct problem • Repeat testing on patient samples and controls • Do not report patient results until problem is solved and controls indicate proper performance QC at Analytical stage
Other additive IQCs QC at Analytical stage
Flags & correlation with PBS QC at Analytical stage
PBS examination • Ensure good drying of the smear • Do not use overused stains • Maintain required pH of the buffer • Maintain the staining time A good smear Quality tips QC at Analytical stage
Flag 1 • ? Suspect BLASTS • ? Monocytosis PBS correlation Analyzer PBS : Monocytes / Blasts QC at Analytical stage
Flag 2 • ? Suspect BLASTS • ? Lymphocytosis • ? Variant lymphocytes PBS correlation Analyzer PBS : Lymphocytes / Blasts QC at Analytical stage
Flag 3 • ? Platelet clumps • Thrombocytopenia PBS correlation Analyzer PBS : Platelet clumps / QC at Analytical stage
Flag 5 • ? Giant Platelets • Thrombocytopenia PBS correlation Analyzer PBS : Giant platelets QC at Analytical stage
External Quality Assessment(Proficiency testing) QC at Analytical stage
Normal Distribution Curve or Gaussian curve Describes events or data that occur symmetrically about the mean. Out of 100 events 68.7 will fall within ±1 SD 95.4 will fall within ± 2 SD 99.7 will fall within ±3 SD QC at Analytical stage
Comparative performance evaluation Z-Score QC at Analytical stage
If EQAS/PT is not available QC at Analytical stage
If EQAS results are out QC at Analytical stage
Summary… Laboratory test results • Pre-analytical stage • Analytical stage • Post-analytical stage Clinical diagnosis Patient management Quality Good Laboratory Practice Thank you…