Noninvasive Blood Pressure Monitoring. 2. Expected Practice. Measure blood pressure (BP) in the upper arm using the oscillatory or auscultatory method. Unable to measure in the upper armMeasure in forearmConsider thigh or calf measurement. Noninvasive Blood Pressure Monitoring. 3. Expected Prac
E N D
1. Noninvasive Blood Pressure Monitoring Issued April 2010
2. Noninvasive Blood Pressure Monitoring 2 Expected Practice Measure blood pressure (BP) in the upper arm using the oscillatory or auscultatory method.
Unable to measure in the upper arm
Measure in forearm
Consider thigh or calf measurement
3. Noninvasive Blood Pressure Monitoring 3 Expected Practice Use appropriate size BP cuff
4. Noninvasive Blood Pressure Monitoring 4 Expected Practice Measure baseline BP bilaterally
5. Noninvasive Blood Pressure Monitoring 5 Expected Practice Position patient
the appropriate reference level for NIBP is the heart.
6. Noninvasive Blood Pressure Monitoring 6 Patient Position—Seated Patient should be seated with back and arms supported, feet on floor, and legs uncrossed with upper arm at heart level (phlebostatic axis: 4th intercostal space, halfway between the anterior and posterior diameter of the chest) (Figure 1)
7. Noninvasive Blood Pressure Monitoring 7 Patient Position—Supine position patient supine
or
with head of bed at a comfortable level and with upper arm supported at heart level.
8. Noninvasive Blood Pressure Monitoring 8 Expected Practice The patient and the caregiver should remain quiet throughout the procedure of taking a BP
9. Noninvasive Blood Pressure Monitoring 9 Scope and Impact of the Problem Blood pressure is measured in virtually all patients
Accurate measurement of blood pressure is essential to guide management decisions
Inaccuracy may lead to over or under-treatment of the patient’s condition
10. Noninvasive Blood Pressure Monitoring 10 Supporting Evidence Use oscillatory devices that meet the Association for the Advancement of Medical Instrumentation standards (mean difference + 5mm Hg and standard deviation < 8mm Hg) when compared to auscultatory method12 and the appropriate size cuff.
11. Noninvasive Blood Pressure Monitoring 11 Supporting Evidence Stiffness of the arteries, influences amplitude of the oscillations and may cause underestimation MAP
Accuracy of the automated device may also be limited in patients with hypertension, hypotension, and cardiac dysrhythmia.
Some studies showed difference < 5mm Hg between BP measurement methods, other studies demonstrated that individual differences may be > 10mm Hg for some individuals.
Vasopressors show no significant effect on difference
12. Noninvasive Blood Pressure Monitoring 12 Supporting Evidence Forearm and upper arm BP's are not interchangeable
Select the proper cuff size and positioning of forearm at heart level
13. Noninvasive Blood Pressure Monitoring 13 Supporting Evidence Cuff Position
Forearm – position the cuff midway between the elbow and wrist
Thigh – position the cuff over the lower third of thigh ( lower edge of cuff approximately 2 to 3cm above popliteal fossa)
14. Noninvasive Blood Pressure Monitoring 14 Supporting Evidence With all BP measurements select the proper size cuff
Calf BP measurements- place the patient in the supine position. Place the patient in the prone position for thigh BP measurements. If the patient cannot be place prone, position the patient supine with knee slightly bent.
Thigh pressures are normally higher than upper arm pressures
Calf pressures are not interchangeable with upper arm pressures
15. Noninvasive Blood Pressure Monitoring 15 Supporting Evidence Calf BP measurement (referred to as an ankle BP) Korotkoff’s sounds are auscultated over either the dorsalis pedis or posterior tibial artery in calf BP or the popliteal artery in thigh BP
In adults, calf BP's should be used only if the upper arm is not accessible
16. Noninvasive Blood Pressure Monitoring 16 Supporting Evidence Reasons an extremity may not be suitable for BP measurement.
BP cuffs should not be used on extremities with a deep vein thrombosis, grafts, ischemic changes, arteriovenous fistula, or arteriovenous graft
BP cuffs should not be applied over a PICC or midline catheter site
In extremities with peripheral IV while an infusion is running or any trauma/incision.
Patients that have had mastectomy or lumpectomy
do not use the involved arm(s) for BP's if there is lymphedema.
17. Noninvasive Blood Pressure Monitoring 17 Supporting Evidence Appropriate cuff size necessary for accurate measurement of BP in all extremities
Cuff too narrow = overestimation of BP
Cuff too wide = underestimation BP.
Cuff is too small = falsely high reading may result
Cuff is too large = falsely low reading may result
Use a cuff with a bladder capable of going around 80% of the arm
18. Noninvasive Blood Pressure Monitoring 18 Supporting Evidence Patients with aortic dissection, congenital heart disease, coarctation of the aorta, peripheral vascular disease, and unilateral neurological and musculoskeletal abnormalities may demonstrate a difference in inter-arm BP
20% - 40% of individuals without the above conditions may have a measurable difference of 10 to 20mm Hg BP between left and right arms
One study showed higher mean differences in systolic and diastolic BP in older participants
When there is a consistent interarm difference, use the arm with higher pressure.
19. Noninvasive Blood Pressure Monitoring 19 Supporting Evidence Body position and arm position influence the measurement of BP
With patient supine and arm placed at heart level the systolic BP is approximately 8mm Hg higher than in the sitting position
Studies show if arm is below heart level BP readings will be higher, conversely, if the arm is above heart level BP readings will be lower
20. Noninvasive Blood Pressure Monitoring 20 Supporting Evidence Systolic and diastolic BP's in hypertensive and normotensive patients increase with talking
21. Noninvasive Blood Pressure Monitoring 21 Actions for Nursing Practice Ensure that your facility has written procedures BP measurement
Ensure proper size cuffs are readily available
Ensure devices meet appropriate standards
Provide routine training healthcare providers in BP measurement
22. Noninvasive Blood Pressure Monitoring 22