Announcements • Exam 1 today – first 40 min. • Second Homework Set will be online soon • Today’s Topics – Chromatographic Theory • Basic definitions (flow – time relationship, distribution constant, retention factor, velocities, plate number, plate height, asymmetry factor, resolution, separation factor) • How to read chromatograms • Meaning of parameters (more when we cover optimization)
Chromatographic TheoryQuestions on Definitions • When is chromatographic separation needed vs. only simple separations? • An analyte interacts with a stationary phase via adsorption. The stationary phase is most likely: a) Liquid b) Liquid-like c) Solid • What are the required two phases in chromatography called? • What are advantages and disadvantages with the three common stationary phases (liquid, liquid-like, and solid)?
Chromatographic TheoryDefinition Section – Flow – Volume Relation • Relationship between volume (used with gravity columns) and time (most common with more modern instruments): V = t·F V = volume passing through column part in time t at flow rate F Also, VR = tR·F where R refers to retention time/volume (time it takes component to go through column or volume of solvent needed to elute compound)
Chromatographic TheoryDefinition Section – More on Volume • Hold-up volume = VM = volume occupied by mobile phase in column • Stationary phase volume = VS • Calculation of VM: VM = Vcolumn – Vpacking material – VS VM = tM·F, where tM = time needed for unretained compounds to elute from column
Chromatographic TheoryDefinition Section – Partition and Retention • Distribution Constant (= Partition Coefficient from LLE) = KC = [X]S/[X]M • KC is constant if T and/or solvent remain constant • Retention Factor (= Capacity Factor = Partition Ratio) = k = (moles X)S/(moles X)M = KC/(VM/VS) • k = KC/β where β = VM/VS • Retention Factor is more commonly used because of ease in measuring, and since β = constant, k = constant·KC (for a given column) • Note: kColumn1 ≠ kColumn2 (because β changes)
Chromatographic TheoryDefinition Section – Partition and Retention • Since the fraction of time a solute molecule spends in a given phase is proportional to the fraction of moles in that phase, k = (time in stationary phase)/(time in mobile phase) • Experimentally, k = (tR– tM)/tM • The same equations can be made with volumes instead: k = (VR– VM)/VM • Note: t’R = tR– tM = adjusted retention time
Chromatographic TheoryReading Chromatograms • Determination of parameters from reading chromatogram (HPLC example) • tM = 2.37 min. (normally determined by finding 1st peak for unretained compounds – contaminant below) • VM = F·tM = (1.0 mL/min)(2.37 min) = 2.37 mL (Note: 4.6 x 250 mm column, so total vol. = (π/4)(0.46 cm)2(25 cm)(1 mL/cm3) = 4.15 mL • Vol. of packing material + stationary phase = 4.15 mL – 2.37 mL = 1.78 mL (note only VS is useful) • 1st peak, tR = 5.93 min.; k = (5.93 -2.37)/2.37 = 1.50
Chromatographic TheoryWhat do all these Parameters Mean? • KC is just like KP in liquid – liquid extractions for HPLC or KH (Henry’s law constant) for GC • Large KC value means analyte prefers stationary phase • In GC: • KC value will depend on volatility and polarity (analyte vs stationary phase) • KC value adjusted by changing T (most common) • The mobile phase or carrier gas (e.g. He vs. N2) has no effect on KC • In HPLC • KC value will depend on analyte vs. mobile phase and stationary phase polarity • KC value adjusted by changing mobile phase polarity
Chromatographic TheoryWhat do all these Parameters Mean? II • Retention Factor is a more useful measure of partitioning because value is related to elution time • Compounds with larger KC, will have larger k, and will elute later • Practical k values • ~0.5 to ~10 • Small k values → usually poor selectivity • Large k values → must wait long time • Higher k values are more practical for complicated samples while low k values are desired for simpler samples to save time
Chromatographic TheoryDefinition Section – Velocity • Mobile phase velocity (u) and analyte average speed (v) can be useful quantities • u = L/tM (L = column length) • v = L/tR • R = retardation factor = v/u (similar to RL used in TLC based on distance migrated)
Chromatographic TheoryReading Chromatogram – cont. • u = L/tM = 250 mm/2.37 min = 105 mm/min • v(1st peak) = L/tR = 250 mm/5.93 min = 42.2 mm/min • R = 42.2/105 = 0.40
Chromatographic TheoryShape of Chromatographic Peak • Gaussian Distribution • Normal Distribution Area = 1 • Widths • σ (std deviation) • wb (baseline width) = 4σ • wh (peak width at half height) = 2.35σ • w’ = Area/ymax = 2.51σ (often given by integrators) Gaussian Shape (Supposedly) Inflection lines Height 2σ Half Height wh wb
Chromatographic TheoryMeasures of Chromatographic Efficiency • Plate Number = N (originally number of theoretical plates – similar to number of liquid-liquid extractions or distillations) • N = (tR/σ)2 (= 16(tR/wb)2 ) • N is an absolute measure of column efficiency but depends on length • Plate Height = H = length of column needed to get N of 1 • H = L/N, but H is constant under specific conditions, while N is proportional to L
Chromatographic TheoryMeasures of Chromatographic Efficiency • Measuring N and H is valid under isocratic conditions • Later eluting peaks normally used to avoid effects from extra-column broadening • Example: N = 16(14.6/0.9)2 = 4200 (vs. ~3000 for pk 3) • H = L/N = 250 mm/4200 = 0.06 mm Wb ~ 0.9 min
Chromatographic TheoryNon-ideal Peak Shapes Fronting Peak (TF < 1) Tailing Peak (actually slow detector) a b Tailing Factor = TF = b/a > 1 (tailing peak)
Chromatographic TheoryDefinitions - More on Peak Shapes • A Gaussian peak shape is assumed for many of the calculations given previously (e.g. peak width and N) • For non-Gaussian peaks, the equations relating specific widths to σ are no longer valid. • New equations are required for equations that have width in them
Chromatographic TheoryDefinitions - Resolution • Resolution is a measure of the ability to separate two peaks from each other • Resolution = RS where d = (tR)B– (tR)A and ave w = [(wb)A + (wb)B]/2
Chromatographic TheoryDefinitions - Resolution • Resolution indicates the amount of overlap between peaks • RS < 1, means significant overlap • RS = 1.5, means about minimum for “baseline resolution” (at least for two peaks of equal height) • RS > 2 often needed if it is important to integrate a small peak near a large peak
Chromatographic TheoryDefinitions - Resolution • RS calculation examples: • 1st two peaks: • tR(1st pk) = 4.956 min., w (integrator) = w’ = 0.238 min, so wb = 0.238·(4/2.5) = 0.38 min. • tR(2nd pk) = 5.757 min., wb = 0.44 min RS = 0.801/0.410 = 1.95 (neglecting non-Gaussian peak shape) • Last two peaks, RS = 3.0
Chromatographic TheoryDefinitions - Resolution • Higher resolution values are needed to quantify small peaks next to large peaks • RS = 1.61 (assuming wb 1st peak equals 2nd peak) • RS is not sufficient for accurate integration of 1st peak (but o.k. for integration of 2nd peak) Expansion of above box Large integration error on 1st pk
Chromatographic TheoryDefinitions - Peak Capacity • Peak Capacity is the theoretical maximum number of peaks that can be separated with RS = 1.0 within a given time period. • We won’t cover calculation, but for example, about 2X more peaks could be possible between 5 and 13 min. • Peak capacity 2.3 to 20 min. would be ~27 peaks. • Greater peak capacity is typical with temperature/gradient programs (like in example).
Chromatographic TheoryDefinitions - Separation Factor • Separation Factor = a = ratio of distribution constants • a = KB/KA = kB/kA = (t’R)B/(t’R)A Where (tR)B > (tR)A so that a > 1 • Smaller a (closer to 1) means more difficult separation • In example chromatogram, (1st 2 peaks) a = (5.77 – 2.37)/(4.96 – 2.37) = 1.31
Chromatographic TheoryDefinitions - Overview • The “good” part of chromatography is separation, which results from differences in KC values giving rise to a > 1 • The “bad” part of chromatography is band broadening or dispersion, leading to decreased efficiency (and also reducing sensitivity) • The “ugly” part of chromatography is non-Gaussian peak shapes (leads to additional band broadening plus need for new equations)
Chromatographic TheoryQuestions on Definitions • List two ways in which a stationary phase is “attached” to a column? • What column component is present in packed columns but not open-tubular columns? • In HPLC, typical packing material consist of μm diameter spherical particles. Even though tightly packing the spheres should lead to > 50% of the column being sphere volume, the ratio of VM/Column Volume > 0.5. Explain this.
Chromatographic TheoryQuestions on Definitions • List 3 main components of chromatographs. • A chemist perform trial runs on a 4.6 mm diameter column with a flow rate of 1.4 mL/min. She then wants to scale up to a 15 mm diameter column (to isolate large quantities of compounds) of same length. What should be the flow rate to keep u (mobile phase velocity) constant? • A chemist purchases a new open tubular GC column that is identical to the old GC column except for having a greater film thickness of stationary phase. Which parameters will be affected: KC, k, tM, tR(component X), β, a.
Chromatographic TheoryQuestions on Definitions • What “easy” change can be made to increase KC in GC? In HPLC? • A GC is operated close to the maximum column temperature and for a desired analyte, k = 10. Is this good? • If a new column for problem 8 could be purchased, what would be changed? • In reversed-phase HPLC, the mobile phase is 90% H2O, 10% ACN and k = 10, is this good? • Column A is 100 mm long with H = 0.024 mm. Column B is 250 mm long with H = 0.090 mm. Which column will give more efficient separations (under conditions for determining H)?
Chromatographic TheoryQuestions on Definitions • Given the two chromatograms to the right: • Which column shows a larger N value? • Which shows better resolution (1st 2 peaks top chromatogram)? • Which shows better selectivity (larger a; 1st 2 peaks on top)? • Should be able to calculate k, N, RS, and α Unretained pk