Fermentation Vessel Automation. SD Team: Dec06-07 December 12, 2006. Client: Stephanie Loveland Department of Chemical and Biological Engineering. Team Members: Andrew Arndt Adam Daters Brad DeSerano Austin Striegel. Advisor: Dr. Degang Chen. Presentation Outline.

saad
Download Presentation

Fermentation Vessel Automation

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Fermentation Vessel Automation SD Team: Dec06-07 December 12, 2006 Client: Stephanie Loveland Department of Chemical and Biological Engineering Team Members: Andrew Arndt Adam Daters Brad DeSerano Austin Striegel Advisor: Dr. Degang Chen

  2. Presentation Outline • Project Overview • Research Activities • Hardware Configuration • Software Development • Implementation • Resources and Scheduling • Lessons Learned • Closing Remarks • Questions

  3. Acknowledgements • Stephanie Loveland • Provided finances, design specifications, and requirements for the project • Dr. Degang Chen • Technical and practical advice

  4. Definitions • DAQ – Data acquisition • Flash – Animated graphics technology and format from Macromedia • GUI – Graphical user interface • LabVIEW – Laboratory Virtual Instrument Engineering Workbench • PPM – Parts per million • RPM – Rotations per minute • RS232 – Standard for serial cable interface • SCC – Signal conditioning system offered by National Instruments • SLM – Standard liters per minute • VI (virtual instruments) – Sub-unit program in LabVIEW that represents the appearance and function of a physical implement

  5. Problem Statement • A mock fermentation vessel is available for use by senior chemical engineering students • Simple methods were used to record data (Paper and Pencil) • An automated data collection system needed to be developed to gather the data • Upgrade equipment as needed

  6. Problem Solution-Approach • Designed and installed new hardware for the mock fermentation vessel apparatus • Data acquisition card • Signal conditioning modules • Oxygen concentration meter • Created automatic data collection software with LabVIEW • Recorded results with software to Excel workbook

  7. Problem Solution-Approach Equipment Data Recorded

  8. Intended Users • Senior level students in the Department of Chemical and Biological Engineering as well as faculty in the department • Users must have knowledge of safety procedures and requirements while conducting experiments within the lab • Users will need to have been exposed to the concepts that the lab is designed to simulate

  9. Intended Uses • Automate the collection of the data from the mock fermentation vessel apparatus • Display data in real-time • Record data into Excel workbook for further analysis • Use of this system is not supported on any other equipment not supported

  10. Operating Environment • Location in 2059 Sweeney • Temperature controlled environment • 60°F to 80°F Laboratory Apparatus

  11. Assumptions (1/2) • The end-user of this project will be someone who is familiar with the fermentation process • Only one experiment will be conducted at a time • Environmental stability of 2059 Sweeney will be maintained • All new components and cables will be paid for by the client • All laboratory components will operate within their given rated power values

  12. Assumptions (2/2) • A computer will be supplied by the client with LabVIEW and Excel already installed • An extra PCI slot will be available on the computer for data acquisition card • The data acquisition card will supply its own clock

  13. Limitations (1/2) • File format type is in Excel format • Software shall be written using LabVIEW • One sample every five second must be recorded from each specified device • Maximum flow rate for the air/nitrogen must be less than 6 SLM • Motor speed must be kept less than 800 RPM • Safety glasses must be worn at all times when working in 2059 Sweeney

  14. Limitations (2/2) • No more than 4 significant digits stored upon measurement • The voltage signals from the stirrer motor control must be electrically isolated • The oxygen concentration meter must read from 0 to 9.5 PPM dissolved oxygen • The oxygen concentration meter must be a benchtop unit

  15. End Product and Deliverables • A fully automated and integrated data collection system • A graphical user interface (GUI) designed in LabVIEW • Instruction manual and documentation for the data collection system

  16. Present Accomplishments • Purchased and installed all hardware for automated data collection • Collected data from each piece of lab equipment • Tested functionality of software as a team • Tested functionality of software with intended users, received feedback • Delivered completed software with software feedback implemented

  17. Future Required Activities • Review user manual with client • Review programmer’s manual with client

  18. Technology Considerations (1/4) • Data Acquisition Board • Signal Conditioning • Oxygen Concentration Meter

  19. Technology Considerations (2/4) Data Acquisition Board • USB DAQ • Inexpensive and Easy Connection • No Signal Conditioning Capability • PXI DAQ System • High Resolution/High Sampling Rate • High Cost • Signal Conditioning Capability Technology Selected: PCI DAQ Board • PCI DAQ Board • Moderate Resolution & Sampling Rate • Moderate Cost • Signal Conditioning Capability

  20. Technology Considerations (3/4) Signal Conditioning • No Signal Conditioning • Less Cost • Unable to interface directly with DAQ board • Signal Conditioning • Isolation requirements met for Stirrer Motor Control • Easy interface with DAQ board • Extra cost of Signal Conditioning Carrier Box Technology Selected: Signal Conditioning

  21. Technology Considerations (4/4) Oxygen Concentration Meter • Omega DOB-930 • 100 data point logging • RS232 Interface • Limited support and availability Technology Selected: Thermo Electron Orion 3-Star • Thermo Electron Orion 3-Star • 200 data point logging • RS232 Interface • 3-year Extended Warranty and availability up to 5 years

  22. Detailed Design (1/8) Hardware Data Flow Configuration

  23. Detailed Design (2/8) Oxygen Concentration Meter and Interface • Thermo Electron Orion 3-Star • Full Scale Measurement of Dissolved Oxygen (0-9.5 PPM) • Interface • Onboard RS232 Connection port for data acquisition • Meter is configured to transfer data every 5 seconds to the PC • Data is acquired using the onboard COM port of the computer supplied

  24. Detailed Design (3/8) Mass Gas Flow Meter and Interface • Omega FMA-5610 • Full Scale Measurement of Gas Flow from 0 to 10 SLM • Analog 0-5V Output Signal • Interface • 9-Pin D Connector: Pins 2-3 voltage output • SCC-AI04 is used to isolate and condition the 0-5V signal • SCC Module is plugged into the SCC Carrier for interface with the DAQ board

  25. Detailed Design (4/8) Signal Conditioning Carrier Unit • SCC Carrier SC-2345 • Direct Cabling to the M-Series DAQ Board • Housing for up to 20 SCC Modules • Powered by DAQ Board with 5V Signal • Interface • Connects to the DAQ board via a 68 pin shielded connector cable

  26. Detailed Design (5/8) Stirrer Motor Control and Interface • Glas-Col GKH-Stir Tester • Two analog voltage outputs (0-5V) • Operates with a floating ground at 70-90V • 60V fast transient spikes on voltage lines • Interface • 4 pin terminal connection (Differential Voltage) • SCC-AI04 is used to isolate the analog input up to 300V • Voltages are measured differentially to protect against transient spikes • SCC Module is plugged into the SCC Carrier to interface with the DAQ board

  27. Detailed Design (6/8) Data Acquisition Card • Interface • Connects with the Signal Conditioning Carrier via the 68 pin shielded cable • Supplies internal clock for data acquisition of signals • 6 Channels of analog inputs are used for acquiring mass gas flow, torque, and speed • Automatic VI’s in LabVIEW define the operation of the DAQ card • NI PCI-6221 M-Series DAQ Board • 16 Analog Inputs, 2 Analog Outputs, 24 Digital I/O Lines, 2 Counters/Timers • 16 Bit Resolution – Accuracy of 70μV • Sampling Rate: 250 kilo-samples/sec

  28. Detailed Design (7/8) Software Design and Implementation

  29. Detailed Design (8/8) Software Interface

  30. Implementation Activities • Determined scaling of devices for proper measurement • Determined proper connection for obtaining stirrer motor control data • No documentation • Contacted manufacturer and obtained more information • Used multimeter to determine correct wiring • Added multiple tab data writing after obtaining beta testing feedback

  31. Testing Activities • Team Testing • Individual unit testing • Overall GUI functionality testing • Beta Testing • Student testing with actual laboratory experiments • Four groups of students tested • Surveys filled out by each group • Changes applied from feedback: • Experiment data on a new worksheet in an Excel file

  32. Resources Personnel Hours

  33. Resources Other Resources

  34. Resources Financial Resources

  35. Schedule

  36. Project Evaluation (1/2) • Technology Research and Selection • 100% Completed • Design • 100% Completed • Implementation • 100% Completed • Testing • 100% Completed • Documentation • 100% Completed

  37. Project Evaluation (2/2) With a score above 90%, the project has fully met and exceeded all expectations Making the project a complete success Legend: Greatly Exceeded (1.1) – Minimum expectations were met with the addition of several extra features Exceeded (1.0) – Minimum expectations were met with the addition of one or more extra features Fully Met (0.9) – Minimum expectation were met Partially Met (0.5) - Some of the minimum expectations were met Not Met (0.0) – None of the minimum expectations were met

  38. Commercialization • Project was not designed to be commercialized • With small software changes the system would be extendable to collect data from similar or newer equipment

  39. Future Recommendations • Total automation of the system via computer controlled laboratory equipment • Current system would allow for computer control following software changes • Dependent upon client preference

  40. Lessons Learned (1/4) Successes • Client relationship • Time management • Project completed earlier than expected • Beta testing occurred early, allowed for more changes • Advisor Advice

  41. Lessons Learned (2/4) Setbacks • Incorrect SCC module purchased initially • Stirrer motor control pin out

  42. Lessons Learned (3/4) Experienced Gain • LabVIEW Programming • Data acquisition and signal conditioning • Troubleshooting problems • Client relations • Delegating responsibilities • Communication skills

  43. Lessons Learned (4/4) What we would do differently • More research into each piece of equipment • Obtain better LabVIEW reference

  44. Risk and Risk Management • Equipment damage • Broken vessel overcome by team • Replacement ordered by client • Wrong module purchase • Initial mass gas flow module wrong input • Used stirrer motor control module during development • Team member loss • No team member lost during duration of project • Human injury • Standard safety procedures are followed by team while working in Sweeney lab

  45. Closing Remarks • Students collected by pencil and paper data from each laboratory equipment every 10-15 seconds • An automatic data collection system was successfully created using data acquisition and LabVIEW software • Users can view real-time data, and do further analysis with electronically saved data

  46. Demonstration