Application Note: CD Measurement for FET Features. NGS 3500 Advanced Metrology System. CD Measurements. Image Data Set. Notes: 20 High Resolution Images of the same site were used in the repeatability test
E N D
Application Note:CD Measurementfor FET Features NGS 3500 Advanced Metrology System
Image Data Set Notes: • 20 High Resolution Images of the same site were used in the repeatability test • Wafer was slightly moved between measurement to create realistic positional and focus variation • No vibration isolation or special fixture was used in the test
Data Summary – CD Measurement • < 0.01 micron Repeatability • 4 seconds Processing Time (80 measurements) Notes: • Gate Width was calibrated as 0.5 um. • Other measurements were not calibrated. • Optics: 100x Objective (FOV ~ 90um x 75 um) • Camera: High-Res. Color camera with 0.036 um/ pixel
ChromaticEdge Contrast Technology * * Patent Pending • This technology based on wavelength detection enables much better resolution for optical CD measurement. Resolutions below 10 nm are possible with standard optical microscopy. • The edges are very sharp based on wavelength contrast. Due to optical resolution limit, the same edge is blurred with noise and low contrast when light intensity (B&W) information is used.