Electronics systems for the ClearPEM -Sonic scanner José C. DA SILVA, LIP-Lisbon Tagus LIP Group * * J.C.Silva , M.Ferreira , S.Ferreira , J.Neves , R.Bugalho , R.Silva , G.Silva , C.Ortigao , J.Varela. Presentation Guidelines. Introduction to ClearPEM-Sonic
E N D
Electronics systems for the ClearPEM-Sonic scannerJosé C. DA SILVA, LIP-LisbonTagus LIP Group **J.C.Silva, M.Ferreira, S.Ferreira, J.Neves, R.Bugalho, R.Silva, G.Silva, C.Ortigao, J.Varela
Presentation Guidelines • Introduction to ClearPEM-Sonic • Front-end electronics and Service Boards • Data Acquisition Electronics (DAE) and test boards • Front End Board (FEB) Validation tests • Status of scanner assembly
Presentation Guidelines • Introduction to ClearPEM-Sonic • Front-end electronics and Service Boards • DAE and test boards • FEB Validation tests • Status of scanner assembly
ClearPem Sonic • ClearPEM-Sonic: • a project in the framework of CERIMED that combines: • a dedicated mammography PET, the ClearPEM • an UltraSound transducer working in elastographic mode from SuperSonicImagine • Same Electronics • Mechanical structure revised • Breast Contention • Image fusion
Presentation Guidelines • Introduction to ClearPEM-Sonic • Front-end electronics and Service Boards • DAE and test boards • FEB Validation tests • Status of scanner assembly MADE IN LIP
Detector Heads Electronics • Two detector plates: • 6144 LYSO 2x2x20 mm3 crystals • 12288 APD pixel channels • Dual readout of crystal pixels • ( for DoI measurement) • 160x180 mm2 active area • Water cooling
Detector Heads Electronics • LYSO 2x2x20 mm3 crystals • Detector modules • 64 Xtals ,4 APDs
Detector Heads Electronics Characteristics Technology: CMOS 0.35μm Area: 70mm2 Input: 192 channels Output: 2 highest channels Shaping: 40ns Analog Memories: 10 pulse samples Clock Frequency: 50-100MHz Power: 3.6 mW/channel
Detector Heads Electronics • Frontend Board • Contains 2 ASICs for signal selection • 2 High-speed dual ADCs (10bit, 100MHz) • 1 LVDS transmitter (600Mb/s) • Supermodule • 2 FE Boards and detector modules • Processes 768 APD channels • Detector Plate • 8 Supermodules • One board for bias-voltage regulation, power supply and clock distribution Detector Plate 4.5 cm
Detector Heads Electronics • PCB Production issues • Bond Pads etching and gold plating • ENIG= electroless nickel immersion gold
Detector Heads Electronics • Auxiliary Boards • Service Board: • LV, HV, Threshold, Reset , • Temperature readout • HV Matrix • Clock Fanout Board M.Ferreira
Service RACK • DAQ Server • LV and HV • DSS • Network Bridge • Patch Pannel • Isolation Transformer
Software • Service Manager • Software Acquisition Manager • (R.Bugalho) R.Bugalho/G.Silva
Presentation Guidelines • Introduction to ClearPEM-Sonic • Front-end electronics and Service Boards • DAE and test boards • FEB Validation tests • Status of scanner assembly
DAE and Trigger Crate • DAE Crate System • Single 19” 4U crate • Uses two cPCI backplanes • 4 DAQ Boards • 1 Trigger & Data Concentrator Board • Coincidence Trigger • Energy and time determination in Virtex FPGA’s • Calibration constants for individual channels • Programmable energy thresholds and coincidence window • Fast link to Data Acquisition Server bandwidth up to 6.4Gb/s
Test Boards FE Emulator 5 boards manufactured and in use Important for the validation of the DAQ Boards FE Receiver 7 boards manufactured, 6 in use Important for the validations of the FE Boards FEB Tester 1 board assembled Important for FEB production tests Currently under tests by INOV and LIP
Presentation Guidelines • Introduction to ClearPEM-Sonic • Front-end electronics and Service Boards • DAE and test boards • FEB Validation tests • Status of scanner assembly J.Neves/R.Silva
Ok to Globtop FEB Validation Tests • Evaluation of FEB´s performance: • CharacterizationandQuantification of the mean pedestal (baseline) and the ASIC Noise (pedestal RMS) • Looking for deadandnoisy ASIC channels
ASIC 1 Good Results ! ClearPEM Sonic FEBs – Test Results ASIC 0
ClearPEM Sonic FEBs – Test Results P36 Bad Results ! P31 P10
FEB Validation Tests 1st Batch 2nd Batch - 22 FEBs CERN (C1, C2,…, C22) -> OK: 12/22 - 10 FEBs Photochemie (P1, P2,…, P10) -> OK: 8/10 - 40 FEBs Photochemie (P11, P12,…, P50) -> OK: 33/38
Presentation Guidelines • Introduction to ClearPEM-Sonic • Front-end electronics and Service Boards • DAE and test boards • FEB Validation tests • Status of scanner assembly
Scanner Assembly Status • Detector Heads • All Electronics in hand • Assembly of the SMs started • Cables and Integration tests • DAQ and Services RACK • LV, HV, DSS , Network, etc : All in hand • DAE • ALL Boards Produced, New Crate fully equipped and in tests
Scanner Assembly Status • Scanner Mechanics • Major components received and assembled • Main assembly in February (cables, pipes, etc) • Expected fully deliver by March
Final words • All major components are in hand or close to be delivered • Important to really keep deadlines now!! • System fully functional from February to have 2 month Calibration/Validation • deliver the ClearPEM Scanner in Marseilles in May Thank you