1 / 24

What data need to be measured?

What data need to be measured?. Scattering from: specimen in its container empty specimen container standard or calibration specimen. What data need to be measured?. Scattering from: specimen in its container empty specimen container standard or calibration specimen dark count (noise)

shanae
Download Presentation

What data need to be measured?

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. What data need to be measured? • Scattering from: • specimen in its container • empty specimen container • standard or calibration specimen

  2. What data need to be measured? • Scattering from: • specimen in its container • empty specimen container • standard or calibration specimen • dark count (noise) • detector dead time • background

  3. What data need to be measured? • Scattering from: • specimen in its container • empty specimen container • standard or calibration specimen • dark count (noise) • detector efficiency • background • transmissions of each specimen • incident beam intensity

  4. What data need to be measured? • Scattering from: • specimen in its container • empty specimen container • standard or calibration specimen • to get absolute intensity measurements, standard must have known scattering intensity

  5. What data need to be measured? • Scattering from: • specimen in its container • empty specimen container • standard or calibration specimen • dark count (noise) • includes errant x-rays - measure w/ strongly absorbing material in place of specimen • may vary with time

  6. What data need to be measured? • Scattering from: • specimen in its container • empty specimen container • standard or calibration specimen • dark count (noise) • detector efficiency • scale each datum to that of a detector w/ constant efficiency

  7. What data need to be measured? • Scattering from: • specimen in its container • empty specimen container • standard or calibration specimen • dark count (noise) • detector efficiency • scale each datum to that of a detector w/ constant efficiency

  8. What data need to be measured? • Scattering from: • specimen in its container • empty specimen container • standard or calibration specimen • dark count (noise) • detector efficiency • background • e.g., pure solvent, or pure matrix material

  9. What data need to be measured? • Scattering from: • specimen in its container • empty specimen container • standard or calibration specimen • dark count (noise) • detector efficiency • background • transmissions of each specimen • cut beam intensity, measure direct beam I w/ & w/o • specimen

  10. What data need to be measured? • General procedure: • 1. Scale all data to incident beam monitor (synchrotron beam decays w/ time)

  11. What data need to be measured? • General procedure: • 1. Scale all data to incident beam monitor (synchrotron beam decays w/ time) • 2. Correct for detector efficiency

  12. What data need to be measured? • General procedure: • 1. Scale all data to incident beam monitor (synchrotron beam decays w/ time) • 2. Correct all data for detector efficiency • 3. Correct all data for empty cell and dark count • Now have: • Ispecimen = (Ispecimen - Idark)- (Tcell full/Tcell empty)(Icell empty -Idark) • Ibkgrd = (Ibkgrd - Idark)- (Tbkgrd/Tcell empty)(Icell empty -Idark) • Istd = (Istd - Idark)- (Tstd/Tcell empty)(Icell empty -Idark) corr corr corr

  13. What data need to be measured? General procedure: Ispecimen = (Ispecimen - Idark)- (Tcell full/Tcell empty)(Icell empty -Idark) Ibkgrd = (Ibkgrd - Idark)- (Tbkgrd/Tcell empty)(Icell empty -Idark) Istd = (Istd - Idark)- (Tstd/Tcell empty)(Icell empty -Idark) 4. Get absolute I s Iabs = (Ispecimen f(q)/ Istd) (tstd Tstd/tspecimen Tspecimen) Ibkgrd = (Ibkgrd f(q)/ Istd) (tstd Tstd/tbkgrd Tbkgrd) corr corr corr corr final corr corr final

  14. What data need to be measured? General procedure: Ispecimen = (Ispecimen - Idark)- (Tcell full/Tcell empty)(Icell empty -Idark) Ibkgrd = (Ibkgrd - Idark)- (Tbkgrd/Tcell empty)(Icell empty -Idark) Istd = (Istd - Idark)- (Tstd/Tcell empty)(Icell empty -Idark) 4. Get absolute I s Iabs = (Ispecimen f(q)/ Istd) (tstd Tstd/tspecimen Tspecimen) Ibkgrd = (Ibkgrd f(q)/ Istd) (tstd Tstd/tbkgrd Tbkgrd) 5. Subtract bkgrd corr corr corr corr final corr corr final

  15. What data need to be measured? Background treatment - complex need to consider what bkgrd should be - w/ solns or dispersions, bkgrd usually includes scattering from solvent or dispersion medium & sample cell

  16. What data need to be measured? Background treatment - complex need to consider what bkgrd should be - w/ solns or dispersions, bkgrd usually includes scattering from solvent or dispersion medium & sample cell Standard specimen reproducible stable known scattering data scattering must be high isotropic scattering

  17. Instrument resolution See Need to consider wavelength spread collimation effects detector resolution

  18. Instrument resolution Wavelength spread Lab x-rays + monochromator Monochromator crystal set to scatter characteristic line - wavelength spread is result of combined effects of natural line width mosaic spread of monochromator crystal collimation before & after crystal

  19. Instrument resolution Wavelength spread Lab x-rays + monochromator Monochromator crystal set to scatter characteristic line - wavelength spread is result of combined effects of natural line width mosaic spread of monochromator crystal collimation before & after crystal Synchrotron spectrum conts - wavelength spread determined only by mosaic spread of monochromator & collimation

  20. Instrument resolution Wavelength spread Lab x-rays + monochromator Monochromator crystal set to scatter characteristic line - wavelength spread is result of combined effects of natural line width mosaic spread of monochromator crystal collimation before & after crystal Synchrotron spectrum conts - wavelength spread determined only by mosaic spread of monochromator & collimation Both conventional & synchrotron X-ray sources - wavelength spread ~/<> typically <10-3 & neglected in most cases

  21. Instrument resolution Collimation

  22. Instrument resolution Detector 3 contributions to spatial resolution division of detector into pixels method of detection method of position determination latter 2 dominant

  23. Instrument resolution Combined resolution fcn arameter related to FWHM of function ex:

  24. Instrument resolution Combined resolution fcn arameter related to FWHM of function ex:

More Related