An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. RPS For Measurement of ECG m5151117 Yumiko Kimezawa

  2. Outline RPS • Our Architecture • Important Things in Measurement of ECG Signal • Book Knowledge • Question • Requirements for Measurement of ECG • Future Work

  3. Our Architecture RPS Processing results are sent through the Ethernet Sensor A/D converter Processing Filter RAM Results From patient

  4. Important Things in Measurement of ECG Signal RPS • Clinically-important frequency component of ECG signal • - 0 ~ 250Hz Change in voltage inside this range must be faithfully reproduced • Change in voltage in chest wall • Less than 3mV (very weak) • Measurement in environments with much noise Voltage gain of high signal-to-noise ratio (250 to 1000 times) must be gained • Signal-to-noise ratio: The ratio of signal power to the noise power

  5. Book Knowledge RPS • An example of block diagram illustrating cardiography equipment Lead Code Head amplifier Isolation circuit Impedance converter Lead network Power Circuit Main amplifier Micro computer ADC Output equipment

  6. Book Knowledge RPS • An example of block diagram illustrating cardiography equipment Lead Code Head amplifier Isolation circuit Impedance converter Lead network Power Circuit • This circuit prevents body’s impedance from being lower than input circuit’s impedance Main amplifier Micro computer ADC Output equipment

  7. Book Knowledge RPS • An example of block diagram illustrating cardiography equipment Lead Code Head amplifier Isolation circuit Impedance converter Lead network Power Circuit • Constitution of 12-lead from the combination of sensors • A~J : Sensors • Example • Ⅰ: Output between A and E • aVR: Output between A and F Main amplifier Micro computer ADC Output equipment

  8. Book Knowledge RPS • ECG signal is very weak • The signal must not be amplified suddenly • Differential amplifier is usedas head amplifier • An example of block diagram illustrating cardiography equipment An amplifier is needed Lead Code Head amplifier Isolation circuit Impedance converter Lead network The signal is amplified in stages Power Circuit • Noises are reduced • signal-to-noise ratio is improved Main amplifier Micro computer ADC Output equipment

  9. Book Knowledge RPS • An example of block diagram illustrating cardiography equipment Lead Code Head amplifier Isolation circuit Impedance converter Lead network Power Circuit • For securing of security, impedance converter, lead network and head amplifier have to be isolated from main amplifier, power circuit and A/D converter Main amplifier Micro computer ADC Output equipment

  10. Question RPS • A/D converter (AD9254) is included in HSMC • Pipelined switched capacitor ADC • Sampling frequency: 150MHz • Resolution: 14-bit • Power: 430mW HSMC Performance is overmuch • Clinically-important frequency component of ECG signal is from 0 to 250Hz 500Hz is enough for sampling frequency • Using a lot of electricity • Conversion process is complex • Using A/D converter in HSMC is of questionable value

  11. Requirements for Measurement of ECG RPS • An amplifier is needed • Securing of security also have to be considered

  12. Future Work RPS • Studying cardiography equipment using digital filtering techniques

  13. RPS

  14. 心電計の回路を表すブロック図の例 RPS 誘導コード インピーダンス変換器 前置 増幅器 アイソレーション回路 誘導ネットワーク 電源回路 主増幅器 マイクロコンピュータ AD変換器 出力装置

  15. 標準12誘導を作るための誘導ネットワークの原理標準12誘導を作るための誘導ネットワークの原理 RPS B C D A~Jの各点の出力を組み合わせる。 例えば、第Ⅰ誘導はAとE、aVRはAとFを組み合わせたものとなる。 A E J F I H G