An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Cosmic Ray DetectionExperimental Report Alisa Bredensteiner Endeavour Institute Santa Cruz Institute of Particle Physics Quarknet August 2010

  2. Objective 1: To establish the relative accuracy of the newly assembled CRMD (DAQ #6416) through analysis of a 5.4 x 1013 nanosecond performance study.

  3. Location of Muon Detection36:59.909452 N, 122:03.625697 W

  4. Methodology Actual run time: 1.77 x 1012 nanoseconds (approx. 30 minutes) Data Settings: • 4 stacked scintillatorpaddles with a fixed separation distance of about 3 cm • a2-fold coincidence level at a 40 ns pre-trig rate with a gate width of 100 ns Voltage Settings: Manually adjusted to allow for a 40 Hz singles rate

  5. Objective 2: To evaluate the cosmic ray flux (фCR) as a function of time over a period of 15 hours.

  6. Flux Events/area/ time Average value: 8651 events/m2/min Area of paddles: .075 m2 Calculated Average Flux: фCR = 10.8 muons/s Bin width = 60 seconds

  7. Error Analysis Absolute uncertainty One standard deviation from the mean √N = √8651 = 93 relative uncertainty = 1.08% Maximum deviations 8053 deviates by 598 (6.9%) 9253 deviates by 602 (7.0%)

  8. Discussion and Conclusions The deviations of muon detection per minute are greater than an expected random sampling of data. This deviation seems to confirm that there is not a uniform bombardment of cosmic rays. Further calibration of the voltage settings would be beneficial