40 likes | 96 Views
Follow a detailed procedure involving shielded syringes, TSB, incubation vials, and simulated Tc-99m for quality testing of radiopharmaceuticals. Includes multiple steps for accurate testing and monitoring.
E N D
Media Fill Procedure 3 mL Shielded syringe 10 mL of TSB 1 mL 20 or 30 mL shielded evac vial (simulatedTc-99m) X 4 10 mL unshielded syringe (repeat 3 more times for a total of 4 vials) Simulated Kit vial #1 Source Vial of TSB 3 mL Draw 0.4 mL of media from Kit (simulated normal saline)
Media Fill Procedure (Cont’d) Kit # 1 Kit #2 Kit #3 Kit #4 Draw 0.4 mL of media from Kit 3 mL shielded syringes Repeat 3 times for a total of 4 vials and 24 syringes 10 or 20 mL incubation vials
1 2 3 4 Media Fill Procedure (Cont’d) • Incubate 4 new vials at 25-35 Degrees Celsius in a controlled environment for 14 days.
1 Media Fill Procedure (Overview) 5 mL Shielded syringe 10 mL of TSB 1 mL 20 or 30 mL shielded evac vial Tc-99m 10 mL unshielded syringe X 4 Vial of TSB (repeat 4 times) Simulated Kit vial #1 3 mL Draw 0.4 mL of media from Kit 3 or 5 mL shielded syringes Repeat process 4 times for a total of 24 incubation vials 10 or 20 mL incubation vials Incubate vials at 25-35 Degrees Celsius