Electro-optical Spectro-polarimeter for Ground- and Space-based Solar Coronagraphy

Electro-optical Spectro-polarimeter for Ground- and Space-based Solar Coronagraphy S. Fineschi, INAF-Osservatorio Astronomico di Torino, Italy. Proba-3: ESA Formation-flying space coronagraph LC Tunable-filter Polarimeter (LCTP) for the “green line” LCTP Lab performances evaluation

stefan
Download Presentation

Electro-optical Spectro-polarimeter for Ground- and Space-based Solar Coronagraphy

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Electro-optical Spectro-polarimeter for Ground- and Space-based Solar Coronagraphy S. Fineschi, INAF-Osservatorio Astronomico di Torino, Italy May 9th, 2012 - MP1104 COST Action 1st WG meeting, Warsaw

  2. Proba-3: ESA Formation-flying space coronagraph LC Tunable-filter Polarimeter (LCTP) for the “green line” LCTP Lab performances evaluation Eclipse 2010 observations with LCTP prototye Case for ground-based emission-line coronagraphy in Europe OUTLINE

  3. ESA PROBA-3 (ESA Startiger: 2009-10 Phase-A) Formation-Flying Coronagraph A coronagraph with the external occulter on one spacecraft and the optical instrument on the other spacecraft at ~150 meters from the first one. Proba-3 – ESA: launch 2017. ASPIICS: 2009 selection; 2010 Phase-A (Startiger); 2012 Phase-B

  4. ASPIICS LC Lyot filter

  5. Liquid-crystal Tunable Lyot Filter for Solar Coronagraphy

  6. LC Tunable-filter & Polarimeter Procurement

  7. LCTP Perfomances

  8. Monochr. slit width= 25 µm; height = 6 mm Monochr. spectrum= CW 530.3 nm; FWHM = 0.1nm Collim. lens (L1) f = 140 mm Camera lens (L2) f = 1130 mm CCD pixel sixe = 25 Spatial scale = 5 arcsec/pix (ASPIICS: 3 arcsec/pix) SET-UP @ TURIN AstronomicalObs. LAB

  9. LCTP SPECTRAL tests

  10. 45° 90° 135°

  11. LC SPACE VALIDATION (rad-hardness) • Proba-3 rad-hard requirement • (Summary Mission Requirements: ESA P3-EST-TN-7002) • 1996 –  1.5 Mrad of < 32 MeV e- (Graham et al., SPIE 2811) • 1998 –  12 krad of  20 MeV protons (Noto et al., Opt. Eng.) • 2007 –  50 krad of -rad; Thermal-vac; Vibe tests • (Heredero et al., Appl. Opt. 46) • 2009-2010 –  50 krad 100 MeV p+, 10 MeV e- • ESA-ITT “LC Space Validation for SolO”

  12. The CorMag was operated during the total solar eclipse of July, 11th 2010 on Tatakoto Atoll (French Polynesia)

  13. LCTP basedspectropolarimeter–cormag

  14. Coronal Temperature contourlevels:Blue: 1.5106 Kgreen: 2106 K K-corona image by M. Drukmuller

  15. HAO CoMP LC Lyotfilter & Polarimeter FeXIII 1074.7 nm S. Tomczyk, et al. Science 317, 1192 (2007);

  16. COMP observations. HAO model . S. Gibson courtesy, 2012

  17. CoMP-S instrument @ Lomnicky Peak Observatory (SLOVAK R.)

  18. LCTP FILTER @ CLIMSO CORONAGRAPH PIc du midi (france)?

  19. LCTP FILTER & ASPIICS telescope @ DOMEc French-italianconcordia base (antartica)? Sky Brightness Monitor @ Concordia J. Arnaud, M. Faroubert, 2009 ESCAPE proposal L. Dame, 2011

  20. Thanks !!!

  21. Liquid-crystal Tunable Lyot Filter for Solar Coronagraphy The LCTP achieves all the spectroscopy performances expected from the F-P etalon The LCTP has additional science capabilities (e.g., polarimetry, simultaneous spectroscopy & high-res imaging) ESA completed a LC space validation program The LCTP operational advantages: no mechs, some science preserved even in “fail state”, OPS monitoring capability also during coronal images acquisition 21 Aug., 2011 - Solar Physics and Space Weather Instrumentation IV – San Diego , CA

  22. SET-UP @ OATo LAB

  23. exit slit 0 -d 0 0 +d Monochromator setting Center wavel.: 530.3 nm FWHM: 0.1 nm

  24. LCTP vs. F-P etalonSpectroscopy 21 Aug., 2011 - Solar Physics and Space Weather Instrumentation IV – San Diego , CA

  25. LC Tunable-filter

  26. Liquid-crystal Tunable Lyot Filter for Solar Coronagraphy LCTP and F-P Characteristics 21 Aug., 2011 - Solar Physics and Space Weather Instrumentation IV – San Diego , CA

  27. LCTP basedSPECTRopolarimeter - cormag 21 Aug., 2011 - Solar Physics and Space Weather Instrumentation IV – San Diego , CA

  28. LCTP basedspectropolarimeter–cormag 21 Aug., 2011 - Solar Physics and Space Weather Instrumentation IV – San Diego , CA