An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Status of Tev IPMsJune 10, 2009 Andreas Jansson

  2. News • Tried turning off select ion pumps • Signals strength improved somewhat • Signal fluctuations still there • Took vertical tbt data at injection • No obvious mismatch, detailed analysis pending • Found almost 1V ripple on 3.3V supply to HIPM timing card • Replaced power supply • Timing jitter now essentially gone IPM status

  3. VIPM IPM status

  4. HIPM (w new power supply) IPM status

  5. Shutdown plans • Before beam disappears, do reference measurements • In particular with horizontal system • Bake ion pumps • Hopefully get rid of N2 saturation • Run tests with internal calibration source • May give a hint on the cause of signal strength fluctuations IPM status