1 / 15

Detailed Measurement of Interface Shapes for Static and Dynamic Contact Angles

Detailed Measurement of Interface Shapes for Static and Dynamic Contact Angles. Geometry Optics Data analysis Extracting contact angle and surface tension Recommendations: when, where. Main students doing the technique development: John A. Marsh Qun Chen Kroum Stoev. Geometry. D T.

tadeo
Download Presentation

Detailed Measurement of Interface Shapes for Static and Dynamic Contact Angles

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Detailed Measurement of Interface Shapes for Static and Dynamic Contact Angles • Geometry • Optics • Data analysis • Extracting contact angle and surface tension • Recommendations: when, where... Main students doing the technique development: John A. Marsh Qun Chen Kroum Stoev

  2. Geometry DT Highest point on line of sight Cap Length • Tube Diameter: DT = 2.5 cm • DT >> Cap Length (1.5 mm): • Azimuthal curvature small effect on shape and flow • Easy to focus, sharp meniscus seen on meridian plane(unlike flat plate) • Unlike spreading drop, outer length scale very large • Cylinder: No "end effects"

  3. Optics: Kohler Illumination Condenser Focal plane • Image of light source forms at condenser aperture

  4. Optics: Kohler Illumination • Image of source aperture forms at object plane • Condenser aperture: controls cone angle • Source aperture: controls illuminated spot size • Uniform illumination key to making physical edge parallel to equi-intensity contour Result: uniformly illuminated, in-focus image of source aperture

  5. Imaging System Schematic

  6. Image Quality Highest curvature in line of sight: Sharpest Image • Uniformly flat surface: • Lowest curvature along line of sight • Fuzzier image • 80° ≤ Contact angles ≤ 100°: Can’t measure because contact line hidden • Usually meniscus edge sharp out to >1.4 mm

  7. Image Quality - 2 • Interfaces meeting at the contact line: • Diffraction patterns interfere& cause distortion • Contact angle < 5°: • No problem • Can get interface all the way through contact line • Larger contact angles: • Safely down to 15µm to 20µm from contact line • Best conditions can get closer

  8. Menisci in Depression • Can be measured • Light path through liquid • PIV possible • Edge finder output: interface slope vs. position • Slope: One derivative closer to curvature than x-y data • Question: do "equal intensity levels" follow physical edge? • A: Calibration

  9. Calibration • Needed due to small distortions near edges • Mechanical shapes (e.g., straight edge) not good enough • How straight is the edge? • Use static capillary shape: • Known exact theoretical form: Young-Laplace Eq. • Use Static Contact Angle and Surface Tension as fitting parameter • Two-parameter fit: contact angle & surface tension uncoupled • Difference (Data-Fit): • No systematic deviation from zero • Strict criterion imposed – cloud of data does not move more than 1/3 width off zero line

  10. Fitting Details • Fitting AWAY from contact line crucial • Why • All surfaces have contact angle hysteresis • With hysteresis comes contact line brokenness • ...which leads to interface shape fluctuations • Fluctuations die out: scale larger than contact line waviness! • Need to fit beyond folding to get “contact angle” & surface tension • Global contact angle: boundary condition for meniscus beyond folds

  11. Analysis • We fit theoretical models to the interface data • Young-Laplace (static theory) • Cox-Dussan composite asymptotics (Newtonian, viscous theory) • Extract: • Static contact angle & surface tension from fit to Young-Laplace • "Dynamic" apparent contact angle from fit to Cox-Dussan • Requirements • (Best fit - Exptal data) free of systematic deviation

  12. Accuracy • Data cloud ~2° thick (but ~1° RMS) • Contact angle accuracy ~1°or less • Mostly Run-to-Run variation • Good accuracy due to calibration with static shape • High precision in local interface angle from fitting to large number of data points to determine one interface angle • Very accurate (~0.25deg) measurement of interface shape

  13. Recommendations • Kohler illumination less important than uniform illumination • Good resolution from 15µm to 1500µm from contact line • Perhaps not strictly necessary for static unless detailed shape needed (i.e., could use "2-point" for statics...) • Necessary when detailed interface shapes needed • Necessary for dynamic contact angle

  14. Back Ups

  15. Optics: Kohler Illumination • Settings: • Source aperture: just large enough to illuminate entire field of view • Larger condenser aperture: more fuzzy, less contrast, more depth of focus • Smaller condenser aperture: more contrast, more diffraction fringing around contact line • Cylindrical geometry requires not-too-large depth of focus Result: uniformly illuminated, in-focus image of source aperture