An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Lab Experience with HIV RNA NAAT Myra Brinson, MT(ASCP) Manager, Virology/Serology North Carolina State Laboratory of Public Health Ph: 919-807-8835 E-mail: Myra.Brinson@ncmail.net

  2. Discussion Topics • History of HIV RNA NAAT at NC State Laboratory • GenProbe APTIMA HIV Method Verification • NC State Lab HIV Test Algorithm

  3. Evolution of NCSLPH HIV NAAT: 2001 to 2008 • Utilization of different assays for the detection of HIV-1 RNA • Different pooling algorithms • Different pooling mechanisms • Consistently demonstrated the ability to detect acute HIV-1 infections

  4. Pilot Study 2001 - Design • All consecutive routine HIV tests submitted to the NC State Laboratory of Public Health over 4 weeks from 110 publicly funded counseling and testing sites (CTS) [n=8505] • Initial Ab testing - OT Vironostika HIV-1 Viral Lysate Microelisa (State Lab) • Manual pooling of Ab NR samples (State Lab) • Roche Amplicor HIV-1 Monitor (UNC) – Standard and US

  5. Pilot Study 2001 - Results • Acute infection: 5 per 10,000 • Chronic infection: 44 per 10,000 • Overall specificity: 99.99% • Estimated additional cost per specimen: $2.01 • Estimated total testing costs/additional case diagnosed: $4109 Pilcher CD et al, JAMA, Vol. 288/No. 2, July 10, 2002

  6. Screening and Tracing Active Transmission (STAT) Program – Year 1 • 11/01/2002 to 10/31/2003 - All consecutive routine HIV tests submitted to the State Laboratory from 110 publicly funded counseling and testing sites (CTS) [n=110,890] • Initial Ab testing - bioMerieux Vironostika HIV-1 Viral Lysate Microelisa • Initial manual pooling • NAAT testing – bioMerieux NucliSens HIV-1 Qualitative assay • Automated pooling added-Beckman Coulter bioMek FX

  7. STAT – Pooling Design

  8. STAT Results – Year 1 Pilcher CD et al, New England Journal of Medicine, May 5 2005;352(18):1873-1883.

  9. Screening and Tracing Active Transmission (STAT) Program – Year 2 • Nov 03 to March 05 - All consecutive routine HIV tests submitted to the State Laboratory from 110 publicly funded counseling and testing sites (CTS) [n=118,656] • Initial Ab testing - bioMerieux Vironostika HIV-1 Viral Lysate Microelisa • NAAT testing – GenProbe Procleix HIV-1 Assay • Automated pooling – Hamilton AT Plus • Detected 17 acute HIV cases (1 False Positive)

  10. STAT Overall Results - 2 years • 224,108 EIA negative sera pooled and tested for HIV-1 RNA • 40 True Positive Acute Infections; 3 False Positive RNA tests • Acute infection: 1.8 per 10,000 • Chronic infection: 65.9 per 10,000 • Overall specificity: 99.8% • At least 4% of the HIV-1 infected patients would have been undetected without the use of NAATs

  11. Manual pooling very labor intensive – 30 to 45 minutes/90 samples Must be particularly diligent in pipetting technique to avoid cross contamination of samples 3 of 4 false positives occurred during the period of manual pooling Automated pooling instrumentation, maintenance, and pipet tips can be expensive Much more efficient – 10-15 minutes/90 samples Improved control of process – decreased human error Positive specimen identification - barcodes Manual vs. Automated Pooling

  12. STAT Project Continues • GenProbe Procleix HIV-1 Assay withdrawn from market due to patent dispute with Chiron • March 05 – Nov 07 bioMerieux Nuclisens Easy Q HIV-1 NAAT • EasyMag Automated Extraction/ Real-Time assay • Initial Ab testing - bioMerieux Vironostika HIV-1 Viral Lysate Microelisa • Automated pooling - Hamilton AT Plus • Continued to detect acute HIV cases, although assay sensitivity was somewhat of concern

  13. 2008 – Changes to HIV Testing CATALYSTS • bioMerieux Vironostika HIV-1 Viral Lysate Microelisa discontinued • Contract for HIV NAAT out for bid – opportunity to bring on a more sensitive assay • New LIMS

  14. Increase in HIV Antibody Screens 2001-2007

  15. New HIV Antibody Assay • Bio-Rad Genetic Systems HIV-1/HIV-2 Plus O EIA • Automation - 3 Evolis instruments • More sensitive assay – detects both IgM and IgG • Ability to detect HIV-2 and Group O

  16. New NAAT Assay and Pooling Algorithm • GenProbe APTIMA HIV-1 NAAT RNA assay • Hamilton STARlet robotic pipetting instrument • Reduced pool size (80 samples/pool) • Increased sensitivity for HIV-1 NAAT

  17. New Pooling Strategy

  18. NAAT Questions??? • Cost of NAAT: $37-$50/test, based upon 4,000/year test volume • Two dedicated staff for pooling and NAAT testing – pool daily and test 2-3 times/week • Currently pool all EIA negative samples vs. separating into low and high risk groups • NAAT has also been useful for resolution of EIA reactive/WB negative or indeterminate samples • Increased TAT – 2-3 days to 2-3 weeks

  19. GenProbe APTIMA HIV-1 NAAT Method Verification • Off-label use of an FDA-approved assay * Plasma vs. serum * Waterbaths vs.SB100s • CAP Molecular Verification Checklist *Accuracy *Precision *Reportable Range *Limit of Detection *Analytical Sensitivity *Analytical Specificity *Specificity/Cross Reactivity

  20. Accuracy • 18 Nonreactive and 7 Reactive B Pools previously tested by current method (bioMerieux Nuclisens EasyQ) • 17 of 18 known NR samples tested NR by APTIMA • 7 of 7 known R samples tested R by APTIMA • One discrepant sample QNS to resolve • 96% (24/25)

  21. Precision-Reproducibility • Tested several replicates of pooled NR sera and a R sera (170 copies /ml) over multiple days • 170 copies/ml- prepared by diluting a sample of known copy number from a purchased linearity panel (HIV RNA Linearity Panel, PRD801, BBI Diagnostics) in pooled NR sera • NR Intra-Assay 49% CV; Inter-Assay 51% CV • R Intra-Assay 4% CV; Inter-assay R: 8% CV

  22. Reportable Range • 2,900,000 (2.9 x 106) copy/ml sample from the purchased linearity panel • Serially diluted in pooled NR sera to yield approx. log concentrations of 1 x 106,105,104,103, 102,and10 copies/ml • Observed typical reaction curve for molecular assays • Precipitous drop-off in signal strength from 100 to 10 copies/ml

  23. Serially diluted the 170 copies/ml sample from the purchased linearity panel in pooled NR sera to yield samples of approx. concentration of 85, 43, 21, 11, 5, and 3 copies/ml Tested five replicates of the seven dilutions 100% at ≥43 copies/ml HIV-1 RNA 80% at 5 to 21 copies/ml HIV-1 RNA Limit of Detection

  24. Analytical Sensitivity/Specificity • Calculated by using accuracy sample results • Analytical Sensitivity: 100% 7 of 7 known R samples tested R by APTIMA • Analytical Specificity: 96% 17 of 18 known NR samples tested NR by APTIMA

  25. Specificity/Cross Reactivity • Spiked pooled NR sera with other blood-borne viral pathogens that might be present in tested patient population: HSV-1, HSV-2, CMV, HAV, HBV, and HCV • Required the purchase of HBV viral isolate from ATCC; other viruses obtained locally • Virus concentrations tested were similar to average concentrations of HIV virus expected to be present in patient serum samples • Samples were run twice, on consecutive days • No cross-reactivity or interference with the six blood-borne viral pathogens

  26. Evaluation Questions??? • Duration: Verification required approx. six weeks to complete • Cost: Test kits provided by GenProbe BBI Linearity Panel - $1,380 ATCC HBV - $188 • Availability of NR and R comparison samples - in our study, we were able to use master pool samples run previously by existing assay - may have to purchase additional panels

  27. HIV-1, HIV-2, Plus O EIA NR R HIV-1 RNA NAAT Repeat EIA x 2 R x 2 Neg NR x 2 Pos R x 1 NR x 1 Report #1 HIV-1 Western Blot CURRENT NC HIV TEST ALGORITHM HIV-1 (Groups M & O) and HIV-2 antibodies were not detected. HIV-1 RNA not detected.

  28. HIV-1, HIV-2, Plus O EIA NR R HIV-1 RNA NAAT Repeat EIA x 2 R x 2 Neg NR x 2 Pos R x 1 NR x 1 HIV-1 Western Blot Report #2 CURRENT NC HIV TEST ALGORITHM HIV-1 RNA detected. Possible Acute HIV Infection. Please consult with Disease Intervention Specialist to arrange for further HIV-1 testing. Recommend clinical evaluation with Infectious Disease Specialist.

  29. HIV-1, HIV-2, Plus O EIA NR R HIV-1 RNA NAAT Repeat EIA x 2 R x 2 Neg NR x 2 Pos R x 1 NR x 1 Report #1 HIV-1 Western Blot Report #2 CURRENT NC HIV TEST ALGORITHM

  30. HIV-1, HIV-2, Plus O EIA NR R HIV-1 RNA NAAT Repeat EIA x 2 R x 2 Neg NR x 2 Pos* R x 1 NR x 1 Report #1 HIV-1 Western Blot Report #2 CURRENT NC HIV TEST ALGORITHM Red, Green, or Blue Dot Samples* *For individually tested samples, repeat any Pos result: If retest is Pos = Report Pos If retest is Neg, repeat again: 2nd retest Neg = Report Neg 2nd retest Pos = Report Pos

  31. Repeatedly Reactive HIV-1, HIV-2, Plus O EIA HIV-1 Western Blot R NR Indeterminate Report #3 HIV-1 RNA NAAT Pos Neg CURRENT NC HIV TEST ALGORITHM HIV-1 (Groups M & O) Antibody was detected. Please consult with Disease Intervention Specialist to determine if further testing is warranted to rule out possible co-infection with HIV-2, based upon epidemiological information.

  32. Repeatedly Reactive HIV-1, HIV-2, Plus O EIA HIV-1 Western Blot R NR Indeterminate Report #3 HIV-1 RNA NAAT CDC for HIV-2 WB or Rapid, by DIS request Pos Neg CURRENT NC HIV TEST ALGORITHM HIV-1 (Groups M & O) Antibody was detected. Please consult with Disease Intervention Specialist to determine if further testing is warranted to rule out possible co-infection with HIV-2, based upon epidemiological information.

  33. Repeatedly Reactive HIV-1, HIV-2, Plus O EIA HIV-1 Western Blot R NR Indeterminate HIV-1 RNA NAAT Pos Neg Report #4 CURRENT NC HIV TEST ALGORITHM HIV-1 RNA detected. Possible Acute HIV Infection with incomplete seroconversion. Please consult with Disease Intervention Specialist to arrange for further HIV-1 testing. Recommend clinical evaluation with Infectious Disease Specialist.

  34. Repeatedly Reactive HIV-1, HIV-2, Plus O EIA HIV-1 Western Blot R NR Indeterminate HIV-1 RNA NAAT Pos Neg Report #5 CURRENT NC HIV TEST ALGORITHM HIV-1 RNA detected. Probable Acute HIV Infection with incomplete seroconversion. Please consult with Disease Intervention Specialist to arrange for further HIV-1 testing. Recommend clinical evaluation with Infectious Disease Specialist.

  35. Repeatedly Reactive HIV-1, HIV-2, Plus O EIA HIV-1 Western Blot R NR Indeterminate HIV-1 RNA NAAT Pos* Neg Report #4 or #5 CURRENT NC HIV TEST ALGORITHM *For individually tested samples, repeat any Pos result: If retest is Pos = Report Pos If retest is Neg, repeat again: 2nd retest Neg = Report Neg 2nd retest Pos = Report Pos

  36. Repeatedly Reactive HIV-1, HIV-2, Plus O EIA HIV-1 Western Blot R NR Indeterminate Report #3 HIV-1 RNA NAAT CDC for HIV-2 WB or Rapid, by DIS request Pos Neg Report #4 or #5 CURRENT NC HIV TEST ALGORITHM

  37. Repeatedly Reactive HIV-1, HIV-2, Plus O EIA HIV-1 Western Blot NR or Ind HIV-1 RNA NAAT Neg HIV-1/HIV-2 Rapid NR R HIV-2 Report #6 or #7 CURRENT NC HIV TEST ALGORITHM HIV-1/HIV-2 infection status is Inconclusive. Please submit another sample for further HIV-1/HIV-2 testing.

  38. Repeatedly Reactive HIV-1, HIV-2, Plus O EIA HIV-1 Western Blot NR or Ind HIV-1 RNA NAAT Neg HIV-1/HIV-2 Rapid NR CDC for HIV-2 WB Confirmation R HIV-2 Report #8 or #9 CURRENT NC HIV TEST ALGORITHM HIV-1 (Groups M or O) was not detected. HIV-2 infection status is inconclusive. Specimen referred to CDC for further HIV-2 testing.