730 likes | 2.7k Views
Objectives. Begin discussing factors that influence image detail or visibility of detailSpatial
E N D
1. Control of Scatter Radiation Bushong Ch. 14
2. Objectives Begin discussing factors that influence image detail or visibility of detail
Spatial & Contrast resolution
Radiographic Noise
Scatter Radiation
Ways to reduce scatter & improve image quality
Primary beam restriction & Grids
Technique adjustments when using grids
3. What are some factors that increase scatter radiation?
4. 3 factors contribute to an increase in scatter Increased kVp
Increased x-ray field size
Increased patient thickness
5. 2 principal characteristics of any image are Spatial & Contrast Resolution Spatial resolution
Resolution is the ability to image two separate objects and visually distinguish one from the other
Spatial resolution is the ability to image small objects that have high subject contrast (eg. bone-soft tissue interface, calcified lung nodules)
Determined by focal-spot size and other factors that contribute to blur
Diagnostic x-ray has excellent spatial resolution. It is measured in line pairs per mm. (CT measured in cm)
7. SMPTE Test Pattern
9. Image-forming x-rays Two kinds of photons are responsible for the OD and contrast on an image: Photons that pass through without interacting and those that are scattered though Compton.
X-rays that exit the patient are remnant and those that interact with the IR are image-forming.
10. Ideally, only those x-rays that do not interact with the patient should reach the IR
. However, scatter radiation is a factor that must be managed
Proper collimation has the PRIMARY effect of reducing patient dose by _________ ?
Proper collimation also improved image contrast by reducing radiographic noise or fog caused by scatter
11. Fog or Noise
13. kVp As x-ray energy increases Photoelectric and Compton interactions decrease. Explain?
At 50 kVp 79% photoelectric, 21% Compton & less than 1% transmission
At 80 kVp 46% photoelectric, 52% Compton & 2% transmission
Pg. 225
14. How does increasing kVp affect patient dose?
15. Patient Thickness Imaging thick parts of the body results in more scatter radiation than thin parts
16. Is patient thickness something the radiographer can control?
17. Patient thickness Normally, No
Compression devices improves spatial resolution by reducing patient thickness and bringing the object closer to the IR. Compression also reduces patient dose and contrast resolution
18. Compression Improves spatial resolution
Reduces OID
Reduces patient dose
Improves contrast resolution (reducing fog or noise)
19. Compression
20. Field Size As field size increases, intensity of scatter radiation also increases rapidly. Especially during fluoroscopy
21. Compare images: What do you think about radiographic contrast & image noise?
22. Control of Scatter Radiation Technologists routinely use two types of devices to reduce the amount of scatter radiation reaching the IR
Beam restrictors
Grids
23. 3 Types of beam-restricting devices Aperture Diaphragm
Cones or Cylinders
Variable aperture collimator
24. Variable Aperture Collimator The most common beam-restricting device is the light-localizing variable aperture collimator
The first part of the collimator serves to control off-focus radiation. What is off-focus radiation?
25. Off - focus Radiation X-ray tubes are designed so that the projectile e- interacts with the target. However, some of the e- bounce off the target and land on other areas
This caused x-rays to be produced out side the focal spot
26. Extrafocal Radiation These rays can also be called off-focus radiation
Extrafocal radiation is undesirable because it extends the size of the focal spot, increases patient skin dose & reduces image contrast
27. Off-focus radiation
28. Fixed diaphragm in the tube housing Using a grid
does not reduce
extrafocal
radiation
29. First-stage entrance shuttering device Has multiple collimator blades protruding from the top of the collimator into the tube housing
30. The collimator lamp & mirror Must be adjusted so that the projected light field coincides with the x-ray beam
Misalignment of the light field and beam can result in collimator cutoff of anatomic structures
31. Total Filtration Filtration review
Total Filtration = _______________ + _____________?
The collimator assembly is usually equivalent to approximately _______ mm Al filtration.
Minimum filtration for tubes that can operate about 70 kVp is _______ mm Al or equiv.
32. The Grid
33. Contrast & Contrast Resolution Two devices are used to reduce Compton effect beam-restricting devices and radiographic grids
Beam-restricting devices effects what reaches the patient. Grids effect the remnant beam
34. Contrast & Contrast Resolution Contrast = the comparison of areas of light, dark and shades of gray on the image
Contrast Resolution = the ability to image adjacent similar tissues
35. Grid Ratio High-ratio grids are more effective in cleaning up scatter radiation than low-ratio grids
The angle of deviation is smaller for high-ratio grids. (the photon must be traveling in a straighter line to make it through the grid)
However, the higher the ratio the more radiation exposure necessary to get a sufficient number of x-rays through the grid to the IR
36. Grid Ratio 3 important dimensions on a grid: The thickness of the grid strips, the width of the interspace material, and the height of the grid
The grid ratio is the HEIGHT of the grid divided by the INTERSPACE WIDTH:
Grid ratio = h
D
37. h = height of the grid, T = thickness of the grid strip, D = width of the interspace material
38. The higher the ratio the straighter the photon must travel to reach the IR Grid ratios range
from 5:1 to 16:1
Most common
8:1 to 10:1
A 5:1 grid will
clean up 85%
16:1 clean up 97%
39. Grid Frequency The number of grid strips or grid lines per inch or centimeter
The higher the frequency the more strips and less interspace material and the higher the grid ratio
As grid frequency increases, patient does is increase because more scatter will be absorbed
40. Grid Frequency Some grids reduce the thickness of the strips to reduce the exposure to the patient, this over all reduces the grid clean up
Grids have frequencies in the range of 25 to 45 lines per centimeter (60 to 110 lines per inch)
41. Higher frequency with the same interspace distance reduces the grid effectiveness
42. Grid Performance The principal function of a grid is to improve image contrast
Contrast Improvement Factor (k) = the ratio of the contrast of a radiograph made with a grid to the contrast of the radiograph made without a grid. A contrast improvement factor of 1 indicates no improvements
The higher the grid ratio & frequency the higher the k
43. Bucky Factor Using grids require more patient dose. Why is this?
When a grid is used technique must be increased to maintain OD
The amount of increase is given by the Bucky factor (B) or grid factor
44. Bucky Factor or grid factor The higher the grid ratio or frequency the higher the bucky factor
The Bucky factor increases with increasing kVp
Pg 235: We will use the average values for calculations.
45. Selectivity or ability to clean upthe heavier the grid the more Pb it contains
46. Grid Types Parallel Grid simplest type of grid
All the lead strips are parallel
Only clean up scatter in one direction (along the axis of the grid)
Easy to make, however can cause grid cutoff with short SIDs.
47. Grid cutoff Distance to cutoff
SID
Grid ratio
With decreasing
SID more potential
for grid cutoff
IR size will also
Influence grid cutoff
48. Grid Cutoff Parallel grid
49. Crossed Grid Have lead strips running along the long and short axes of the grid
Made by placing two parallel grid on top of each other
50. Crossed Grid
Have twice the grid ratio as linear
grids
However, CR vs grid placement is critical. The CR must align with the center of the grid and the grid and CR must be exactly parallel or grid cutoff will occur
51. Focused Grid Designed to minimize grid cutoff
Lead strips are aligned with the divergence of the x-ray beam
Each focused grid must be identified with the appropriate SID
Wrong SID = Grid cutoff
52. Focused grid have a little SID latitude (eg. 100cm grid could be used at 90cm 110cm)
53. Moving Grids All stationary grids will give you grid lines on your radiograph. Thinner Pb strips will give you less noticeable lines. However, thinner strips have less Pb content not cleaning up as well
Grid Lines are made when primary x-rays are absorbed in the grid strips.
54. Focused grids are usually used as moving grids The grid is placed in a holding mechanism that begins moving just before the x-ray exposure and continues moving after the exposure ends
2 types of movement Reciprocating & Oscillating
55. Grid Motion Reciprocating = moves several times about 2cm back and forth during the exposure
Oscillating = moves several times about 2 3 cm in a circular pattern
Most grids are moving. Except for portable imaging
56. Grid Problems Increased OID, especially with moving grids
The biggest problem with grids is misalignment
57. Grid Problems Off Level
58. Grid Problems Off Center
A problem with focused & crossed grids
59. Grid Problems Off Focus (wrong SID)
60. Grid Problems Upside-Down
A problem with focused & crossed grids
61. Grid Selection Patient Dose
Pg 241 mAs changes
Exam
Detail required
Part thickness
Desired technique (kVp)
Equipment availability
62. Questions
.?