1 / 61

Control of Scatter Radiation

Objectives. Begin discussing factors that influence image detail or visibility of detailSpatial

tevin
Download Presentation

Control of Scatter Radiation

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


    1. Control of Scatter Radiation Bushong Ch. 14

    2. Objectives Begin discussing factors that influence image detail or visibility of detail Spatial & Contrast resolution Radiographic Noise Scatter Radiation Ways to reduce scatter & improve image quality Primary beam restriction & Grids Technique adjustments when using grids

    3. What are some factors that increase scatter radiation?

    4. 3 factors contribute to an increase in scatter Increased kVp Increased x-ray field size Increased patient thickness

    5. 2 principal characteristics of any image are Spatial & Contrast Resolution Spatial resolution Resolution is the ability to image two separate objects and visually distinguish one from the other Spatial resolution is the ability to image small objects that have high subject contrast (eg. bone-soft tissue interface, calcified lung nodules) Determined by focal-spot size and other factors that contribute to blur Diagnostic x-ray has excellent spatial resolution. It is measured in line pairs per mm. (CT measured in cm)

    7. SMPTE Test Pattern

    9. Image-forming x-rays Two kinds of photons are responsible for the OD and contrast on an image: Photons that pass through without interacting and those that are scattered though Compton. X-rays that exit the patient are remnant and those that interact with the IR are image-forming.

    10. Ideally, only those x-rays that do not interact with the patient should reach the IR…. However, scatter radiation is a factor that must be managed Proper collimation has the PRIMARY effect of reducing patient dose by _________ ? Proper collimation also improved image contrast by reducing radiographic noise or fog caused by scatter

    11. Fog or Noise

    13. kVp As x-ray energy increases Photoelectric and Compton interactions decrease. Explain? At 50 kVp 79% photoelectric, 21% Compton & less than 1% transmission At 80 kVp 46% photoelectric, 52% Compton & 2% transmission Pg. 225

    14. How does increasing kVp affect patient dose?

    15. Patient Thickness Imaging thick parts of the body results in more scatter radiation than thin parts

    16. Is patient thickness something the radiographer can control?

    17. Patient thickness Normally, No Compression devices improves spatial resolution by reducing patient thickness and bringing the object closer to the IR. Compression also reduces patient dose and contrast resolution

    18. Compression Improves spatial resolution Reduces OID Reduces patient dose Improves contrast resolution (reducing fog or noise)

    19. Compression

    20. Field Size As field size increases, intensity of scatter radiation also increases rapidly. Especially during fluoroscopy

    21. Compare images: What do you think about radiographic contrast & image noise?

    22. Control of Scatter Radiation Technologists routinely use two types of devices to reduce the amount of scatter radiation reaching the IR Beam restrictors Grids

    23. 3 Types of beam-restricting devices Aperture Diaphragm Cones or Cylinders Variable aperture collimator

    24. Variable Aperture Collimator The most common beam-restricting device is the light-localizing variable aperture collimator The first part of the collimator serves to control off-focus radiation. What is off-focus radiation?

    25. Off - focus Radiation X-ray tubes are designed so that the projectile e- interacts with the target. However, some of the e- bounce off the target and land on other areas This caused x-rays to be produced out side the focal spot

    26. Extrafocal Radiation These rays can also be called off-focus radiation Extrafocal radiation is undesirable because it extends the size of the focal spot, increases patient skin dose & reduces image contrast

    27. Off-focus radiation

    28. Fixed diaphragm in the tube housing Using a grid does not reduce extrafocal radiation

    29. First-stage entrance shuttering device Has multiple collimator blades protruding from the top of the collimator into the tube housing

    30. The collimator lamp & mirror Must be adjusted so that the projected light field coincides with the x-ray beam Misalignment of the light field and beam can result in collimator cutoff of anatomic structures

    31. Total Filtration Filtration review… Total Filtration = _______________ + _____________? The collimator assembly is usually equivalent to approximately _______ mm Al filtration. Minimum filtration for tubes that can operate about 70 kVp is _______ mm Al or equiv.

    32. The Grid

    33. Contrast & Contrast Resolution Two devices are used to reduce Compton effect beam-restricting devices and radiographic grids Beam-restricting devices effects what reaches the patient. Grids effect the remnant beam

    34. Contrast & Contrast Resolution Contrast = the comparison of areas of light, dark and shades of gray on the image Contrast Resolution = the ability to image adjacent similar tissues

    35. Grid Ratio High-ratio grids are more effective in cleaning up scatter radiation than low-ratio grids The angle of deviation is smaller for high-ratio grids. (the photon must be traveling in a straighter line to make it through the grid) However, the higher the ratio the more radiation exposure necessary to get a sufficient number of x-rays through the grid to the IR

    36. Grid Ratio 3 important dimensions on a grid: The thickness of the grid strips, the width of the interspace material, and the height of the grid The grid ratio is the HEIGHT of the grid divided by the INTERSPACE WIDTH: Grid ratio = h D

    37. h = height of the grid, T = thickness of the grid strip, D = width of the interspace material

    38. The higher the ratio the straighter the photon must travel to reach the IR Grid ratios range from 5:1 to 16:1 Most common 8:1 to 10:1 A 5:1 grid will clean up 85% 16:1 clean up 97%

    39. Grid Frequency The number of grid strips or grid lines per inch or centimeter The higher the frequency the more strips and less interspace material and the higher the grid ratio As grid frequency increases, patient does is increase because more scatter will be absorbed

    40. Grid Frequency Some grids reduce the thickness of the strips to reduce the exposure to the patient, this over all reduces the grid clean up Grids have frequencies in the range of 25 to 45 lines per centimeter (60 to 110 lines per inch)

    41. Higher frequency with the same interspace distance reduces the grid effectiveness

    42. Grid Performance The principal function of a grid is to improve image contrast Contrast Improvement Factor (k) = the ratio of the contrast of a radiograph made with a grid to the contrast of the radiograph made without a grid. A contrast improvement factor of 1 indicates no improvements The higher the grid ratio & frequency the higher the k

    43. Bucky Factor Using grids require more patient dose. Why is this? When a grid is used technique must be increased to maintain OD The amount of increase is given by the Bucky factor (B) or grid factor

    44. Bucky Factor or grid factor The higher the grid ratio or frequency the higher the bucky factor The Bucky factor increases with increasing kVp Pg 235: We will use the average values for calculations.

    45. Selectivity or ability to “clean up” the heavier the grid the more Pb it contains

    46. Grid Types Parallel Grid – simplest type of grid All the lead strips are parallel Only clean up scatter in one direction (along the axis of the grid) Easy to make, however can cause grid cutoff with short SID’s.

    47. Grid cutoff Distance to cutoff SID Grid ratio With decreasing SID more potential for grid cutoff IR size will also Influence grid cutoff

    48. Grid Cutoff – Parallel grid

    49. Crossed Grid Have lead strips running along the long and short axes of the grid Made by placing two parallel grid on top of each other

    50. Crossed Grid Have twice the grid ratio as linear grids However, CR vs grid placement is critical. The CR must align with the center of the grid and the grid and CR must be exactly parallel or grid cutoff will occur

    51. Focused Grid Designed to minimize grid cutoff Lead strips are aligned with the divergence of the x-ray beam Each focused grid must be identified with the appropriate SID Wrong SID = Grid cutoff

    52. Focused grid have a little SID latitude (eg. 100cm grid could be used at 90cm – 110cm)

    53. Moving Grids All stationary grids will give you grid lines on your radiograph. Thinner Pb strips will give you less noticeable lines. However, thinner strips have less Pb content not “cleaning up” as well Grid Lines are made when primary x-rays are absorbed in the grid strips.

    54. Focused grids are usually used as moving grids The grid is placed in a holding mechanism that begins moving just before the x-ray exposure and continues moving after the exposure ends 2 types of movement Reciprocating & Oscillating

    55. Grid Motion Reciprocating = moves several times about 2cm back and forth during the exposure Oscillating = moves several times about 2 – 3 cm in a circular pattern Most grids are moving. Except for portable imaging

    56. Grid Problems Increased OID, especially with moving grids The biggest problem with grids is misalignment

    57. Grid Problems – Off Level

    58. Grid Problems – Off Center A problem with focused & crossed grids

    59. Grid Problems – Off Focus (wrong SID)

    60. Grid Problems – Upside-Down A problem with focused & crossed grids

    61. Grid Selection Patient Dose Pg 241 – mAs changes Exam Detail required Part thickness Desired technique (kVp) Equipment availability

    62. Questions….?

More Related