An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Preventing Adverse Effects from Medical Devices A Saudi Food & Drug Administration Program January 2008 Tim Ritter, Senior Project Engineer

  2. “Intensive Care” • Credit to the artist – Yan Nascimbene • Illustration for an Annals of Medicine article in New Yorker magazine (10.12.07) newyorker.com/reporting/2007/12/10/071210fa_fact_gawande/ • The Checklist by Atul Gawande

  3. Intensive Care is Extremely Complicated • Dr. Gawande concludes: Delivery of intensive care has become so complicated that checklists should be developed and followed for each procedure or “action” • He presents a study that determined the average number of actions ICU patients require per 24 hr

  4. Actions include • Suctioning a patient • Responding to fibrillation • Programming an infusion pump • How many actions per day??? • . . . . .178

  5. How many errors occurred? • Remarkably, nurses and doctors were observed to make an error in just 1% of 178 actions • However . . . that is still 2 errors/day with every patient! • Unknown ─ the percentage of errors that produced an adverse effect

  6. Preventing Adverse Effects from Medical Devices • Today, in this workshop, we will look at several technologies and case studies to learn how some of those errors and adverse effects might be prevented • We want to look at failures and errors to identify • equipment that needs repair or modification • supplies and accessories that are defective • the need for user training • problems that should be reported to SFDA

  7. Preventing Adverse Effects from Medical Devices Session 1 (45 min) Introduction Tubing and cable misconnections Surgical fires / oxygen-enriched environments* Session 2 (60 min) Reducing infusion pump errors – “smart pumps”** Defibrillators Hospital beds *Printed with Laser Safety handouts **On SFDA MDS Web site

  8. Preventing Adverse Effects from Medical Devices Session 3 (60 min) Electrosurgical/surgical diathermy units Imaging equipment Laser safety Questions and discussion

  9. Any questions or concerns about: • Aspirators or suction regulators? [2 points] • Hospital electrical or medical gas/vacuum systems? • A technology that is not listed on the program? • How many nurses work in: • Intensive care? Emergency care? • Operating theater? Med/surg? • Emergency care? Pediatrics? • What technology is associated with the most problems/hazards???

  10. Top 10 Health Technology HazardsListed alphabetically • Alarm hazards • Burns during electrosurgery/surgical diathermy • Burns during magnetic resonance imaging • Caster failures • Infusion pump programming errors [Number 1] Health Devices – November 2007

  11. Top 10 Health Technology Hazardscontinued • Misconnections of blood pressure monitors to intravenous lines • Needlesticks and other sharps injuries • Radiation dose in computed tomography • Radiation therapy errors • Surgical fires

  12. Preventing Misconnections of Lines and Cables A Saudi Food & Drug Administration Program January 2008 Tim Ritter Senior Project Engineer

  13. Misconnections – Why do they occur? • Patients are connected to an increasing number of medical devices – with so many lines and cables, it’s not obvious which components should be connected • If it can happen it will – many medical devices which should not be connected share a common type of connector (e.g. Luer) or can be connected by using an adapter

  14. Misconnection Awareness is Needed • Misconnections are preventable errors – when misconnections occur they often result in severe injury or fatalities • The frequency of misconnections in healthcare is unknown – many misconnections are not reported, especially the “near misses” recognized and corrected before patient injury occurs

  15. Classification of Misconnections • Three types of lines in the patient area • tubes carrying liquids • tubes carrying gas • cables • Generally, misconnections involving gas and liquid lines are more severe because they involve direct bodily access routes

  16. Misconnection Examples • Cable Misconnections • cable plugged into the wrong jack • cable plugged into the correct jack in the wrong orientation • cable not fully inserted • Outcome: at best a device that does not operate; in some cases, patient burn, equipment damage, electrocution; damage to plug and/or jack

  17. Misconnection Examples • Gas misconnections • misconnecting a ventilator breathing circuit • delivering the wrong gas (e.g. N2O instead of Oxygen) • Outcome: serious injury, death

  18. Misconnection Examples • Liquid Misconnections • delivering a substance through the wrong delivery route (e.g. enteral solution delivered intravenously, IV medication delivered epidurally) • Outcome: serious injury (sepsis) or death

  19. Standards • Few design standards and regulations have addressed known misconnection problems • FDA requires electrode lead wires be protected to prevent connection to AC power sources • Currently almost no standards address the potential for Luer connector misconnections – CEN and AAMI standards are being developed which may restrict the use of Luer connectors to certain types of devices • Exception: ANSI/AAMI ID54 – requires enteral feeding set connectors and associated adapters not be compatible with Luer connectors to reduce the risk of an intravenous connection

  20. Opportunities for Misconnection • Misconnecting devices outfitted with Luer connectors – many functionally incompatible devices share this common connector • Breathing circuit misconnections – often facilitated by adapters • Spiking pre-filled enteral feeding bags or sterile water bags with IV administration sets – even though the bags are labeled “not for IV use”, they’re often mistaken for IV fluids • Misconnecting gas or vacuum supply lines

  21. What Can Hospitals Do? • Purchase well-designed devices • Develop a multidisciplinary “task force” to perform a hospital-specific misconnection risk assessment – this team should continue to track misconnection reports to identify areas needing improvement and report to SFDA • Alter policies and procedures to reduce misconnections

  22. Device Design • Most effective strategy – make it physically impossible to make a misconnection • Locking connectors are preferred – they prevent disconnections and subsequent mistakes upon reconnecting lines • Good design incorporates more than one type of equipment design strategy • However, color coding or labeling are ineffective strategies by themselves – they rely on users to recognize and use prompts

  23. Connector Design Examples • Good design – the 3 jacks look similar but have unique pin configurations which only accept the correct connector; jacks are also labeled • An even better design – the plug will only physically mate with one of the jacks; colors and shapes also serve as user prompts

  24. Specific Purchasing Tips • Where possible, avoid buying non-IV equipment (nebulizers, pneumatic compression devices, NIBPs) that can mate with Luer connectors • Only purchase enteral feeding sets which do not mate with Luer connectors • Avoid purchasing sterile water and enteral feeding solutions in pre-filled bags which can be mis-spiked with IV tubing sets

  25. Work Practices for Clinicians • Trace all lines back to their origins before making connections • Don’t force connections – if it takes a lot of effort to make a connection, chances are it’s the wrong one • Don’t use adapters unless they are clearly required for the application • Make sure you’ve been trained to use a device and its accessories – if unsure, ask for help • Report all misconnections – even near-misses – to SFDA

  26. Policy Level Prevention • Encourage reporting of misconnections and ensure these reports are being tracked for trends and problem areas • Provide periodic training about misconnection prevention to all personnel • Conduct pre-purchase evaluations so nurses and clinical engineers can “play” with devices and identify misconnection hazards before devices are acquired • Inspection/maintenance procedures should look for unlabelled jacks or worn connectors

  27. For More Information • This slide show summarizes information presented in the March 2006 issue of Health Devices • Guidance Article: Preventing Misconnections of Lines and Cables • Questions? Stephanie Joseph Senior Project Engineer ECRI Institute sjoseph@ecri.org