A Rapid Analysis of Actinides In Water Samples Using Eichrom’s Resin Cartridges

EIChroM. I N D U S T R I E S , I N C. A Rapid Analysis of Actinides In Water Samples Using Eichrom’s Resin Cartridges. Anil H.Thakkar Eichrom Industries, Inc.. Darien, IL. EIChroM. I N D U S T R I E S , I N C. Objective.

tim
Download Presentation

A Rapid Analysis of Actinides In Water Samples Using Eichrom’s Resin Cartridges

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. EIChroM I N D U S T R I E S , I N C. A Rapid Analysis of Actinides In Water Samples Using Eichrom’s Resin Cartridges Anil H.Thakkar Eichrom Industries, Inc.. Darien, IL

  2. EIChroM I N D U S T R I E S , I N C. Objective • To perform analysis for Am, Pu and U in water samples using Eichrom’s new chromatographic configuration of UTEVA and TRU Resin

  3. EIChroM I N D U S T R I E S , I N C. Current Eichrom Method • UTEVA and TRU Resin Column • Industry Standard • Flow Rate Limitation

  4. EIChroM I N D U S T R I E S , I N C. UTEVA TRU Strip Strip U Pu Am Tandem Column Arrangement Load Discard

  5. EIChroM I N D U S T R I E S , I N C. Eichrom UTEVA Resin Column Separation Load: 3M HNO3, 1.0M Al(NO3)3 to TRU Resin column 0.01M HCl U

  6. EIChroM I N D U S T R I E S , I N C. Eichrom TRU Resin Column Separation Load: 3 M HNO3, 1.0M Al(NO3)3 to waste 9m HCl & 4M HCl Am 0.1M ammonium bioxalate Pu

  7. EIChroM I N D U S T R I E S , I N C. Decontamination Factors (DF) • To determine the ruggedness of the method • To determine the degree of separation of the analyte from potentially interfering matrix components

  8. Experiment for DF • Approximately 100 pCi (3.7 Bq) of each actinide was added to a load solution and run through the method. • DF in uranium, plutonium and americium fractions were calculated.

  9. EIChroM I N D U S T R I E S , I N C. Equation for Decontamination Factor DF = Ao Af Where; Ao = Original activity of the analyte added Af = Final activity of analyte in the fraction

  10. EIChroM I N D U S T R I E S , I N C. Criteria for Acceptance • For actinides, a DF value of > 1000 is acceptable • Tracer Recovery: 70%-105%

  11. EIChroM I N D U S T R I E S , I N C. Decontamination Factors

  12. EIChroM I N D U S T R I E S , I N C. New Configuration:Resin Cartridge • Packed with 0.7 gram of resin • Cartridges with Luer lock • Ease of tandem arrangement for sequential analysis

  13. EIChroM I N D U S T R I E S , I N C. Testing the procedure with Cartridges • 1 liter aliquots of sample were spiked with Am, Pu, and U

  14. EIChroM I N D U S T R I E S , I N C. Am/Cm, Pu and U Analyses Calcium phosphate precipitation Dissolve in load solution Eichrom U/TEVA Resin U Eichrom TRU Resin Am/Cm, Pu Preparation for counting Alpha spectrometry

  15. EIChroM I N D U S T R I E S , I N C. Sample Preparation • l liter aliquots of acidified water samples were spiked with Am, Pu and U • Samples were heated to near boiling for at least 3 hours • A Ca3(PO4)2 precipitation was performed • Decant supernatant, centrifuge and collect the precipitate

  16. EIChroM I N D U S T R I E S , I N C. Load Solution for UTEVA and TRU Resin Columns • Dissolve precipitate in 3M HNO3/1M Al(NO3)3 solution • Add ferrous sulfamate and ascorbic acid • Load on UTEVA and TRU Resin cartridge in tandem

  17. EIChroM I N D U S T R I E S , I N C. Flow rate • Typically 2-3 mL/minute • Achievable with vacuum pump

  18. Recoveries for Direct Spikes

  19. Recoveries in DI Water

  20. Recoveries in Simulated Hanford Ground Water

  21. EIChroM I N D U S T R I E S , I N C. Summary • Eichrom Resin cartridges provide fast separation for actinides • Good, reproducible and consistent recoveries achievable • Rapid turn-around possible

  22. EIChroM I N D U S T R I E S , I N C. Future Objectives • Evaluate effect of flow rate on chemical recovery • To test other matrices such as; urine, soil and vegetation for actinides using Eichrom Resin cartridges • To evaluate separation of Sr using Eichrom Sr Resin cartridges