320 likes | 559 Views
QUALITY PATIENT CARE THROUGH PROFICIENCY TESTING. Not Just Talking Quality, But Investing in Quality Naval Medical Center San Diego Laboratory Department Quality Assurance Unit. What I really need to know besides my “A,B,Cs”?. QA QC POCT PT. OBJECTIVES.
E N D
QUALITY PATIENT CARE THROUGH PROFICIENCY TESTING Not Just Talking Quality, But Investing in QualityNaval Medical Center San DiegoLaboratory DepartmentQuality Assurance Unit
What I really need to know besides my “A,B,Cs”? QA QC POCT PT
OBJECTIVES • Discuss regulatory aspects and impact of proficiency testing (PT) • Approaches to effectively manage and fully utilize PT • List investigative techniques used to determine causes of poor PT performance and describe effective corrective actions • Understand how “Linearity” relate to the accreditation process • Explain techniques to perform the appropriate verifications of AMR and CRR • Common Analytical Exception codes
CAP Proficiency Testing • Allows laboratories to evaluate their performance and improve the accuracy of the patient results • Results evaluated using comparable peer groups from comprehensive database of laboratories
PROFICIENCY TESTING DO’s • Samples are tested on all shifts by technicians who routinely test patient samples • All surveys performed are treated as real patient samples. This includes using the same test methods for quality control, patient testing, workload accounting, verification and reporting of results. • NOTE: This only applies to analytes that are tested in-house. • Personnel who routinely analyze patient samples will integrate all PT within the routine workload
Proficiency Testing DON’Ts • Test samples differently than patient specimens • Have the same people always do testing • Refer samples to another laboratory • Discuss results before reporting • Use PT sample for other purposes before submission date deadline
ENROLLMENT • Required for all analytes listed in the PT Enrollment Guide - Updated annually - Available at CAP website www.cap.org -Included with order renewal forms
Why perform Linearity and Calibration Verification??? • Requirements: CAP Laboratory Accreditation Program Checklist Question CHM.13400 (one of several relating to Calibration Verification) and Current CLIA Regulations Section 493.1255 for most analytes. • Linearity evaluations give information about what’s occurring “near the extremes," the high and low ends of the measurement range. • LN Surveys are particularly useful in calibration verification – identify imprecision problems before they become noticeable in other ways. • Calibration Verification provides proof that the measurement system calibration has not changed since the last calibration event.
Why does AMR need to be verified? • Example: typical serum glucose method • Low calibrator = reagent blank (0 mg/dL) • High calibrator = 356 mg/dL • AMR = 10-600 mg/dL • The AMR is based on the expectation that the signal is linearly related to glucose concentration over the 10-600 mg/dL range. • Verification of the AMR expectation is required because the calibrators do not define the complete measurement range.
Clinically Reportable Range • Extension of the AMR based on clinical requirements. • Established once when method is implemented • Does not need to be verified on a recurring schedule • Values outside the CRR are reported as < or > a numeric value. • Checklist question: Are dilution protocols and diluents (or concentration protocols) specified for all methods for which the CRR exceeds the AMR?
AMR: how close to low and high limits • To verify acceptable clinical performance • For low limit: use clinical judgment, e.g. • creatinine, 0.2 or 0.3 mg/dL • For high limit: within 10-15% of upper limit of AMR
Reviewing your LN Evaluation Report Example of LN2 Report (Calcium)
Example of a Non-Linear Result Loss of recovery at the high end of the AMR -Check reagents, calibrators, QC -Are the results clinically significant?
PARTICIPATION • In addition to enrolling and testing samples, you must: -Ensure results are sent either electronically or via fax -Ensure they are sent by the deadline -Ensure all faxed pages are received NOTE: Non-participation (failure to receive results)= UNSATISFACTORY
SATISFACTORY PT PERFORMANCE • 80% for regulated analytes and other analytes that have at least 5 challenges per PT event • Percentage varies for analytes with fewer than 5 challenges
CLERICAL ERRORS • Post Analytic Phase • Same importance as testing errors • Examples: • Transcription • Method/Reagent/Instrument codes • Missing Information (TNP, </>, etc)
TECHNICAL ISSUES • Directly attributable to human actions: • Reconstitution • Specimen Mix-up • Improper Specimen Handling • Incorrect Instrument Handling • Failure to follow testing kit instructions • Morphologic Misinterpretations
Methodologic Issues • Mechanical difficulties • Instrument Software Problems • Frequency of CALIBRATION • Inadequate Reagent Performance • Inadequate Maintenance • Other instrument malfunctions
Exception Codes ANALYTICAL PROBLEM (Leave the result area blank and fill the bubble for the appropriate Exception Code). • 11 Unable to analyze (documentation be provided by laboratory). • 22 Result is outside method/ instrument reportable range. • 33 Specimen determined to be unsatisfactory after contacting the CAP. *It is the laboratory’s responsibility to document the appropriate use of these exception codes should this be requested during a laboratory inspection.
Additional Tips • Keep original worksheets and instrument print-outs with the PT file • Make sure every less than acceptable result is followed up and documented • Confirm that corrective actions make sense and actually fix the problem
UNSATISFACTORY PT PERFORMANCE • Laboratories that have unsatisfactory performance for 1 or more analytes on an event will receive PTES Exception Summary (PTES) report. • Investigate the problem • Determine cause • Implement Corrective Action • RESPONSE TO LAP is not required
UNSUCCESSFUL PT PERFORMANCE • Is unsatisfactory performance on 2 of 3 PT events • Laboratory will receive a PT Exception Summary (PTES) report • Laboratory must suspend testing or implement plan of corrective action • LAP must approve laboratory’s action
CORRECTIVE ACTIONS • Design a process to verify clerical entries prior to result submission • Retrain testing personnel as to the proper procedures for sample preparation, testing and reporting • Modify/Narrow QC acceptance range to detect problems sooner • Evaluate/increase frequency of calibration • Perform Instrument Function Verification (Precision, Linearity, Correlation) • Revise Procedure to reflect corrective actions
Example of Unsuccessful Proficiency Testing Performance:IgA FAILED on two out of 3 events. What will be your course of action: • Continue testing and provide CAP with credible program of corrective action • Voluntarily cease patient testing and investigate the failure • Conduct thorough investigation
IgA Unity Laboratory Comparison Report <2.0 SDI/CVR
IgA Levy Jennings POOR CALIBRATION
Identified Cause of Failure: Poor Calibration Corrective Measures to prevent reoccurrence: • Daily review of calibration and QC • More frequent monitoring of LJ charts to catch trends and shifts that could indicate calibration curve issue • Method validation of IgA method was performed Precision Linearity Correlation Studies
CONCLUSION: Proficiency Testing monitors laboratories' continuing performance in a way that: • –Improves patient care • –Protects the public’s health
References: www.cap.org