High-precision efficiency calibration of a HP Germanium detector

High-precision efficiency calibration of a HP Germanium detector. Bertram Blank CEN Bordeaux-Gradignan. main physics goals detector scans source measurements MC simulations. ISOLDE workshop, December 5-7, 2011. Ft = ft (1 + d ’ R ) (1 + d NS - d C )

varen
Download Presentation

High-precision efficiency calibration of a HP Germanium detector

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. High-precision efficiency calibration of a HP Germanium detector Bertram Blank CEN Bordeaux-Gradignan • main physics goals • detector scans • source measurements • MC simulations ISOLDE workshop, December 5-7, 2011

  2. Ft = ft (1 + d’R) (1 + dNS- dC) = for T=1 K 2 G2v (1+DvR) Super-allowed Fermi transitions: Introduction Determination of experimental values for 0+  0+:  measurement of T1/2, BR, QEC vector coupling constant Gv(CVC hypothesis) Vudelement of CKM quark matrix Precision required: < 10-3 Ft = 3073.660.75 s • Future subjects: • - experimental tests of isospin corrections dc • - test of CVC on largerrange • - measurementswith more exoticisotopes • Measurements for A > 54 • Measurements for Tz ≤ -1

  3. Super-allowed Fermi transitions: Ca-38 ISOLDE 2007 Half-life of 38Ca: beampreparation withREXtrap T1/2 = 443.8 ± 1.9 ms B. Blank et al., EPJA 44 (2010) 363 Kavanagh et al. Zioni et al. Presentwork Gallman et al. Wilson et al. BR = 75.45(58)%

  4. Super-allowed Fermi transitions for Tz = 0 • basically 100% super-allowed decay • very small non-analog decays • precision of e.g. 10% for a 1% transition:  0.1%

  5. Super-allowed Fermi transitions for Tz = -1 • many decay channels open • strong non-analog transitions • high precision of g efficiency needed  0.1%

  6. 70% n-type Germanium detector • thickdead zone on interiorside • thinexternaldead layer • request of high-precision for • crystal size • crystal position • dead zones • large dewar • aluminum entrance window

  7. Calibration Procedure • X-ray radiography • g-ray detector scans • source measurements • MC simulations • (GEANT4 or CYLTRAN) •  develop a model of the detector • to calculateefficiencyatanyenergy • at a fixed distance of 15 cm

  8. X-ray photography of detector • rough size of crystal • tilt of crystalwith respect to detector housing of 1° • according to GEANT4 simulations no influence on results

  9. Gamma-ray scan of detector • AGATA scan table at CSNSM: strongly collaminated 137Cs source HPGe 137Cs (477MBq) X-Y table A. Korichi et al.

  10. Gamma-ray scan of detector • AGATA scan table at CSNSM: strongly collaminated 137Cs source (662 keV) HPGe 137Cs (477MBq) X-Y table

  11. Gamma-ray scan of detector • Scan at CENBG: strongly collaminated 241Am source (60 keV)

  12. Longitudinal scan: 662 keV total full-energy peak • excellent full-energy peak spectrum • good total-energy spectrum • problem with thickness of entrance window?

  13. Front scan: 662 keV total full-energy peak • tilt…. • full-energy peak: excellent • total spectrum: raisonable • precision of the stepping motors: 10-2

  14. 70% HP Germanium on precision test bench Source position high-precision X-Y-Z table • all source measurements at exactly 15 cm from entrance window •   position precision of better than 0.1%

  15. Pile-up correction • Fixed 137Cs and movable 241Am: 6ms of shaping time

  16. Comparison: experiment - simulations • To develop detector model: • which code to use: GEANT4, CYLTRAN, EGGS…. • GEANT4: (J. Souin, PhDstudent…. hequit….) • almost all featuresneeded • errors in decayschemes externaleventgenerator • no g-gcorrelations (1% effect)  externaleventgenerator • not extremelywelltested for low-energyg rays • different packages, which one to use? • CYLTRAN: (BB) • verywelltested J. Hardy • FORTRAN77… • only single g rays or electrons in the input •  upgrade to « source-type » input via an externaleventgenerator • positron annihilation-in-flight not included •  to beincluded; significanteffect on escape peaks • which code to use: GEANT4, CYLTRAN, EGGS….

  17. Comparison: experiment - simulations • twoways: • correct experimental data for summingeffects to get single g-rayefficiencies • certain number of approximations • « collimated » simulations, openingcone: •  muchlessevents to simulate • peak-to-total canbeincluded a posteriori •  J. Hardy et al. • to simulatewholedecayschemes • no approximations needed as long as decayschemeisknown • need of eventgeneratorincludingg-gangularcorrelations • peak-to-total needed in simulations • need of a lot of simulatedevents to getstatistics for all g rays!! • B. Blank et al. • presentlyonly CYLTRAN simulations

  18. Calibration sources • peak-to-total sources: •  one single g ray with 100% branching ratio • 57Co, 51Cr, 85Sr, 137Cs, 58Co, 54Mn, 60Co, 22Na • relative efficiency sources: •  a few well-known branches (BR error <1%) atlargelydifferentenergies • standard sources: • 60Co, 88Y, 133Ba, 134Cs, 152Eu, 207Bi • short-lived online source at ISOLDE: • 24Na, 27Mg, 56Co, 66Ga, 75Se • one absoluteefficiency source: • 60Co withactivityprecision of 0.7‰

  19. Total - to - peak 57Co 51Cr 85Sr 54Mn 60Co

  20. Comparison of source measurements and simulations

  21. Comparison of source measurements and simulations +133Ba, 56Co

  22. Detector model ! !

  23. To do list..... • rescan front with241Am • check entrance windowthickness • measure entrance window to cristal distance • test pile-up correction for differentenergies and shaping times • improve sources 133Ba and 56Co • addsources 66Ga, 75Se, 134Cs, 152Eu • measureother sources: • peak-to-total: 139Ce, 7Be, 65Zn, 28Al, 37S • efficiency: 46Sc, 59Fe, 108mAg, 120Sb, 180mHf, 66Ga • add annihilation in flight to CYLTRAN • compare with GEANT4 simulations • analyse for single g-rayefficiency  detector model with 0.1-0.2% efficiencyprecision

  24. Acknowledgement • J. Souin, PhDstudent • A. Korichiand Hoa Ha, CSNSM • A. Herlert and K. Johnston, ISOLDE • J. Hardy, Texas A & M • ExoticNuclei group at CENBG Born to bewild….. … or what???? Thanks for your attention!

  25. Longitudinal scan: 60 keV full-energy peak • nice over-all agreement • holding structure reasonably well reproduced

  26. Front scan: 60 keV full-energy peak • slight difference… • there was a problem during the scan…

  27. Perpendicular scan: 662 keV total full-energy peak • good agreement for full-energypeak • reasonableagreement for total spectrum

  28. Perpendicular scan: 60 keV full-energy peak • does not help a lot… • effect of holding structure very important

  29. 0+ T1/2 QEC BR g 0+ Super-allowed Fermi transitions: experimental device b detector Tape transport system Ge detectors 38Ca ISOLDE beam from REXtrap

  30. DAQ: deadtime correction • fixed 137Cs source and pulser on other channel

  31. 137Cs: conversion electrons Total - to - peak 57Co 51Cr 85Sr 137Cs 54Mn 60Co

More Related