Status

Status. DETECTORS: Baseline for 2mm: NICA 5a (144 pixels, IRAM, 30nm sputtering); average sensitivity: 6mK/Hz 0.5 (1Hz) corresponding roughly to NEP  6 · 10 -16 W/Hz 0.5 (1Hz). Before de-correlation. Tested one array made in Trento/Roma for 2mm. Found about 12mK/Hz 0.5 (1Hz).

Download Presentation

Status

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Status • DETECTORS: • Baseline for 2mm: NICA 5a (144 pixels, IRAM, 30nm sputtering); average sensitivity: • 6mK/Hz0.5 (1Hz) corresponding roughly to NEP  6·10-16 W/Hz0.5 (1Hz). Before de-correlation. • Tested one array made in Trento/Roma for 2mm. Found about 12mK/Hz0.5 (1Hz). • Still working to improve the 2mm IRAM-Néel 2mm array (e.g. cross-talk, thickness, HF, film) • Best pixels tested had 2.7mK/Hz0.5 (1Hz). Before de-correlation. • - Baseline for 1.25mm: SRON (IRAM, 30nm sputtering); average sensitivity: 50mK/Hz0.5 (1Hz) • NEP  8·10-15 W/Hz0.5 (1Hz). PRELIMINARY. Imaging OK. Resonances OK. • Tested one LEKID array made in Trento/Roma for 1.25mm. Found about 30mK/Hz0.5 (1Hz) but • some problems with resonances (coupling not adapted). • As agreed, the priority is for the SRON array if the Roma/Trento one isn’t 2.5 times better in • Sensitivity. Improvements ongoing on both sides. Time until first week of October for testing. • OPTICS: • Full dual-band tested last week. • Polarizer not flat at 300K but imaging seems OK (full system test 25/08/2010) • Still waiting one last HPE filter with the good diameter (should arrive) • Background seems lower with the new optics (40K vs. 70K equivalent). • ELECTRONICS: • - ROACH board 112 channels version working (up to 220MHz band) (DDC program) • ROACH board FFT program is also working (Loren) • LPSC electronics available and to e tested in September (128 channels over 125MHz band) • Andrey working on another electronics (FFTS, 1GHz band)

  2. Summary • WE THINK A SHORT TECHNICAL RUN (5x12h) AT THE 30-m IS JUSTIFIED BY : • - IMPROVEMENT on 2mm BAND sensitivity (factor estimated in 24) • - NEW 1.25mm BAND (simultaneous !) • - LARGER NUMBER OF PIXELS (> 3 times more than last year) • IF ACCEPTED WE WOULD NEED : • 8 dewars of helium (we’re carrying 2 more from Grenoble) • more information on receiver cabin dimensions • to check again that the Ethernet cable from the cabin is OK • ………………….

  3. 3-6dB deep  coupling OK Problem. 2-3 resonances lost. NICA 5a Process : IRAM (sputtering), 30nm, HR silicon, 300um thick. Design : NICA 5a

  4. NICA 5a BEAM FWHM = 28mm S/N on the « planet » = 3000 Hz0.5 • Assuming T=20K (200K of 8mm • diameter planet diluted over 28mm • beam), means • 20K/3000  7mK/Hz0.5 •  consistent with Sky simulator estimation

  5. OPTICAL CONFIGURATIONS TELESCOPE FOCAL PLANE TELESCOPE LABORATORY 620 mm

  6. For 6mK difference T on focal plane we calculate (assuming 0.35 instrument Transmission and 130-165GHz band) 5.5·10-16 W difference per pixel (2.25x2.25mm) DECORRELATION VERY IMPORTANT.

  7. 132 pixels, 2mm array. Al box (could not be a good idea) Resonances spacing : 2MHz or less SRON 256 pixels array. Resonances Spacing : 4MHz