The Photon Spectrometer for RHIC and beyond

20×20× 200 mm 3 produced by Furukawa co. T.Sugitate. Y.Furuhashi. K.Homma. R.Kohara. The Photon Spectrometer for RHIC and beyond. PbWO 4 Crystal Density 8.29 g/cm 3 Radiation length 0.89 cm Moliere radius 2.2 cm. Presented by Toru Sugitate Hiroshima University.

Download Presentation

The Photon Spectrometer for RHIC and beyond

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 20×20× 200 mm3 produced by Furukawa co. T.Sugitate Y.Furuhashi K.Homma R.Kohara T. Sugitate / Hiroshima / PHX031 / Nov.01 The Photon Spectrometer for RHIC and beyond PbWO4 Crystal Density 8.29 g/cm3 Radiation length 0.89 cm Moliere radius 2.2 cm Presented by Toru Sugitate Hiroshima University

  2. T. Sugitate / Hiroshima / PHX031 / Nov.01 Motivation of the project • Motivation • Photon is an unique signal to explore the QGP matter. • higher granularity in a large multiplicity environment. • higher energy resolution for better PID capabilities. • PWO crystal is a good candidate for the EMCal upgrade. Scope of the work • Choice of the crystal; Apatity, Shanghai, Japan. • Optimize the optical readout device; PMT, PIN, APD. • Design and fabricate a FEE prototype. • Prototyping of an array and testing with beams. • Physics measurements in a RHIC experiment. Time profile and costs • To be discussed in the workshop.

  3. T. Sugitate / Hiroshima / PHX031 / Nov.01 A PWO signal for a MeV photon PMT Hamamatsu R7056 single photon sensitive (107 @1900V) PWO crystal (Covered with Aluminum) 60Co -source PMT Hamamatsu R2083 plastic scintillator BC-404

  4. T. Sugitate / Hiroshima / PHX031 / Nov.01 PWO light yield and decay time constant Pulse height distribution Scintillation decay curve Sample (B) single p.e. peak double p.e. peak 60Co- photo peak Time spectrum [ns]

  5. T. Sugitate / Hiroshima / PHX031 / Nov.01 A beam test with 150 MeV electron at the Hiroshima VBL A 150 MeV electron microtron and a storage ring

  6. Geant4 simulation of a 150 MeV electron injection in a crystal Single electron peak Ee = 116 MeV Deposit energy 77 % Energy resolution 16 % The simulation does not include; photon and p.e. statistics, backgrounds, PMT and electronics noises, and the beam energy fluctuation. double electron peak Ee = 231 MeV Ee = 117 MeV /E = 13 % T. Sugitate / Hiroshima / PHX031 / Nov.01 Crystal response to the 150 MeV electrons

  7. T. Sugitate / Hiroshima / PHX031 / Nov.01 A Monte Carlo simulation of the PWO photon spectrometer GEANT4 Simulation 10 0 ’s in the acceptance (dN/d~76) d/dpexp(-p/0.3) (1<p<5 GeV/c) flat distribution in y and  π0 → γγ The Photon Spectrometer coverage: 1m ×1 m crystal size: 20×20×200 mm3 array size: 50 × 50 distance from IP: 3 m  coverage: 0.17

  8. The invariant mass spectrum shows the mass reconstruction resolution of 3.2% A sophisticated clustering algorithm may considerably improve the mass resolution . Invariant Mass [MeV/c2] T. Sugitate / Hiroshima / PHX031 / Nov.01 Invariant mass spectrum Deposit Energy vs. Hit Position at the PWO Spectrometer

  9. T. Sugitate / Hiroshima / PHX031 / Nov.01 Snapshots of APD readout from the crystal Blue-enhanced APD Hamamatsu S5345 active area: 5mm  Photon signal for 137Cs source Charge amplifier Hamamatsu H4083 Gain: 0.5V/coulomb

  10. T. Sugitate / Hiroshima / PHX031 / Nov.01 Summary and outlook of the “Hiroshima” project Summary • A crystal produced by Furukawa was tested: • Transmission; 50%@350nm and 80%@400nm. • Light yield; 4-8 p.e./MeV. • Decay constant; 1=1.2-1.5ns (16-23%), 2= 5.0-5.6ns (72-81%), and 3= 21.9-38.2ns (3-5%). • Light output in 50 (100) ns; 94.9 (98.6)%. • Energy deposit in 20x20x200mm: 77% consistent with Geant4 • A MC simulation code was setup. • APD readout underway. Outlook in 1.5 years • Build a mini-tower (3x3) in this Dec. • Mini-tower tests with PMT at Hiroshima and at KEK-PS. • Improve the clustering algorithm in the MC code. • Continue to study the APD readout. • Evaluation of crystals from Apatity and Shanghai. • Mini-tower tests with APD readout in the next JFY. • Optimization of APD device.