250 likes | 390 Views
Automated Chip QC. Michael Elashoff. Chip QC. Transition from mostly manual/visual chip QC to mostly automated chip QC Database of passing and failing chips to serve as the training set (5K passing, 2K failing). Chip QC: Defect Classes. In order of occurrence: Dimness High Background
E N D
Automated Chip QC Michael Elashoff
Chip QC • Transition from mostly manual/visual chip QC to mostly automated chip QC • Database of passing and failing chips to serve as the training set (5K passing, 2K failing)
Chip QC: Defect Classes • In order of occurrence: • Dimness • High Background • Unevenness • Spots • Haze Band • Scratches • Brightness • Crop Circle • Cracked • Snow • Grid Misalignment • Training set of 7K chips (Human, Rat, Mouse)
Dimness/Brightness A chip Low Scan Passing Chips Bright/Dim Chips
Dimness/Brightness A chip Low Scan Passing Chips Bright/Dim Chips
Dimness/Brightness • Each chip type has a different typical brightness range • Typical brightness range depends on scanner setting • tuned-up versus tuned-down • scanners must be calibrated to achieve consistency
Implementation of Li-Wong • With training set of 5K passing chips, apply Li-Wong algorithm • For each probe set, algorithm yields: • “outlier” status for each probe-pair • probe weights for non-outlier probe-pairs
Implementation of Li-Wong • For QC, new chips are screened individually • For each probe set: • Ignore “model outlier” probes • Using training ‘s, compute • Compute residuals for each probe pair • Flag residuals that are large
Implementation of Li-Wong • Compare distributions of outlier count for passing and failing chips in training set • Determine upper bound of acceptable outlier count:
Limitations of Li-Wong • Must estimate 1.8 million probe weights for human/rat chip sets • Works poorly for rare genes • Probe weights may vary • Tissue Type • RNA Processing • Chip Lot • Training Set
Using Spike-Ins Spike-in R2 must be >96.5%
QC Metrics • Mean of Non-control Oligo Intensity • Mean OligoB2 Intensity • Spike-in R2 • Li-Wong Outlier Count • Several measures of LiWong Outlier “clustering” • Vertical profiles • Horizontal profiles • Thresholds differ for each chip type
QC Metrics: Performance Two week validation run False Negative Rate = 0.4% These will not be manually QC’d anymore False Positive Rate = 46.8% These are still manually QC’d
Conclusions • Automated QC has: • reduced the number of chips in visual QC • made the process more objective • Automated QC has not: • eliminated the need for visual QC • incorporated the impact on real world data quality/analysis
Thanks • Peter Lauren • Chris Alvares • John Klein • Michelle Nation • Jeff Wiser