Radiation hardness test of PD’s and APD’s, Energy resolution of CsI/PD and CsI/APD. ACCULINNA Group, FLNR JINR, Dubna Russia (http://aculina.jinr.ru) & Lund University, Sweden. Si PIN characteristics : Face size – 10.5 by 11.5 mm 10 by 10 mm Resistance – 8 k W 4 k W

zoltin
Download Presentation

ACCULINNA Group, FLNR JINR, Dubna Russia (aculina.jinr.ru) & Lund University, Sweden

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Radiation hardness test of PD’s and APD’s, Energy resolution of CsI/PD and CsI/APD ACCULINNA Group, FLNR JINR, Dubna Russia (http://aculina.jinr.ru)&Lund University, Sweden Si PIN characteristics : Face size – 10.5 by 11.5 mm 10 by 10 mm Resistance – 8 kW 4 kW Front layer materials: RTV, Epotex301, Bicron, Naked Radiation flux: Light charged particles (E=1-25 AMeV) 2×1010 1/cm2 Fast and slow neutrons 5×1011 1/cm2

  2. earlier warning on effectiveness of full energy deposition for CsI Test of CsI detection response for 179 MeV tritons 2 Si strip 61x61x1mm3 detectors followed by 25x25x50mm3 CsI

  3. Conditions: Iinitial (U=50V) = 2-3 nA DE/E=2.5% (5.8 MeV) Air, T = 22o 1o C Intensity: <104 pps/cm2 Time: ~ 60 days Dark current values (nA) of PD’s and ΔE/E (%) of CsI/PD’safter an integrated flux of ~2×1010 LCP/cm2

  4. Beam components: 8He (99%) & 3H (1%) Dimension – 25 mm on a D2 gas target Energy resolution after radiation (2×1010 LCP/cm2)ΔE (1 mm Si) vs. Eloss (CsI/PD) plots

  5. Effect of irradiation by fast neutrons, 5 days after(Flux ~ 5×1011 n/cm2 at ACCULINNA - F2) few hours after irradiation

  6. From literature: Increase of the dark current of APD’s during irradiation with 70 MeV proton beam of PSI (top) and 90 MeV protons at KVI (bottom) Prop. CMS PANDA-EMC APD’s S8148 S8664-1010 Area 5x5 mm2 10x10 mm2 Idark 5 nA 10 nA Cterm 80 pF 270 pF (according to authors the beam rate of 9×1012 protons/cm2 is equivalent to 2×1013 of 1 MeV neutrons/cm2) Jump of Idark (510 nA  4 μmA) takes place after an integrated flux of 5×1011 neutron/cm2 PSI That could be dangerous for low energy gamma spectroscopy! KVI

  7. Comparison of CsI/APD and CsI/PD energy resolution Eγ=0.66 MeV ΔE/E=5.8% ΔE/E=12.0% Eγ=1.33 MeV ΔE/E=4.8% ΔE/E=7.1%

  8. Conclusions 1H • For single CsI there is a large loss of events of full energy deposition for high proton energy (E > 200 MeV). • DE-E method for LCP identification will be ambiguous with increasing energy. • High energy γ will fire several CsI crystals. How to reconstruct from the neighbouring crystals (5-10?) the total energywith a good enough resolution ~3-5% ? • Low energy γ spectroscopy will suffer from radiation damage of PD/APD and crystals activation. • A simulation of a neutron field is desirable to estimate deterioration of energy resolution caused by the radiation. γ Has each crystal its specific light-energy relation and has be calibrated individually by LCP and/or γ ?