MS-based methods for protein identification & phosphorylation site analysis

MS-based methods for protein identification & phosphorylation site analysis. 생명과학부 구용의. M ass S pectrometer(MS). 분자가 MS 내로 들어가면 분자는 이온화됨과 동시에 더 작은 이온들 (fragments) 로 쪼개진다 . 쪼개진 이온들은 그들의 질량 / 하전 (m/z) 비에 따라 선택적으로 분리되어 이온수에 비례하게 signal 을 만든다 .

hoshi
Download Presentation

MS-based methods for protein identification & phosphorylation site analysis

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. MS-based methods for protein identification & phosphorylation site analysis 생명과학부 구용의

  2. Mass Spectrometer(MS) • 분자가 MS 내로 들어가면 분자는 이온화됨과 동시에 더 작은 이온들(fragments)로 쪼개진다. 쪼개진 이온들은 그들의 질량/하전(m/z)비에 따라 선택적으로 분리되어 이온수에 비례하게 signal을 만든다. • 이온들의 질량/하전비는 이온들의 생성된 양(Abundance)의 함수로 표시되어 mass spectrum이 그려지며, 이 mass spectrum을 이용하여 미지성분의 정성확인을 할 수 있다. • 이온은 양이온과 음이온 모두 사용

  3. MS Component

  4. MS component • Sample inlet 시료도입장치로 시료를 MS내로 효율적으로 보내주는 역할을 한다 • Ion source 시료분자를 이온화시키고 더 작은 이온으로 쪼갠다. 생성된 이온들을 MS analyzer 쪽으로 이동시킨다 • Mass analyzer 이온들을 m/z ratio에 따라 선택적으로 분리시킨다 • Ion detector 이온 흐름을 그 양에 비례하게 전기적인 흐름으로 전환, 증폭시켜 signal을 생성한다 • Vacuum system MS 내 진공상태를 10-4 ~ 10-9 Torr 로 만들어 주어 최적의 상태로 분석이 진행될 수 있도록 한다 • Data System MS 내 각 구성성분들의 조절이 가능하며 시료분석과 동시에 데이터해석을 할 수 있는 곳이다

  5. Ion source • Gaseous sample introduction - EI(electron ionization) - CI(chemical Ionization) • Liquid sample introduction - FAB(fast atom bombardment) - ESI(electrospray ionization)(soft ionization) • Solid sample introduction - MALDI(soft ionization) (matrix-assisted laser desorption/ionization)

  6. EI(electron ionization) • 전자를 발생시키기 위한 filament가 가열되고 (+) 극판에 전압이 걸리면 가속된 전자들이 흐르고 그 속을 지나던 기체화된 sample은 전자와 충돌하여 에너지를 얻고 전자 하나를 잃어 분자이온 M+가 된다 • 이 방법은 큰 에너지를 사용하기 때문에 복잡한 spectrum을 이루며 분자 이온을 얻기 힘들다 • M + e-⇒ M+ + 2e-

  7. CI(chemical Ionization) • 가열된 filament에서 발생, 가속된 전자는 106정도로 많은 reagent gas와 충돌하여 이들을 이온화시키고 이 reagent gas ion은 sample gas와 충돌 • 이 sample gas는 fragmentation 되기도 하고 때로는 reagent gas ion 과 complex를 이루기도 한다. • 매우 낮은 에너지로 충돌하기 때문에EI 보다 수백배 이상 많은 수의 분자 이온을 만들어 내기 때문에 분자량 확인에 많이 쓰인다. • R(CH4) + e-⇒ R+ + 2e- R+ + M ⇒ M1+ + N1 M1+ ⇒ M2+ + N2

  8. M A L D I

  9. MALDI process • matrix와 sample을 각 1000~10,000 : 1의 농도 비율로 적당한 용매(acidic organic solvent- TFA+MeCN)에 혼합한다 • 이 혼합물을 시료용 probe 에 올려 놓고 진공조건을 만들어 주면 유기용매는 기화되고 시료는 matrix 와 함께 균질하게 결정화 된다 • 이 때 펄스 레이져(UV at 337nm, IR at 2.94um) 를 sample-matrix 결정에 쏘아주면 에너지가 matrix를 통해 sample로 전달되어 약한 이온화가 일어난다. • 이온화는 protonation/deprotonation, cation attachment/cation detachment, oxidation/reduction으로 일어난다.

  10. Matrix in MALDI • Sinapinic acid : peptide, protein>2kDa • 2,5-Dihydroxybenzoic acid : glycoprotein, glycolipids, carbohydrate • A-Cyano-4-hydroxycinnamic acid : low MW peptides, peptide • 2-Benzoic acid : sulfonated dyes • Nicotinic acid/Anthranilic acid(1;1) : oligonucleotide, sialylated glycopeptides • 3-Hydroxypicolinic acid : oligonucleotide adducts, oligonucleotide • 2,4(6) Trihydroxyacetophenone : oligonucleotide, proteins 1-30kDa

  11. E S I

  12. ESI(electrospray ionization) • 시료용액이 고전압이 걸려 있는 capillary를 통과하면서 분무되어 전하를 많이 띤 droplet이 생성됨 • drpolet이 capillary에서 orifice를 지나면서 inert gas(or heat)에 의해 desolvation 됨 • Desolvation 과정에서 ion의 charge는 더 증가하고 “Coulombic expolsion”에 의해 droplet의 ion이 gas phase로 된다. • smaple에 가해지는 충격이 약하며, multiple charge를 가진 peptide이온이 생긴다.

  13. E S I

  14. Mass analyzer and Detector

  15. MALDI – MS(Linear)

  16. TOF(time of flight) • Ion source에서 생성된 이온을 높은 전압으로 가속화시켜 tube(field-free drift tube)를 통과시켜 검출기에 도달하게 한다. • 모든 이온들이 ion source에서 같이 떠난다면 m/z가 작을수록 더 빨리 검출기에 도착한다. • TOF는 이온들의 이동시간을 측정하여 m/z를 계산한다

  17. T O F

  18. Mass resolution Resolution = m/△m

  19. MALDI-TOF MS resolution • Reflectron(ion mirror) - 비행관 끝에 위치한 일련의 정전기장과 자기장으로 이루어 있음 - m/z 값은 같으나 약간씩 다른 운동에너지를 가지고 있는 이온들은 reflectron 에 가까이 도달할수록 속도가 늦어져 정지하여 이온들의 속도차가 줄어들고 reflectron에서 반사되어 detector 쪽으로 비행한다 • Time-lag focusing(delayed extraction) - 이온의 생성과 가속사이에 약간의 시간차를 주어 이온들의 초기운동 속도 분포차를 줄여주는 방법

  20. MALDI – MS(Reflectron)

  21. Quadruple analyzer(mass filter) • 4개의 molybdenum 막대로 이루어져 있으며, 한쌍(1,2)은 dc voltage가 또다른 한쌍(3,4)은 radio frequency voltage가 가해진다. • Dc voltage가 0이고 RF voltage만 있으면 모든 이온이 통과할 수 있다 • 가해지는 전압의 진폭은 선택된 m/z의 비에 해당되는 ion만 ion source에서 detector까지 통과하도록 한다 • 이 quadropole의 전압을 바꾸면서 주어진 mass범위의 이온을 scanning 하여 mass spectrum을 얻는다.

  22. PSD(postsource decay) • The metastable fragmentation of ions after full acceleration that occurs in the field free region of TOF-MS • If peptides are subjected to PSD they will fragment predominantly along th polypeptide backbone, thus generating series of fragment ions which, in principle, contain the amino acid sequence information of the peptide • To obtain primary structural information

  23. MALDI – MS(Reflectron)

  24. MS/MS

  25. ESI –TQ-MS

  26. ESI – TQ - MS • CID(collision-induced dissociation) The term used to describe fragmentation in MS/MS experiment. The precursor ion is isolated and allowed to collide with neutral gas molecules in a collision cell. The translational energy of the precursor ion is converted to internal energy after the collisions resulting in fragmentation of the precursor ion. • quadruple을 사용하는 방법에 따라 여러가지 mode로 사용할 수 있음 - MS mode - MS/MS mode - Neutral loss scan mode - Precusor(or parent) ion scanning - In-source CID : fragmentation occurs in the high-pressure region of an ESI source as a result of collision with atmospheric gases.

  27. Ring electrode와 end-cap electrode로 구성되어 있으며, quadruple처럼 dc voltage와 RF voltage가 흐른다. RF only인 경우 모든 이온들이 trap에 갇히게 됨

  28. ESI –IT-MS

  29. ESI – IT -MS • Tandem mass spectrometer in which ions can be accumulated and stored prior to analysis • The ion trap is both a mass analyzer and collision cell • “Resonance ejection” refers to ions becoming unstable in the trap and being ejected axially through th end-cap electrodes where they are detected • Through this process of trapping and selective ejection of ions, ions of specific m/z can be isolated in the trap

  30. Mass spectrum of CO2

  31. EI-MS spectrum of propionic acid Total ion chromatogram PA

  32. M+ Experimental data Data base Propionic acid

  33. MALDI-MS spectrum of protein

  34. MS/MS spectrum of peptide

  35. Peptide mass searching • Peptides are generated by digestion of the protein of interest using specific cleavage reagents(usually enzymes) • The masses of these peptides are accurately determined experimentally using MALDI-MS(or ESI-MS) • Theoretical peptide masses are calculated for each sequence entry in the database using the same cleavage specificity as the reagent employed experimentally. • A score(or ranking) is then calculated to provide a measure of fit between the experimentally derived and calculated peptide masses.

  36. Measured peptide mass 와 sequence가 맞지 않는 경우 • The additional masses are due to posttranslational or artifactual modifications or post-translational processing • Unspecific proteolysis had occurred or contaminating protease was present • Protein was part of a mixture of ‘contaminating’ proteins

  37. Critical experimental parameter in peptide mass searching • The accuracy of the peptide-mass measurement - time-lag focusing/delayed extraction - internal standard with isotope • The specificity of the enzyme(or chemical reagent) employed - ‘missed cleavage’, ‘ragged termini’를 control할 수 있도록 program화

  38. Orthogonal method in peptide mass searching • Site-specific chemical modification • Determination of partial amino acid composition of the peptide • Identification of the N-terminal amino acid residue • Identification of different cleavage sites within the peptides • Identification of the C-terminal residue(s)

  39. Fragment ion • “b ion” - the fragment with c-terminal deletions and intact N-terminal • “y ion” - the fragment with N-terminal deletions and intact C-terminal • internal fragment - internal acyl ion - immonium ion(represent individual amino acid)

  40. Uninterpreted fragment ion searching • Fragmentation can be induced by PSD-MALDI-MS as well as by CID in triple quadruple or ion-trap mass spectrometer • Fragment ion spectra contain reduntant pieces of information

  41. How to interpret fragment ion • Manual interpretation • Interpret with database - A partial manual interpretation of the spectrum to identify consecutive elements of a particular (b or y) ion series - Uninterpreted fragment ion search program(SEQUEST)

  42. De novo sequencing • Data base independent • “peptide ladder sequencing” - different peptide in length by one amino acid - by chemical(Edman) ⇒ N-terminal blocking : Gln(128.13),Lys(128.17)구별, Ile, Leu 구별못함 - enzymatic degradation : Ile and Leu, Gln and Lys 구별 못함 - are analyzed by MALDI-TOF MS • CID spectra(fragment ion spectra) - are manually interpretated - missing frgament(incomplete ion series) 확인을 위해 trypsin 분해시 H218O사용 (50% H218O+50%H216O, intact C-terminal peptide ion series have doublet by 2u) - methyl esterfication of the carboxyl groups in the peptide (14u increase, derivatized and underivatized 비교 )