1 / 26

Initiating Positive Pressure CMV

Initiating Positive Pressure CMV. A generic approach!. Do these BEFORE attaching the patient. Assemble the ventilator circuit Attach humidifier and fill with water Attach high pressure O2 hose (may also need hose for air) Plug ventilator in. Set the FIO2. If no prior ABGs, start on 100%.

Anita
Download Presentation

Initiating Positive Pressure CMV

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Initiating Positive Pressure CMV A generic approach!

  2. Do these BEFORE attaching the patient • Assemble the ventilator circuit • Attach humidifier and fill with water • Attach high pressure O2 hose (may also need hose for air) • Plug ventilator in

  3. Set the FIO2 • If no prior ABGs, start on 100%

  4. Set desired Vt • 6-12 ml/kg

  5. Set rate • Choose a rate that in conjunction with Vt, gives a minimum Ve of 6-8 LPM • ALWAYS THINK OF MINUTE VENTILATION WHEN SETTING RATE

  6. Set flowrate • Start between 40-60 LPM • Make sure it meets patients inspiratory demand • Steady rise on manometer • Make sure it gives good I:E ratio • The larger the Vt, the faster the flowrate!

  7. Set sigh parameters ( if using) • Sigh Vt = 1.5 x Vt • Sigh rate = 6-10 per hour

  8. Set sensitivity • If pressure triggering, set for 1-2 cmH2O pressure drop • If flow triggering, usually set for 3-5 LPM drop in flow

  9. Turn ventilator on and attach patient Note: make sure high pressure alarms are set at 50 cmH2O to start the patient on CMV

  10. Set high pressure alarms • For Vt, set 5-10 cmH2O greater than peak pressure for Vt • For sigh, set 5-10 cmH2O greater than peak pressure for sigh

  11. Set low exhaled volume alarm • Set 100 ml less than Vt

  12. Set low pressure alarm • Set 5-10 cmH2O less than peak pressure for Vt

  13. Alarms and Monitors What they mean and what to do

  14. Assist Light • Shows patient triggered the breath • If patient is not triggering and light is on, then ventilator is too sensitive • autocycling

  15. Pressure Light • Shows ventilator pressure limited the breath • Also has audible alarm

  16. Pressure limiting may be caused by: • Secretions • Water in the tubing • Physical obstruction or a kink in tubing • Change in patient position • ET tube in right main stem bronchus • Pneumothorax • Maladjusted control

  17. Indicates increased resistance Also a decrease in compliance

  18. Remember: Flow =pressure/resistance

  19. Ratio Light • Indicates inverse I:E • Inspiration is as long or longer than expiration • Increase flow!!!!!

  20. Oxygen Lights • Red = FIO2 set to greater than 21% but not hooked up to source of O2 • Usually has audible too • Green = FIO2 is greater than 21% • Many ventilators only check for pressure on the O2 line

  21. Low exhaled Vt/Low pressure • Indicates a leak between ventilator and patient • The closer the leak is to the ventilator, the lower the manometer pressure

  22. Low exhaled Vt/pressure causes: • Patient disconnect • Leaking humidifier (very common) • Leak in exhalation valve • Loose nebulizer • Loose tubing connection • Leak in ET tube • Alarm maladjustment

  23. Tubing compliance • Tubing expands as it is pressurized • Some of Vt from ventilator is lost to tubing expansion • Nondisposable circuit compliance is about 3 ml/cmH2O

  24. To figure out how much Vt is lost to tubing expansion: • Multiply peak pressure by tubing compliance (usually 3 ml/cmH2O) • Subtract this number (tubing expansion volume) from the set Vt • The difference is known as effective or corrected Vt • Effective or corrected Vt is the Vt the patient actually gets!

  25. Flow sheet • Accumulated data • Makes sure patient is being ventilated effectively • Allows trend analysis to aid in decisions about ventilator management • Filled out periodically, eg every 2 hours

  26. Time to Rock and Roll!

More Related