330 likes | 997 Views
SCATTER. The Issue: 50-90% of x-rays reaching receptor are scatter, which degrades image contrastFactors Affecting ScatterScatter increases with amount of tissue exposedScatter increases with kVp. Scatter and X-ray Field Size. Scatter Escape Angle:. scatter is emitted in all directionsonly scatter reaching image within area of interest (field size) degrades contrast. The smaller the field, the larger the escape angle.
E N D
1. X-RAY SCATTER
2. SCATTER
The Issue: 50-90% of x-rays reaching receptor are scatter, which degrades image contrast
Factors Affecting Scatter
Scatter increases with amount of tissue exposed
Scatter increases with kVp
3. Scatter and X-ray Field Size
4. Scatter Escape Angle: scatter is emitted in all directions
only scatter reaching image within area of interest (field size) degrades contrast.
The smaller the field, the larger the escape angle
5. FACTORS AFFECTING SCATTER Scatter increase with amount of tissue exposed
Field size: irradiated volume and escape angle
Thickness: irradiated volume
Levels off at some thickness/size (scatter range)
Scatter increases with kVp
Higher fraction of Compton (source of scatter)
Scatter in more penetrating: more reaches film
Scatter is more biased forward (toward film)
6. CONTROLLING SCATTER
Minimize Scatter Production
Lower kVp (but usually fixed by penetrability)
Reduce thickness (eg, compression in mammo)
Minimize X-ray field size: Collimation
Prevent Scatter From Reaching Image
Grids
Air gap
7. COLLIMATION
8. Functions of X-ray Beam Restrictors Minimize Patient Radiation Dose:
Irradiates only necessary tissue
“Square” Function: Proportional to beam area e.g.: reducing 10x10” field (100 in2) to 8x8” field (64 in2) reduces irradiated tissue by 36%
Minimize Scatter:
Improves image contrast
Scatter reduced in ~proportion to beam area
9. Types of Restrictors: Apertures, Cones
10. Apertures and Cones Beam defined by fixed “hole” (aperture)
Penumbra depends on distance of defining aperture from focal spot (ie-top or bottom)
Used if small number of beam sizes/shapes
Examples: Dental, Mammography
11. X-ray Beam Collimators
13. Functions ofX-ray Beam Collimators:
14. Collimator Requirements Continuous range of size (down to 2x2” or smaller) and shape (rectangular) by pairs of independent blades
Light field indication (must indicate x-ray edges to <2% os SID)
Penumbra reduced by upper/lower shutters
15. Collimator Congruence Testing
16. SCATTER AND GRIDS
17. GRID TERMINOLOGY Types of Grids
Linear grid: Strips parallel to table (and tube)
Crossed grid: superimposed perpendicular grids
Focused grid: grid lines “converge” to focal line
Parallel grid : focused at infinity
Grid Numbers:
Grid Ratio: predicts scatter-removal ability
Grid Frequency: grids lines per inch
18. X-RayScatterRemoval Grids
19. Bucky’s Original Grid
20. Linear Grid
21. Crossed Grid (Superimposed Linear
22. Focused Grid
23. Parallel Grid (Focused at Infinity)
24. Grid Ratio and Scatter Clean-up
25. Grid Ratio and Scatter Clean-up-con’t
26. Grid Ratio Selection Trade-off: primary absorption and scatter removal
High scatter exams (high kV, thick part): 10:1up to 16:1
Otherwise: 8:1
27. Grid Performance Grid Primary Transmission
Bucky Factor (Tube Load, Patient Dose)
Contrast Improvement Factor (“real” test)
Lead Content (assuming “good” grid design)
Grid Cutoff
28. Grid Performance: Bucky Factor The Bucky Factor is the ratio of incident radiation intensity reaching the grid to the transmitted radiation intensity passing through the grid. This is the factor by which mAs --- and thus patient dose --- must be increased when using the grid.
29. Grid Performance: Bucky Factor
30. Contrast Improvement Factor The Contrast Improvement Factor (K) is the ratio of contrast with the grid to contrast without the grid. Since the purpose for using the grid is to improve contrast (scatter removal is just a means to that end), many consider this the “real” test of a grid
31. Contrast Improvement Factor
32. Grid Performance: Lead Content
33. Grid Cutoff Upside Down
Lateral Decentering
Distance Decentering
Combined lateral+distance decentering
Angulation
34. Grid Cutoff Upside Down Lateral Shift
35. Lateral Decentering Grid Cutoff (40” Focal Distance)
36. Grid Cutoff: Distance
37. Distance Decentering Grid Cutoff (40” Focal Distance)
38. Grid Cutoff: Combined/Angulation
39. Reciprocating Grid: Bucky Factor
40. Scatter Removal: Air Gap