1 / 8

Q-Lab Thermal Cycler

<br><br>Q-Lab Thermal Cycler has high precision 6 Peltier heating units to heat up the sample with 3 variations in temperature which can process more than 100 cycles. It can <br>accommodate 120u00d780 mm wells or tubes. Continual information of the operation status is seen on graphical 8-inches TFT colour touch-screen. Sophisticated operating <br>concept with windows operation system for error free processing of samples. This device can be applied across a temperature range of 0 to 100 u2103 with maximum <br>ramping rate is 4 u2103/s. The user can secure his data using password and user login method and use pre-installe

Kimber4
Download Presentation

Q-Lab Thermal Cycler

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Q-Lab Thermal Cycler www.labmate.com | info@labmate.com Overview

  2. Q-Lab Thermal Cycler has high precision 6 Peltier heating units to heat up the sample with 3 variations in temperature which can process more than 100 cycles. It can accommodate 120×80 mm wells or tubes. Continual information of the operation status is touch-screen. Sophisticated operating concept with windows operation system for error free processing of samples. This device can be applied across a temperature range of 0 to 100 ℃ with maximum ramping rate is 4 ℃/s. The user can secure his data using password and user login method and use pre-installed 11 standard protocols. The files can be transferred via USB, printer support. It can data in the10,000 files data in internal flash memory. seen on graphical 8-inches TFT colour Specifications : Capacity 120×80 mm

  3. Temperature Range 0 to 100 °C Maximum. Heating 4 °C/s Rate Maximum. Cooling 3 °C/s Rate Uniformity ≤ ± 0.2 °C Accuracy ≤± 0.1 °C Display Resolution 0.1 °C Temperature Control Block /Tube Ramping Rate 0.1 to 4 °C Adjustable

  4. Hot Lid Temperature 30 to 110 °C Hot Lid Height Step less Adjustable Adjustable Number Of Programs 10000+(USB FLASH) Maximum. No. Of Step 30 Maximum. No. Of Cycle 100 Increment/Decrement 1Sec to 600 Sec Time Increment/Decrement 0.1 to 10.0 °C Temperature Pause Function Yes

  5. Auto Data Protection Yes Holding 4 °C Landing Experiment Yes PCR Experiment Yes Print Yes LAN To Computer Yes LCD 8inch,800×600Pixels,2600 00 colors Interface USB2.0,LAN Dimensions(L×W×H) 390×270×255 mm

  6. Weight 9.0 kg Power Consumption 600 W Power Supply 85 to 264 V AC,47 to 63 Hz Features : - It has graphical 8-inches touch screen display which enables the user to change the commands - Uniform heating with 3 different temperature of the samples is achieved with Peltier heating units with more than 100 cycles - It is well built with anodizing technology with aluminum block treatment which makes the durable and corrosion resistant - Maximum heating rate is 4 ℃/s with temperature range of 0 to 100 ℃

  7. - It has the capability to accommodate 120×80 mm wells with moveable hot lid which can fit different sizes of tubes - It has the internal flash memory to store the 10000 files data for analysis - The connectivity between computer is via USB port and mini printer and can store the GLP data a?er every step to get reliable results - All the operations are controlled by a computer so?ware system which instrument PCR protocols and also has in built 11 standard protocols - It has power of operation recovery option which helps in continuing the same command a?er a recovering from power cut - The user can access to his individual experimental data by setting a password and data management system - To satisfy the customers need it is installed with special feature like the user can add sample

  8. while the process is going on to encounter different variations - The user will be sent notifications via email about the process to alert when operation ends Applications : Used detection, DNA methylation analysis, Genotyping and gene scanning, miRNA research etc. in Gene expression analysis, Pathogen Labmate Scientific Email: info@labmate.com | Website: www.labmate.com