An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. First TimePix tests at CERN David Attié ECFA Meeting – Valencia – 8 November 2006 ECFA Meeting, Valencia – November 8, 2006

  2. Introduction • TimePix Proposal: November 2004 • develop • produce Medipix2  TimePix • test • TimePix test at CERN ~ November 2006 Medipix2 TimePix ECFA Meeting, Valencia – November 8, 2006

  3. TimePix measurements • Chip design was submitted to foundry on the beginning of July • 12 wafers arrived at CERN 2nd week of September • Modifications in the Medipix readout system (~1 week): • “Timepix-Medisoft” version of the software • MUROS2 firmware updated (thanks to Hans Verkoojen, Nikhef) • Preliminary TimePix characterization has been done on-wafer • using the Medipix2 probing card in the DSF clean room facility at CERN ECFA Meeting, Valencia – November 8, 2006

  4. Test procedures for TimePix * Alignment monitor (moving platform + microscope) to position the chip * Readout using probing bench + MUROS2 * Check the current supply (~200 mA/chip) and ID chip • * Software used : Medisoft 4.x • Analogic tests (ANIN) • TOT mode • MEDIPIX mode • Digital tests using calibration of DACs  Classification (A, B, C, …) ECFA Meeting, Valencia – November 8, 2006

  5. Test procedures for TimePix * Alignment monitor (moving platform + microscope) to position the chip * Readout using probing card of Medipix2 + MUROS2 * Check the current supply (~200 mA/chip) and ID chip • * Software used : Medisoft 4.x • Analogic tests (ANIN) • TOT mode • MEDIPIX mode • Digital tests using calibration of DACs  Classification (A, B, C, …) ECFA Meeting, Valencia – November 8, 2006

  6. Pixel measurement • From 2 special pixels at the bottom of the matrix • we can measure: • preampOut • discOut • internal Ref_Clk • counter clock • State Machine of • the counter Modes (P0, P1) • Medipix (0, 0) • TOT (0, 1) • TimePix-1hit (1, 0) • TimePix (1, 1) • work as expected in simulations • No visible coupling of • the Ref_Clk signal • into the analog signals • (preampOut and discOut) TOT mode (P0=1; P1=0) TimePix mode (P0=1; P1=0) TimePix-1hit mode (P0=0; P1=1) ―Shutter ―PreampOut ―Thr ―discOut ―CounterClk ECFA Meeting, Valencia – November 8, 2006

  7. TimePix DACs • Some DACs show a non-monotonic curve • Non- Monotonic DACs are: • Preamp • Disc • BufferA • BufferB • Problem may be related • to the latching of the DACs register • The midrange value (code 01111111=127) • is always stable (1 complete wafer) • These DACs are unchanged • under normal operating conditions ECFA Meeting, Valencia – November 8, 2006

  8. TimePix chip classification 14 mm • Classification of a tested chip: • Class A: no dead column • Class B: 1 dead column • Class C: 2 dead columns • Class D: more than 2 dead columns • Class E: bad DACs • Class F: bad digital test, no communication, etc… •  dead chip ECFA Meeting, Valencia – November 8, 2006

  9. First TimePix test Wafer #14: A5FWQTX Tested by Xavier Llopart (CERN) ECFA Meeting, Valencia – November 8, 2006

  10. First TimePix test Wafer #15: AQFWTPX Tested by: - David Attié (CEA/Saclay) - Uwe Renz (Freiburg) ECFA Meeting, Valencia – November 8, 2006

  11. Future development + Medipix2/TimePix Micromegas • no ExB effect • good for 2-track separation • high gain • very fast electron signal on the • anode plane • very small ion feedback in the • space cheap • robust and easy to implement • 55 mm pixels • Direct X-ray conversion • selection of window in energy • 13-bit counter per pixel Mini-chamber SiTPC endplate+Large Prototype ECFA Meeting, Valencia – November 8, 2006

  12. Next steps • First test of a Medipix2 + Micromegas in the gas • Improve the gas mixture (“non-sparking” state) • Medipix2  TimePix readout chip  LP endplate M. RIALLOT (DAPNIA/SEDI) ECFA Meeting, Valencia – November 8, 2006

  13. Box design Metallic box of 24 cm x 14 cm x 6 cm size M. RIALLOT (DAPNIA/SEDI) ECFA Meeting, Valencia – November 8, 2006

  14. Mini-chamber for the Medipix2/TimePix chip Design plan: Windows for X-rays source Box 2.5 cm 6 cm Windows for b source Field cage Micromegas mesh M. RIALLOT (DAPNIA/SEDI) Medipix2 readout chip ECFA Meeting, Valencia – November 8, 2006

  15. Conclusions * 2 wafers and half have been tested at CERN * ~ 50 % of class A ~ 20 % of class B ~ 30 % of class C/D/E/F * Test to be continued in the next weeks * First test beam in DESY right now with Medipix/TimePix * CEA/Nikhef: test in gas chamber in the next months with Medipix/TimePix NIKHEFMaximilien Chefdeville, Martin Fransen, Harry van der Graaf, Jan Timmermans, Jan Visschers CEA/Saclay/David Attié, Dan Burke DAPNIAPaul Colas, Arnoud Giganon, Yannis Giomataris Univ. Twente/Jurriaan Schmitz, Cora Salm, Mesa+Victor Blanco Carballo, Sander Smits CERNMichael Campbell, Erik Heine Xavi Llopart Thanks to: Wim Gotink Joop Rovenkamp Marc Riallot ECFA Meeting, Valencia – November 8, 2006