eLigo Photon Calibrator investigation: long-term stability of DARM actuation

eLigo Photon Calibrator investigation: long-term stability of DARM actuation. Rick Savage Jonathan Berliner, Pablo Daveloza, Roberto Grosso, Greg Mendell, Michael Sakosky, Ryan DeRosa, Matt West. Photon Calibrator method.

alika
Download Presentation

eLigo Photon Calibrator investigation: long-term stability of DARM actuation

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. eLigo Photon Calibrator investigation:long-term stability of DARM actuation Rick Savage Jonathan Berliner, Pablo Daveloza, Roberto Grosso,Greg Mendell, Michael Sakosky, Ryan DeRosa, Matt West LVC meeting - Arcadia, CA March 2010

  2. Photon Calibrator method Power-modulated auxiliary laser to displace the mass via the recoil of the photons (in “Science mode” configuration). E. Goetz, et al. Class. Quantum Grav. 26 (2009) LVC meeting - Arcadia, CA March 2010

  3. eLigo Pcal upgrades Integrating spheres for absolute power calibration 1047 nm Nd:YLF laser Acousto- optic modulator Improved optical layout and enclosures LVC meeting - Arcadia, CA March 2010 • Upgraded Pcals installed for S6. H1 and L1 X-arm end test masses (Matt West, Ryan DeRosa, Michael Sakosky)

  4. Calibrating DARM_CTRL_EXC with Pcal Coil – 393.1 Hz Pcal – 401.2 Hz Simultaneously drive ETM position with both Pcal and voice coil actuators with sinusoidal excitations separated by 7.1 Hz Ratio of peaks together with Pcal absolute power calibration yields on-line, continuous calibration of DARM_ERR LVC meeting - Arcadia, CA March 2010

  5. DARM_ERR lines with ~ 60-sec FFTs ASD of DARM_ERR Voice coils Pcal Excitation frequencies: H1: coils – 393.1 Hz; Pcal – 400.2 Hz L1 : coils – 396.7 Hz; Pcal – 403.8 Hz LVC meeting - Arcadia, CA March 2010

  6. Relevant LLO Pcal peaks – high BW LVC meeting - Arcadia, CA March 2010

  7. Relevant LLO Pcal peaks – Low BW LVC meeting - Arcadia, CA March 2010

  8. DARM actuation long-term trend investigation S/N ~ 200 S/N ~ 70 • C++ code running on cluster atCaltech (LHO or LLO)- level 1 RDS frames- one S6 day per core • 60 sec. data segmentsfor three channels- DARM_ERR- 2 excitation readback channels • Hann window data- all lines bin-centered • FFT using FFTW3 • Write- GPS time at start of data segment- minimum state vector value during segment- complex amplitude for four lines 2 DARM_ERR 2 excitation readbacks LVC meeting - Arcadia, CA March 2010

  9. 60-sec FFT Science Mode data: H1 Higher S/N Pcal beam spot loctions changed Violin notches removed - response function change LVC meeting - Arcadia, CA March 2010

  10. 60-sec FFT Science Mode data: L1 Swap of coil driver Shorted wire in Pcal internal power sensor LVC meeting - Arcadia, CA March 2010

  11. Characterization of data LVC meeting - Arcadia, CA March 2010 • Outliers • Glitches • Data errors • Distribution • Correlated noise due to laser speckle • etc., etc. • Focus on goal of assessing long-term stability of DARM acutation

  12. One-hour averages – Science Mode: H1 LVC meeting - Arcadia, CA March 2010 • DARM_CRTL_EXC calibration (> 8 mos of S6) • Mean 1.28 nm/ ct ; standard deviation 0.28%

  13. One-hour averages – Science Mode: L1 • DARM_CTRL_EXC calibration (> 8 mos of S6) • Mean 0.53 nm/ ct ; standard deviation 0.40% LVC meeting - Arcadia, CA March 2010

  14. Conclusions – Future work • Long-term stability of actuation • Photon calibrator calibration and DARM actuation (coil driver and voice coils) stable within fraction of a percent over ~ 8 months • Continue to monitor during remainder of S6 • Future work: • Use relative phase between DARM_ERR and Pcal readback peaks to investigate phase jumps due to computer reboots, etc. • Generate Pcal data using DMT monitor (R. Grosso) • Drive Pcal with GPS-synched aLIGO timing module for absolute phase reference (with I. Bartos, et al.) • Investigate source of outliers • Investigate if laser speckle is fundamental limit to Pcal precision using integrating spheres (J. Berliner) LVC meeting - Arcadia, CA March 2010