1 / 88

CT Scanning

CT Scanning. Dr. Craig Moore Medical Physicist & Radiation Protection Adviser Radiation Physics Service CHH Oncology. Brief History of CT Scanning. First CT Scanner - 1972. Originally called CAT A = axial 80 x 80 resolution 4 min. per rotation 8 grey levels overnight reconstruction.

anikam
Download Presentation

CT Scanning

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. CT Scanning Dr. Craig Moore Medical Physicist & Radiation Protection Adviser Radiation Physics Service CHH Oncology

  2. Brief History of CT Scanning

  3. First CT Scanner - 1972 • Originally called CAT • A = axial • 80 x 80 resolution • 4 min. per rotation • 8 grey levels • overnight reconstruction

  4. Here and Now • 512 x 512 or 1024 x 1024 resolution • Sub second rotation • 4096 grey levels • 100’s slicesper rotation

  5. Components of a CT Scanner

  6. Principal components of CT scanner • X-ray tube, collimator, and detector array on a rotating gantry • Rotation axis is referred to as Z axis • Fan beam wide enough to cover patient cross-section • Narrower width in the z-axis • Behind the patient is a bank of detectors • Patient lies on a couch that is moved longitudinally through the gantry

  7. X-ray Tube • Tube parallel to patient movement – minimise anode heel effect • X-rays are produced by firing electrons at a metal target – typically tungsten • Capable of producing long exposure times at high mA – get very hot (require heat capacities up to 4MJ and active cooling mechanisms) • Continuous scanning limited to around 90s • Focal spot size typically 0.6 - 1mm • High kVps (80 – 140) and beam heavily filtered (6-10 mm Al filters) to optimise spectrum • Stops attenuation coefficients varying with depth via beam hardening

  8. Collimation and Filtration • Want a monoenergetic beam to avoid beam hardening artefacts • As beam passes through patient low energies are filtered • This results in the apparent reduction of attenuation and CT number of tissues • Computer reconstruction assumes monoenergetic beam • Not possible with X-ray tubes so they are heavily filtered • At least 6 mm aluminium or copper and typically high kVp • Some manufacturers use shaped filters such as bow-tie filters to even out the dose distribution (conform to the shape of an elliptical patient) • Pre-patient collimator is mounted on the X-ray tube • Beam is approx 50cm wide to cover cross section of patient • The size is variable in the z-axis • Multi-slice scanners between 0.5 and 40 mm thick beams • Post patient collimation is not used with multi-slice scanner

  9. Bow Shaping Filters • Body is approx circularly shaped • Shape of beam that reaches detector is therefore non-uniform which can lead to image artefacts • Shaping filters are applied to the beam prior to make the dose distribution more uniform

  10. Detectors • Requirements: • Small enough to allow good spatial resolution • Up to 1000 detectors per scanner • Typically 1.5 mm width but can be a small as 0.5mm • High detection efficiency • Fast response • Wide dynamic range – massive variation in X-ray intensity • Stable and noise free • No afterglow • There needs to be separation between detectors to prevent light crossover • This reduces efficiency from 98% to 80%

  11. Detectors • Modern CT detectors are fabricated as detector modules • A side view shows individual detector elements coupled to photodiodes that are layered on electronics • Spaces in between filled with optical filler to reduce cross talk

  12. Detectors

  13. Generations of CT Scanner

  14. CT Imaging • Conventional radiography suffers from the collapsing of 3D structures onto a 2D image • However, CT scanning has extremely good low contrast resolution, enabling the detection of very small changes in tissue type • Almost true depiction of subject contrast • CT gives accurate diagnostic information about the distribution of structures inside the body • Generation of images in transaxial section • Perpendicular to the axis of rotation of the X-ray tube about the body • Perpendicular to the craniocaudal axis of patient

  15. Number of detectors and projections • Typically, for a 3rd generation scanner: • 650 – 900 detectors • 1000 to 2000 projections per rotation

  16. Image contrast 2:1 Collapse of 3D Data into 2D Plane • Planar imaging • 2D representation of 3D Distribution of Tissue • No depth information • Structures at different depths are superimposed • Loss of contrast Subject Contrast 4:1 X rays

  17. Typical CT Image

  18. CT Images • Commonly calculated on 512x512 matrix, but 256x256 and 1024x1024 are also used • Each pixel is more accurately described as a voxel, because it has depth information • The value stored in each voxel is referred to as the CT number which is related to the attenuation of a particular tissue: • CTn = 1000 x (µt - µw)/ µw • Sometimes referred to as Hounsfield Units • Each CT number is assigned a certain shade of grey in the resulting image • CT number represents x-ray attenuation coefficient of the corresponding voxel within the patient

  19. CT Numbers

  20. Image Display • CT image represented by a range of CT numbers from -1000 to + 3000 (ie 4000 levels of grey) • Human eye dose not have the capacity to distinguish so many grey levels • If 4000 shades of grey displayed altogether there would be very little difference between different tissues

  21. Window Width and Level • The appearance of the image on the screen can be changed by altering the window width and level • Window width refers to the range of CT numbers selected for display • This range of CT numbers is centred at a particular level called the window level • e.g. if imaging bone window level should be ~1000 • Can spread a small range of CT numbers over a large range of grayscale values

  22. Window Level –593 Window Width 500 Good contrast in lungs Only see CT numbers +/- 250 around -593 Window Level –12 Window Width 400 Good soft tissue contrast Only see CT numbers +/- 200 around -12

  23. How do we get the images? • Tube and detector rotate smoothly around the patient • X-rays are produced continuously and the detectors sample the X-ray beam approx 1000 times during one rotation • Typically 2 to 4 revolutions per second

  24. In reality not always parallel to detectors • Each voxel is traversed by one or more x-ray beams for every measurement (1000 per rotation) • Number of measurements taken in single axial section depends on • number of detectors • Number of samples per rotation • Assume 800 detectors measured at 0.5° intervals per 360 ° rotation • This is 576,000 measurements • More than needed as we only need 260,000 measurements (512 x 512)

  25. How do we get the picture? • Back Projection • Reverse the process of measurement of projection data to reconstruct image • Each projection if smeared back across the reconstructed image

  26. Back Projection – the basics • Consider cylindrical uniform body with a hole down the centre • A beam passing through this body from one direction will have a transmitted profile in its central region • This single measurement cannot determine the position of the hole other than identifying that it is in the line of the pencil beam passing through the centre of the body • Pixel values along this line are decreased by the amount of attenuation measured • These values are projected back along the field of view • A second projection at 90° provides a second band of grey • This is then projected back across field of view • Progressive projections are shown in the final figure – a star like pattern • We now have an image that looks similar to what we are scanning

  27. Back Projection • Back Project each planar image onto three dimensional image matrix 3 6 3 3 3 6 6 3 3 3 6 3

  28. Back Projection • Back Project each planar image onto three dimensional image matrix 3 6 3 1 2 1 1 2 1 1 2 1

  29. Back Projection • Back Project each planar image onto three dimensional image matrix 3 6 3 3 2 2 1 3 2 1 2 1 2 1 6 3 4 3 3 2 1 2 3 1 2

  30. Back Projection • Back Project each planar image onto three dimensional image matrix 3 6 3 6 3 4 4 3 6 6 6 8 6 6 3 4 4 3 3 6 3

  31. Back Projection • Back Project each planar image onto three dimensional image matrix 3 6 3 6 3 4 4 3 6 6 6 8 6 6 3 4 4 3 3 6 3

  32. Back Projection • More views – better reconstruction • 1/r blurring, even with infinite number of views

  33. Filtered Back Projection • Back projection produces blurred transaxial images • Projection data needs to be filtered before BP • Different filters can be applied for different diagnostic procedures • Smoother filters for viewing soft tissue • Sharp filters for high resolution imaging • After filtration, back projection same as before • Data from neighbouring beams are used • Some data is subtracted • Some data is added • Filters are convolved with the blurred image data in Fourier Space

  34. Filtered Back Projection

More Related