An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Atomic beam studies in the H-jet polarimeter. A.Zelenski PSTP2007, September 11,BNL

  2. H - jet polarimeter. Dissociator

  3. Compression tube calibration. Ø -10 mm. H0 • Hydrogen mass-flow controller MKS. Full range: 0.0-1.0 scc/min. Absolute accuracy 1-2%. • Conventional technique: pressure drop in calibrated volume. • Independent AB intensity measurement from the well known TMP pumping speed. Length-10 cm H2 Hot filament ion gage.

  4. The compression tube calibration system for the absolute AB intensity measurements. H-beam Ion gage Ion gage H2 mass-flow controller and a pressure drop measurement in the calibrated volume were used for compression tube calibration.

  5. Atomic beam intensity vs H2 flow in dissociator. RF-power was kept constant at 260 W Nozzle temperature 75 K Slope is 1.75 steeper Than simulations H2 flow, cm3/s

  6. QMA upgrade for H-jet measurements. Original QMA geometry (right) Expanded QMA sensitive volume (left) 10 mm

  7. Atomic beam intensity and density measurements in the collision region • H-beam intensity and density vs. H2 flow in dissociator.

  8. QMA calibration. 10-7 torr.H2 0.98·10 10 H2/cm3 Measured H2 / H density ratio: ~1.5%× 2  3%-polarization dilution. • QMA atomic beam signal vs. H2 partial gas pressure in the collision chamber.

  9. Layout of the electron beam ionizer and magnetic ion analyzer in the collision chamber. Electron gun-600 eV Ionization, extraction region RHIC beams Analyzing magnet FC Electrostatic optics for H+, H2+ ions transport. Einzel lenses Recoil arms.

  10. Ion spectra measured with the new diagnostic device. Atomic beam “on”. Calibration with H2. H H2 H H~1.2% H2~3% H2 H2O Cross-section ratio H2/H ~15 at 600 eV electron beam energy.

  11. H-jet 12.5·1016 at/s Operational atomic beam sources intensities. T.Wise calc. ANKE HERMES Maximum H-jet beam intensity is at Tnozzle~75 deg. K

  12. Atomic beam intensity dependence on H2 flow in dissociator. AB intensity at 65 scc/s H2 flow is 12.5·1016 at/s.

  13. AB intensity vs. nozzle temperature. The maximum AB intensity was measured at Tnozzle =75 deg. K

  14. Atomic beam intensity profile measurements. Beam profile at the entrance of 6-pole #5. Beam profile at the RHIC beam collision point.

  15. H-jet sextupole separation magnet system. 24 sectors separating magnets with 1.5 T field at the pole tips.

  16. FWHM 5.0 mm

  17. H-jet target intensity profile at the RHIC collision point. Profile was measured from elastic pp scattering events. FWHM ~ 6 mm in agreement with a compression tube measurements. Number of elastic pp events Hor. position of the JET ,10000 steps = 2.5 mm.